Environmental determinants of sterility maintenance in single-layer non-woven wrapped instrument packs stored on clinical units: A four-week prospective study
Aim/objective To determine whether instruments in packs wrapped in a single layer of SMMS non-woven fabric remain sterile for up to 4 weeks in clinical-unit cabinets, and whether cabinet temperature and humidity are associated with contamination. Methods and results In a prospective observational study at a specialist cancer hospital in northern Thailand, 1,056 steam-sterilized packs containing forceps were stored in closed cabinets in seven clinical areas and tested after 1, 2, 3, or 4 weeks. Forceps were swabbed and cultured by two-step enrichment; cabinet temperature and humidity were recorded twice each working day. Growth occurred in 15 packs (1.42%), with no significant difference by storage duration (p = 0.327) or location (p = 0.134). In cluster-robust logistic regression, higher temperature (adjusted OR 1.26 per °C, 95% CI 1.07–1.49) and humidity (adjusted OR 1.08 per percentage point, 95% CI 1.00–1.16) were associated with contamination. All isolates were skin or environmental commensals, mainly coagulase-negative staphylococci. Discussion The environmental association is hypothesis-generating: the enrichment method detects very low bioburden, the organisms are typical procedural contaminants, and no swab-handling controls were included. Significance and impact A four-week shelf life is supportable for single-layer SMMS-wrapped packs in closed clinical-unit cabinets under an event-related approach. Storage areas markedly above recommended temperature, such as this study’s intensive care unit, need environmental remediation or a shorter local limit.
Authors
- Piyanut Xuto (ORCID: https://orcid.org/0000-0003-1338-0789)
- Lawitra Khiaokham (ORCID: https://orcid.org/0000-0002-4008-5446)
- Nida Matejitkul
- Pipat Na-ngam
- Jatuporn Chaiyanon
- Patompong Treeboonmueang
- Wipada Sukarunrote
- Ammara Pawong
Institutions
- Chiang Mai University (TH)
Publication Details
- Journal
- Journal of Infection Prevention
- Published
- 2026-11-26
- DOI
- https://doi.org/10.1177/17571774261513010
- Primary Topic
- Medical Device Sterilization and Disinfection
- Type
- article
- Field-Weighted Citation Impact
- 0.00