Hydraulic Resistance of the Internal Shunt of a Double-Balloon Catheter for Percutaneous Hepatic Perfusion: An In Vitro Study

Background: Percutaneous hepatic perfusion (PHP) isolates the liver circulation with a double-balloon catheter whose internal shunt maintains venous return from the lower body during balloon occlusion of the inferior vena cava (IVC). Clinically, PHP is associated with hypotension and high catecholamine requirements. This in vitro study quantified the hydraulic resistance of the shunt. Methods: One 16 F double-balloon catheter was tested in a bench model (12.7 mm PVC tubing, open reservoir) using a blood analog (43% glycerol by mass) at 21.4–22.4°C. A centrifugal pump generated flows at 500–4000 RPM with uninflated and inflated balloons (4 mL NaCl); flow and pressure were recorded (n = 5 per step). Results: With uninflated balloons, flow reached 7.55 ± 0.05 L/min at a pressure gradient of 31.6 ± 0.5 mmHg (4000 RPM). After inflation, flow was restricted to the shunt lumen and reduced by 96.9% (500 RPM) to 93.3% (4000 RPM). The maximum shunt flow was 0.51 ± 0.01 L/min at 387 mmHg; at 10 mmHg it was approximately 0.02 L/min. The resistance (ΔP/flow) rose from 493 to 762 mmHg·min/L (p < 0.001). Flow was laminar (Reynolds number < 1800). Conclusion: In this bench model, the internal shunt was a severe hydraulic bottleneck that is unlikely to carry physiological venous return at low venous driving pressures. Occlusion time should be kept short with close hemodynamic monitoring, and shunt design optimization should be investigated. The results are limited to one catheter in a rigid-tube model.

Authors

Publication Details

Journal
Journal of ExtraCorporeal Technology
Published
2026-11-27
DOI
https://doi.org/10.1051/ject/2026054
Primary Topic
Hepatocellular Carcinoma Treatment and Prognosis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Hydraulic Resistance of the Internal Shunt of a Double-Balloon Catheter for Percutaneous Hepatic Perfusion: An In Vitro Study

Jan Turra, Folker Wenzel, Philippe Grieshaber, Patrick Rehn et al.
Journal of ExtraCorporeal Technology
Hepatocellular Carcinoma Treatment and Prognosis
article

Hydraulic Resistance of the Internal Shunt of a Double-Balloon Catheter for Percutaneous Hepatic Perfusion: An In Vitro Study

Jan Turra, Folker Wenzel, Philippe Grieshaber, Patrick Rehn, Maik von der Forst, Max Dietrich, Oliver Biecker, Dania Fischer, Osman Oecal, Jessica Hassel
article en

Abstract

Background: Percutaneous hepatic perfusion (PHP) isolates the liver circulation with a double-balloon catheter whose internal shunt maintains venous return from the lower body during balloon occlusion of the inferior vena cava (IVC). Clinically, PHP is associated with hypotension and high catecholamine requirements. This in vitro study quantified the hydraulic resistance of the shunt. Methods: One 16 F double-balloon catheter was tested in a bench model (12.7 mm PVC tubing, open reservoir) using a blood analog (43% glycerol by mass) at 21.4–22.4°C. A centrifugal pump generated flows at 500–4000 RPM with uninflated and inflated balloons (4 mL NaCl); flow and pressure were recorded (n = 5 per step). Results: With uninflated balloons, flow reached 7.55 ± 0.05 L/min at a pressure gradient of 31.6 ± 0.5 mmHg (4000 RPM). After inflation, flow was restricted to the shunt lumen and reduced by 96.9% (500 RPM) to 93.3% (4000 RPM). The maximum shunt flow was 0.51 ± 0.01 L/min at 387 mmHg; at 10 mmHg it was approximately 0.02 L/min. The resistance (ΔP/flow) rose from 493 to 762 mmHg·min/L (p < 0.001). Flow was laminar (Reynolds number < 1800). Conclusion: In this bench model, the internal shunt was a severe hydraulic bottleneck that is unlikely to carry physiological venous return at low venous driving pressures. Occlusion time should be kept short with close hemodynamic monitoring, and shunt design optimization should be investigated. The results are limited to one catheter in a rigid-tube model.

Journal of ExtraCorporeal Technology
Hepatocellular Carcinoma Treatment and Prognosis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Hydraulic Resistance of the Internal Shunt of a Double-Balloon Catheter for Percutaneous Hepatic Perfusion: An In Vitro Study — Jan Turra, Folker Wenzel, et al. · Journal of ExtraCorporeal Technology (2026) | TGRS Research Map | TGRS