A Compact Dual-Axis Homodyne Grating Interferometer Using a Wollaston-Prism-Based Quadrant Detection Module
A compact dual-axis homodyne grating interferometer with an integrated detection module based on Wollaston prisms and quadrant photodetectors is developed for displacement measurement in space-constrained precision motion systems. Rotating the upstream non-polarizing beam splitter by 45° about the incident optical axis adjusts the spatial arrangement of two parallel beams so that the four spots formed by each Wollaston prism are centered on the four photosensitive quadrants of the corresponding detector. This arrangement adapts the integrated detection module to dual-axis grating interferometry. The prototype measures 49.2 mm × 72.5 mm × 35 mm. In the separate-axis measurement experiments using two phasemeters, the 3σ values of the displacement fluctuations along the x- and z-axes were 6.06 nm and 8.10 nm, respectively, over 10 s stationary windows.
Authors
- Di Chang (ORCID: https://orcid.org/0000-0002-2124-8043)
- Dunbo Song (ORCID: https://orcid.org/0009-0007-1840-4547)
- Jiubin Tan
Institutions
- Harbin Institute of Technology (CN)
- Ministry of Industry and Information Technology (CN)
Publication Details
- Journal
- Photonics
- Published
- 2026-11-26
- DOI
- https://doi.org/10.3390/photonics13121117
- Primary Topic
- Advanced Measurement and Metrology Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00