Environmental determinants of sterility maintenance in single-layer non-woven wrapped instrument packs stored on clinical units: A four-week prospective study

Aim/objective To determine whether instruments in packs wrapped in a single layer of SMMS non-woven fabric remain sterile for up to 4 weeks in clinical-unit cabinets, and whether cabinet temperature and humidity are associated with contamination. Methods and results In a prospective observational study at a specialist cancer hospital in northern Thailand, 1,056 steam-sterilized packs containing forceps were stored in closed cabinets in seven clinical areas and tested after 1, 2, 3, or 4 weeks. Forceps were swabbed and cultured by two-step enrichment; cabinet temperature and humidity were recorded twice each working day. Growth occurred in 15 packs (1.42%), with no significant difference by storage duration (p = 0.327) or location (p = 0.134). In cluster-robust logistic regression, higher temperature (adjusted OR 1.26 per °C, 95% CI 1.07–1.49) and humidity (adjusted OR 1.08 per percentage point, 95% CI 1.00–1.16) were associated with contamination. All isolates were skin or environmental commensals, mainly coagulase-negative staphylococci. Discussion The environmental association is hypothesis-generating: the enrichment method detects very low bioburden, the organisms are typical procedural contaminants, and no swab-handling controls were included. Significance and impact A four-week shelf life is supportable for single-layer SMMS-wrapped packs in closed clinical-unit cabinets under an event-related approach. Storage areas markedly above recommended temperature, such as this study’s intensive care unit, need environmental remediation or a shorter local limit.

Authors

Institutions

Publication Details

Journal
Journal of Infection Prevention
Published
2026-11-26
DOI
https://doi.org/10.1177/17571774261513010
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Environmental determinants of sterility maintenance in single-layer non-woven wrapped instrument packs stored on clinical units: A four-week prospective study

Piyanut Xuto, Lawitra Khiaokham, Nida Matejitkul, Pipat Na-ngam et al.
Journal of Infection Prevention
Medical Device Sterilization and Disinfection
article

Environmental determinants of sterility maintenance in single-layer non-woven wrapped instrument packs stored on clinical units: A four-week prospective study

Piyanut Xuto, Lawitra Khiaokham, Nida Matejitkul, Pipat Na-ngam, Jatuporn Chaiyanon, Patompong Treeboonmueang, Wipada Sukarunrote, Ammara Pawong
article en

Abstract

Aim/objective To determine whether instruments in packs wrapped in a single layer of SMMS non-woven fabric remain sterile for up to 4 weeks in clinical-unit cabinets, and whether cabinet temperature and humidity are associated with contamination. Methods and results In a prospective observational study at a specialist cancer hospital in northern Thailand, 1,056 steam-sterilized packs containing forceps were stored in closed cabinets in seven clinical areas and tested after 1, 2, 3, or 4 weeks. Forceps were swabbed and cultured by two-step enrichment; cabinet temperature and humidity were recorded twice each working day. Growth occurred in 15 packs (1.42%), with no significant difference by storage duration (p = 0.327) or location (p = 0.134). In cluster-robust logistic regression, higher temperature (adjusted OR 1.26 per °C, 95% CI 1.07–1.49) and humidity (adjusted OR 1.08 per percentage point, 95% CI 1.00–1.16) were associated with contamination. All isolates were skin or environmental commensals, mainly coagulase-negative staphylococci. Discussion The environmental association is hypothesis-generating: the enrichment method detects very low bioburden, the organisms are typical procedural contaminants, and no swab-handling controls were included. Significance and impact A four-week shelf life is supportable for single-layer SMMS-wrapped packs in closed clinical-unit cabinets under an event-related approach. Storage areas markedly above recommended temperature, such as this study’s intensive care unit, need environmental remediation or a shorter local limit.

Journal of Infection Prevention
Chiang Mai University (TH)
Medical Device Sterilization and Disinfection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon