Evaluation of Dose Calibrator (CRC-55t) Linearity Test Using Different Photon Energies (F-18, Tc-99m and I-131)

This present study investigates and compares the linearity performance for three isotopes: technetium-99m (140 keV γ, T½ = 6.01 h), fluorine-18 (511 keV, positron emitter, T½ = 109.8 min), and iodine-131 (364 keV γ + β emitter, T½ = 8.02 days) whose activities were measured up to values lower than 1 MBq using three different windows. The mean ratio between activities indicated by the dose calibrator (DC) for the 99mTc source as measured with the equipment precalibrated to measure 99mTc, iodine-131, and 18F was 1.431 ± 0.004, 1.431 ± 0.003,and 1.434 ± 0.007, respectively, values considered constant over the measurement time. The mean ratio of the Tc-99m window relative to the F-18 window remained remarkably stable across all experiments, for Tc-99m this ratio was 3.257 ± 0.0093, for iodine-131 source this ratio was 3.247 ± 0.007, and for the 18F source this ratio was 3.247 ± 0.019. The DC tested was linear across different activity range using different isotopes Tc-99m, F-18, and I-131. Tc-99m remains suitable for routine linearity checks, while inclusion of F-18 and I-131 strengthens validation across photon energies. These results support the clinical reliability of the device.

Authors

Institutions

Publication Details

Journal
World Journal of Nuclear Medicine
Published
2026-11-26
DOI
https://doi.org/10.1055/s-0046-1830099
Primary Topic
Radioactive Decay and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Evaluation of Dose Calibrator (CRC-55t) Linearity Test Using Different Photon Energies (F-18, Tc-99m and I-131)

Naima Khamis Al-Bulushi, Ayman Qawark, Khalsa Al-Nabhani, Fatema AlGhafri
World Journal of Nuclear Medicine
Radioactive Decay and Measurement Techniques
article

Evaluation of Dose Calibrator (CRC-55t) Linearity Test Using Different Photon Energies (F-18, Tc-99m and I-131)

Naima Khamis Al-Bulushi, Ayman Qawark, Khalsa Al-Nabhani, Fatema AlGhafri
article en

Abstract

This present study investigates and compares the linearity performance for three isotopes: technetium-99m (140 keV γ, T½ = 6.01 h), fluorine-18 (511 keV, positron emitter, T½ = 109.8 min), and iodine-131 (364 keV γ + β emitter, T½ = 8.02 days) whose activities were measured up to values lower than 1 MBq using three different windows. The mean ratio between activities indicated by the dose calibrator (DC) for the 99mTc source as measured with the equipment precalibrated to measure 99mTc, iodine-131, and 18F was 1.431 ± 0.004, 1.431 ± 0.003,and 1.434 ± 0.007, respectively, values considered constant over the measurement time. The mean ratio of the Tc-99m window relative to the F-18 window remained remarkably stable across all experiments, for Tc-99m this ratio was 3.257 ± 0.0093, for iodine-131 source this ratio was 3.247 ± 0.007, and for the 18F source this ratio was 3.247 ± 0.019. The DC tested was linear across different activity range using different isotopes Tc-99m, F-18, and I-131. Tc-99m remains suitable for routine linearity checks, while inclusion of F-18 and I-131 strengthens validation across photon energies. These results support the clinical reliability of the device.

World Journal of Nuclear Medicine
Royal Hospital (OM)
Radioactive Decay and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon