Statistical process control and error curve modelling identify superior pre-treatment verification metrics for nasopharyngeal carcinoma and cervical cancer VMAT: a retrospective study
Background and purpose Gamma analysis is widely used for volumetric modulated arc therapy (VMAT) verification, but its clinical relevance and sensitivity to delivery errors remain limited. We integrated statistical process control (SPC) with error curve modelling (ECM) to identify metrics detecting clinically significant errors, derive tolerance limits, and quantify undetected error ranges. Materials and methods This retrospective study included 120 VMAT plans (68 nasopharyngeal carcinoma [NPC], 52 uterine cervical cancer [UCC]). Five mechanical error types were introduced, yielding 1831 datasets. Under three global gamma criteria (3%/2 mm, 2%/2 mm, 1%/1 mm), gamma passing rates (GPRs) and mean gamma index (μGI) were computed in multiple regions. Metrics were screened by Spearman correlation, tolerance limits derived by SPC, undetected ranges quantified by ECM, and discrimination assessed at 3%/2 mm by ROC analysis in an independent patient-level validation set (628 datasets, 55 patients), against a clinical (|DE%PTV| ≥ 3%) and a mechanical (any error) standard. Results GPRswhole, μGIwhole, GPRs90% and μGI90% correlated strongly with PTV dose error (|r| ≥ 0.5). SPC on the development cohort gave 3%/2 mm lower limits for GPRswhole of 0.509 (NPC) and 0.470 (UCC) and upper limits for μGIwhole of 1.016 and 1.087; 97.4–100% of error-free validation plans fell within them. ECM, interpreted only where R² > 0.8, responded to MU and PTV dose errors, whereas random MLC, gantry and most systematic MLC curves were too poorly constrained to yield a detection limit. Against the clinical standard, AUCs were 0.999 (NPC) and 0.874 (UCC) for GPRswhole and 0.997 and 0.841 for μGIwhole, versus 0.605 and 0.810 for GPRs10%; in UCC the 90%-region metrics reached 0.940 and 0.947. At 90% specificity, sensitivities were 0.98–1.00 (NPC) and 0.70–0.83 (UCC) versus 0.37 and 0.58. Under the mechanical standard the advantage was smaller in NPC and absent in UCC. Conclusions Modulation-based metrics on clinically relevant regions identified clinically relevant dose deviations better than GPRs10%, which remained more sensitive to the mere presence of a perturbation; the two are complementary. These limits and undetected error ranges are site-specific and hypothesis-generating, requiring multi-centre validation before clinical acceptance.
Authors
- hanyin zhang
- furong lv
- Wenli Lu
- Qingfeng Jiang
- Haixia Cui
- Wei Huang
- Xin Yi
Institutions
- The Affiliated Yongchuan Hospital of Chongqing Medical University (CN)
- Chongqing Municipal Health Commission (CN)
- Xi'an Jiaotong University (CN)
Publication Details
- Journal
- Frontiers in Oncology
- Published
- 2026-11-27
- DOI
- https://doi.org/10.3389/fonc.2026.1960046
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00