Agarose–Gelatin Hydrogel Phantoms with Tunable Hemoglobin Absorption for Hematocrit-Graded Calibration of Near-Infrared Optical Sensors

Optical tissue phantoms are essential for biomedical spectroscopy and sensor calibration, particularly for non-invasive hematocrit (Hct) sensing, for example, in dengue, where hemoconcentration signals plasma leakage. Reproducible blood-mimicking hydrogels with tunable, Hct-graded absorption and a temporally stable matrix remain challenging to develop. Agarose, gelatin and agarose–gelatin composite matrices were evaluated for background absorption, spectral neutrality, homogeneity and stability over 40 h. A composite of 0.15% agarose and 4% gelatin gave a featureless near-infrared (NIR) background (μa ≤ 0.015 cm−1 at 805 nm) and lower optical drift than either single-polymer matrix. Ten Hct-graded phantoms (10–90% equivalent) containing dissolved bovine hemoglobin (2.1–23.4 g/L) were characterized by transmission spectrophotometry at 740, 805, 900 and 940 nm. Absorption was linear in hemoglobin loading at all wavelengths (R2 = 0.9999), and two independent preparations agreed within 5.8–8.0% (mean absolute deviation), equivalent to between-series prediction deviations of 2.7–4.4 Hct points. Phantom absorption was about one tenth that of whole blood and spectrally flatter: per gram of hemoglobin, it matched oxygenated whole blood at 900–940 nm (0.93–1.04×) but exceeded it at 740 and 805 nm, consistent with non-oxygenated hemoglobin species. Hemoglobin-loaded phantoms drifted by up to 26% over 40 h at intermediate Hct. The composite provides a tunable, reduced-concentration absorption standard for wavelength-specific calibration of NIR Hct sensors, best used within a defined period after preparation.

Authors

Institutions

Publication Details

Journal
Biosensors
Published
2026-11-27
DOI
https://doi.org/10.3390/bios16120691
Primary Topic
Optical Imaging and Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Agarose–Gelatin Hydrogel Phantoms with Tunable Hemoglobin Absorption for Hematocrit-Graded Calibration of Near-Infrared Optical Sensors

Murad Althobaiti, Gameel Saleh, Haya Hamed, Batool Bokhedher et al.
Biosensors
Optical Imaging and Spectroscopy Techniques
article

Agarose–Gelatin Hydrogel Phantoms with Tunable Hemoglobin Absorption for Hematocrit-Graded Calibration of Near-Infrared Optical Sensors

Murad Althobaiti, Gameel Saleh, Haya Hamed, Batool Bokhedher, Jana Aldulaijan
article en

Abstract

Optical tissue phantoms are essential for biomedical spectroscopy and sensor calibration, particularly for non-invasive hematocrit (Hct) sensing, for example, in dengue, where hemoconcentration signals plasma leakage. Reproducible blood-mimicking hydrogels with tunable, Hct-graded absorption and a temporally stable matrix remain challenging to develop. Agarose, gelatin and agarose–gelatin composite matrices were evaluated for background absorption, spectral neutrality, homogeneity and stability over 40 h. A composite of 0.15% agarose and 4% gelatin gave a featureless near-infrared (NIR) background (μa ≤ 0.015 cm−1 at 805 nm) and lower optical drift than either single-polymer matrix. Ten Hct-graded phantoms (10–90% equivalent) containing dissolved bovine hemoglobin (2.1–23.4 g/L) were characterized by transmission spectrophotometry at 740, 805, 900 and 940 nm. Absorption was linear in hemoglobin loading at all wavelengths (R2 = 0.9999), and two independent preparations agreed within 5.8–8.0% (mean absolute deviation), equivalent to between-series prediction deviations of 2.7–4.4 Hct points. Phantom absorption was about one tenth that of whole blood and spectrally flatter: per gram of hemoglobin, it matched oxygenated whole blood at 900–940 nm (0.93–1.04×) but exceeded it at 740 and 805 nm, consistent with non-oxygenated hemoglobin species. Hemoglobin-loaded phantoms drifted by up to 26% over 40 h at intermediate Hct. The composite provides a tunable, reduced-concentration absorption standard for wavelength-specific calibration of NIR Hct sensors, best used within a defined period after preparation.

BiosensorsVol. 16(12)
Imam Abdulrahman Bin Faisal University (SA)
Optical Imaging and Spectroscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Agarose–Gelatin Hydrogel Phantoms with Tunable Hemoglobin Absorption for Hematocrit-Graded Calibration of Near-Infrared Optical Sensors — Murad Althobaiti, Gameel Saleh, et al. · Biosensors (2026) | TGRS Research Map | TGRS