Evaluation of Dose Calibrator (CRC-55t) Linearity Test Using Different Photon Energies (F-18, Tc-99m and I-131)
This present study investigates and compares the linearity performance for three isotopes: technetium-99m (140 keV γ, T½ = 6.01 h), fluorine-18 (511 keV, positron emitter, T½ = 109.8 min), and iodine-131 (364 keV γ + β emitter, T½ = 8.02 days) whose activities were measured up to values lower than 1 MBq using three different windows. The mean ratio between activities indicated by the dose calibrator (DC) for the 99mTc source as measured with the equipment precalibrated to measure 99mTc, iodine-131, and 18F was 1.431 ± 0.004, 1.431 ± 0.003,and 1.434 ± 0.007, respectively, values considered constant over the measurement time. The mean ratio of the Tc-99m window relative to the F-18 window remained remarkably stable across all experiments, for Tc-99m this ratio was 3.257 ± 0.0093, for iodine-131 source this ratio was 3.247 ± 0.007, and for the 18F source this ratio was 3.247 ± 0.019. The DC tested was linear across different activity range using different isotopes Tc-99m, F-18, and I-131. Tc-99m remains suitable for routine linearity checks, while inclusion of F-18 and I-131 strengthens validation across photon energies. These results support the clinical reliability of the device.
Authors
- Naima Khamis Al-Bulushi (ORCID: https://orcid.org/0000-0002-9819-651X)
- Ayman Qawark
- Khalsa Al-Nabhani
- Fatema AlGhafri
Institutions
- Royal Hospital (OM)
Publication Details
- Journal
- World Journal of Nuclear Medicine
- Published
- 2026-11-26
- DOI
- https://doi.org/10.1055/s-0046-1830099
- Primary Topic
- Radioactive Decay and Measurement Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00