Clinical Validation and Diagnostic Performance of a Multiplex Real-Time RT-PCR Assay for the Detection of SARS-CoV-2 and Influenza A Virus in Clinical Specimens Collected in Kazakhstan in 2021 and 2026

The co-circulation of SARS-CoV-2 and influenza A virus highlights the need for multiplex molecular diagnostics. This study aimed to clinically validate a newly developed multiplex real-time reverse transcription polymerase chain reaction (real-time RT-PCR) assay for their simultaneous detection. A total of 195 nasopharyngeal and oropharyngeal swab specimens collected in Kazakhstan were analyzed, including 68 archived specimens from 2021 and 127 prospectively collected specimens from 2026. Specimen validity was assessed using the endogenous RNase P control. Diagnostic agreement with commercial assays was evaluated using positive percent agreement (PPA), negative percent agreement (NPA), overall percent agreement (OPA), and Cohen’s kappa (κ). Of the 195 specimens, 188 (96.4%) yielded valid results. Influenza A virus and SARS-CoV-2 were detected in 123 (65.4%) and 55 (29.3%) valid specimens, respectively. Borderline results (Ct > 38.0) were excluded separately for each viral target from the binary comparative analysis, resulting in 180 specimens per target. No SARS-CoV-2-positive specimens were identified in the prospective 2026 cohort; therefore, positive percent agreement for this target could not be estimated in this panel. The developed assay demonstrated high agreement with commercial comparator assays, with positive percent agreement values of 98.40% for influenza A virus and 98.21% for SARS-CoV-2. Both viruses were simultaneously detected in seven specimens. The multiplex format enables the simultaneous detection of two clinically important respiratory viruses in a single reaction, potentially reducing specimen and reagent consumption and simplifying laboratory workflows. These findings support the potential application of the developed assay in routine molecular diagnostics of respiratory viral infections.

Authors

Institutions

Publication Details

Journal
Microorganisms
Published
2026-11-26
DOI
https://doi.org/10.3390/microorganisms14122762
Primary Topic
SARS-CoV-2 detection and testing
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Clinical Validation and Diagnostic Performance of a Multiplex Real-Time RT-PCR Assay for the Detection of SARS-CoV-2 and Influenza A Virus in Clinical Specimens Collected in Kazakhstan in 2021 and 2026

Tangat T. Yermekbay, Olga V. Chervyakova, Bekbolat S. Usserbayev, Nurlan Kozhabergenov et al.
Microorganisms
SARS-CoV-2 detection and testing
article

Clinical Validation and Diagnostic Performance of a Multiplex Real-Time RT-PCR Assay for the Detection of SARS-CoV-2 and Influenza A Virus in Clinical Specimens Collected in Kazakhstan in 2021 and 2026

Tangat T. Yermekbay, Olga V. Chervyakova, Bekbolat S. Usserbayev, Nurlan Kozhabergenov, Nurdos A. Aubakir, Kuandyk D. Zhugunissov, Meirzhan Shirinbekov, М.D. Almezhanova, Kulyaisan T. Sultankulova, Yerbol D. Burashev, Assankadyr T. Junushov
article en

Abstract

The co-circulation of SARS-CoV-2 and influenza A virus highlights the need for multiplex molecular diagnostics. This study aimed to clinically validate a newly developed multiplex real-time reverse transcription polymerase chain reaction (real-time RT-PCR) assay for their simultaneous detection. A total of 195 nasopharyngeal and oropharyngeal swab specimens collected in Kazakhstan were analyzed, including 68 archived specimens from 2021 and 127 prospectively collected specimens from 2026. Specimen validity was assessed using the endogenous RNase P control. Diagnostic agreement with commercial assays was evaluated using positive percent agreement (PPA), negative percent agreement (NPA), overall percent agreement (OPA), and Cohen’s kappa (κ). Of the 195 specimens, 188 (96.4%) yielded valid results. Influenza A virus and SARS-CoV-2 were detected in 123 (65.4%) and 55 (29.3%) valid specimens, respectively. Borderline results (Ct > 38.0) were excluded separately for each viral target from the binary comparative analysis, resulting in 180 specimens per target. No SARS-CoV-2-positive specimens were identified in the prospective 2026 cohort; therefore, positive percent agreement for this target could not be estimated in this panel. The developed assay demonstrated high agreement with commercial comparator assays, with positive percent agreement values of 98.40% for influenza A virus and 98.21% for SARS-CoV-2. Both viruses were simultaneously detected in seven specimens. The multiplex format enables the simultaneous detection of two clinically important respiratory viruses in a single reaction, potentially reducing specimen and reagent consumption and simplifying laboratory workflows. These findings support the potential application of the developed assay in routine molecular diagnostics of respiratory viral infections.

MicroorganismsVol. 14(12)
National Academy of Sciences of the Republic of Kyrgyzstan (KG)
SARS-CoV-2 detection and testing
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon