Influence of the volume of laboratory-developed simulators on the counting efficiency (kCPS/MBq) of the PET scanner belonging to LIM/CDTN

This study investigated the influence of phantom volume on the conversion coefficient between count rate (CPS) and activity (MBq) for a preclinical Positron Emission Tomography (PET) scanner, the Triumph® LabPET Solo 4, installed at the LIM/CDTN (Laboratório de Imagem Molecular do Centro de Desenvolvimento da Tecnologia Nuclear). Three cylindrical phantoms (S, M, and L) with different internal volumes (17.7 mL, 22.0 mL, and 26.5 mL, respectively) were fabricated. PET images were acquired using two approaches: spatial and temporal. The spatial acquisition mode was used to verify the quality and homogeneity of the reconstructed PET images for the different phantom volumes. In this analysis, no relevant variations in image quality were observed among the phantoms. The temporal acquisition mode was applied to evaluate the behavior of count rates as a function of activity, analyzing total, true, random, scattered, and noise-equivalent count (NEC). The results demonstrated a progressive reduction in counting efficiency with increasing phantom volume, indicating an inverse relationship between phantom size and detection efficiency. These findings emphasize the importance of considering phantom geometry in calibration and quantitative procedures in preclinical PET imaging studies to ensure accurate measurements.

Authors

Institutions

Publication Details

Journal
Brazilian Journal of Radiation Sciences
Published
2026-11-27
DOI
https://doi.org/10.15392/2319-0612.2026.3058
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Influence of the volume of laboratory-developed simulators on the counting efficiency (kCPS/MBq) of the PET scanner belonging to LIM/CDTN

Rodrigo Modesto Gadelha Gontijo, Andréa Vidal Ferreira, Bruno Melo Mendes, Alexandre Vilaça Silva
Brazilian Journal of Radiation Sciences
Medical Imaging Techniques and Applications
article

Influence of the volume of laboratory-developed simulators on the counting efficiency (kCPS/MBq) of the PET scanner belonging to LIM/CDTN

Rodrigo Modesto Gadelha Gontijo, Andréa Vidal Ferreira, Bruno Melo Mendes, Alexandre Vilaça Silva
article en

Abstract

This study investigated the influence of phantom volume on the conversion coefficient between count rate (CPS) and activity (MBq) for a preclinical Positron Emission Tomography (PET) scanner, the Triumph® LabPET Solo 4, installed at the LIM/CDTN (Laboratório de Imagem Molecular do Centro de Desenvolvimento da Tecnologia Nuclear). Three cylindrical phantoms (S, M, and L) with different internal volumes (17.7 mL, 22.0 mL, and 26.5 mL, respectively) were fabricated. PET images were acquired using two approaches: spatial and temporal. The spatial acquisition mode was used to verify the quality and homogeneity of the reconstructed PET images for the different phantom volumes. In this analysis, no relevant variations in image quality were observed among the phantoms. The temporal acquisition mode was applied to evaluate the behavior of count rates as a function of activity, analyzing total, true, random, scattered, and noise-equivalent count (NEC). The results demonstrated a progressive reduction in counting efficiency with increasing phantom volume, indicating an inverse relationship between phantom size and detection efficiency. These findings emphasize the importance of considering phantom geometry in calibration and quantitative procedures in preclinical PET imaging studies to ensure accurate measurements.

Brazilian Journal of Radiation SciencesVol. 14(4)
Universidade Federal de Minas Gerais (BR), Centro de Desenvolvimento da Tecnologia Nuclear (BR)
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon