A Low-Cost Dual-Site Photoplethysmography Platform for Millisecond-Resolution Pulse Arrival Time Difference: Design, Timing-Error Analysis, and Validation Protocol
This research presents the design and timing-error analysis of a low-cost dual-site photoplethysmography (PPG) platform for measuring millisecond-resolution pulse arrival time differences. The system integrates two MAX30102 optical sensors, an ESP32-C3 microcontroller, and a TCA9548A I2C multiplexer to acquire synchronized pulse signals from two measurement sites. It employs hardware interrupt-based timestamping and an advanced pulse-foot detection algorithm to improve timing accuracy. Monte Carlo simulations investigate the effects of sampling frequency, signal noise, and detection methods on measurement error. Additionally, a three-experiment validation protocol evaluates signal quality, timing accuracy, and physiological plausibility. The platform aims to provide an affordable research tool for investigating peripheral vascular characteristics without making clinical diagnostic claims.
Authors
- Moaaz Ahmed Abdlhamed Mohammed Salem (ORCID: https://orcid.org/0009-0001-8982-2128)
Publication Details
- Journal
- Zenodo (CERN European Organization for Nuclear Research)
- Published
- 2026-11-27
- DOI
- https://doi.org/10.5281/zenodo.24595148
- Primary Topic
- Non-Invasive Vital Sign Monitoring
- Type
- preprint