Quantitative and task‐based performance of spectral deep learning reconstruction in low‐dose dual‐layer coronary CT angiography: A phantom study

Background Coronary CT angiography (CCTA) is an established noninvasive test for coronary artery disease, but balancing image quality and radiation dose remains challenging. Dual‐layer spectral CT (DLCT) enables reconstruction of virtual monoenergetic images (VMI) that enhance coronary artery contrast, but are inherently more sensitive to noise, making this trade‐off especially critical. Deep learning reconstruction (DLR) improves image quality and enables radiation dose reduction in CCTA, but its performance on spectral reconstructions from dual‐layer CT (DLCT) remains unexplored. Purpose This phantom study compares spectral DLR at five different noise reduction strengths with hybrid‐iterative reconstruction (hybrid‐IR) at one fixed, clinically used noise reduction strength across conventional and VMI at multiple reduced dose levels. Methods An anthropomorphic thorax phantom with a coronary insert containing two calcifications (100 and 800 mg CaHA/mL) was scanned on a DLCT system at six dose levels (volumetric CT dose index 12.2–1.4 mGy) using a clinical sequential CCTA protocol. Data were reconstructed with one clinical hybrid‐IR protocol, and DLR at five noise reduction strengths (Smoother, Smooth, Standard, Sharp, Sharper) for conventional images and VMI at 7 energy levels (40–70 keV). Image noise, luminal contrast‐to‐noise ratio (CNR lumen), noise power spectrum means frequency (NPS mean), task transfer function at 50% (TTF 50), and detectability index (d') of coronary plaques were assessed. Statistical significance was inferred when 95% confidence intervals did not overlap. Results At full dose, DLR significantly reduced noise up to 64% in conventional images and 67% in 40 keV VMI, while increasing CNR lumen up to 300% and 230%, respectively, compared with hybrid‐IR. DLR Standard, Sharp, and Sharper significantly increased NPS mean by 7%–13% for conventional images and 40 keV VMI, whereas DLR Smoother shifted noise toward lower frequencies, reducing NPS mean by 22%–32%. All DLR strengths significantly improved spatial resolution of calcified plaque, with TTF 50 increasing 9%–29% relative to hybrid‐IR. VMI at 40 keV with DLR significantly increased d' by 307% for high‐density and 283% for low‐density calcifications compared to conventional hybrid‐IR. At 50% dose reduction, VMI at 40 keV with DLR Smooth improved CNR lumen (+ 46%), and TTF 50 (+ 18%) and d’ (+ 24%) of the high‐density calcification relative to full‐dose hybrid‐IR 40 keV VMI, whereas confidence intervals overlapped for NPS mean and for TTF 50 and d’ of the low‐density calcification. Conclusions On dual‐layer CT, spectral DLR outperforms hybrid‐IR on both conventional and VMI images in noise, CNR, spatial resolution, and plaque detectability. At 50% dose reduction, DLR Smooth on 40 keV VMI exceeded full‐dose hybrid‐IR on 40 keV VMI in CNR, and in spatial resolution and detectability of high‐density calcifications, with overlapping confidence intervals for noise texture and for low‐density calcifications. Patient studies are needed to validate clinical translation.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2026-11-25
DOI
https://doi.org/10.1002/mp.70778
Primary Topic
Cardiac Imaging and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Quantitative and task‐based performance of spectral deep learning reconstruction in low‐dose dual‐layer coronary CT angiography: A phantom study

Arnold M. R. Schilham, Marcel J. W. Greuter, Dominika Suchá, Stefan Johannes Fransen et al.
Medical Physics
Cardiac Imaging and Diagnostics
article

Quantitative and task‐based performance of spectral deep learning reconstruction in low‐dose dual‐layer coronary CT angiography: A phantom study

Arnold M. R. Schilham, Marcel J. W. Greuter, Dominika Suchá, Stefan Johannes Fransen, Eran Langzam, Niels R. van der Werf, Lennart R. Koetzier, Birgitta K. Velthuis
article en

Abstract

Background Coronary CT angiography (CCTA) is an established noninvasive test for coronary artery disease, but balancing image quality and radiation dose remains challenging. Dual‐layer spectral CT (DLCT) enables reconstruction of virtual monoenergetic images (VMI) that enhance coronary artery contrast, but are inherently more sensitive to noise, making this trade‐off especially critical. Deep learning reconstruction (DLR) improves image quality and enables radiation dose reduction in CCTA, but its performance on spectral reconstructions from dual‐layer CT (DLCT) remains unexplored. Purpose This phantom study compares spectral DLR at five different noise reduction strengths with hybrid‐iterative reconstruction (hybrid‐IR) at one fixed, clinically used noise reduction strength across conventional and VMI at multiple reduced dose levels. Methods An anthropomorphic thorax phantom with a coronary insert containing two calcifications (100 and 800 mg CaHA/mL) was scanned on a DLCT system at six dose levels (volumetric CT dose index 12.2–1.4 mGy) using a clinical sequential CCTA protocol. Data were reconstructed with one clinical hybrid‐IR protocol, and DLR at five noise reduction strengths (Smoother, Smooth, Standard, Sharp, Sharper) for conventional images and VMI at 7 energy levels (40–70 keV). Image noise, luminal contrast‐to‐noise ratio (CNR lumen), noise power spectrum means frequency (NPS mean), task transfer function at 50% (TTF 50), and detectability index (d') of coronary plaques were assessed. Statistical significance was inferred when 95% confidence intervals did not overlap. Results At full dose, DLR significantly reduced noise up to 64% in conventional images and 67% in 40 keV VMI, while increasing CNR lumen up to 300% and 230%, respectively, compared with hybrid‐IR. DLR Standard, Sharp, and Sharper significantly increased NPS mean by 7%–13% for conventional images and 40 keV VMI, whereas DLR Smoother shifted noise toward lower frequencies, reducing NPS mean by 22%–32%. All DLR strengths significantly improved spatial resolution of calcified plaque, with TTF 50 increasing 9%–29% relative to hybrid‐IR. VMI at 40 keV with DLR significantly increased d' by 307% for high‐density and 283% for low‐density calcifications compared to conventional hybrid‐IR. At 50% dose reduction, VMI at 40 keV with DLR Smooth improved CNR lumen (+ 46%), and TTF 50 (+ 18%) and d’ (+ 24%) of the high‐density calcification relative to full‐dose hybrid‐IR 40 keV VMI, whereas confidence intervals overlapped for NPS mean and for TTF 50 and d’ of the low‐density calcification. Conclusions On dual‐layer CT, spectral DLR outperforms hybrid‐IR on both conventional and VMI images in noise, CNR, spatial resolution, and plaque detectability. At 50% dose reduction, DLR Smooth on 40 keV VMI exceeded full‐dose hybrid‐IR on 40 keV VMI in CNR, and in spatial resolution and detectability of high‐density calcifications, with overlapping confidence intervals for noise texture and for low‐density calcifications. Patient studies are needed to validate clinical translation.

Medical PhysicsVol. 53(12)
Philips (Finland) (FI), University Medical Center Groningen (NL), University of Groningen (NL), University Medical Center Utrecht (NL), Philips (Netherlands) (NL)
Cardiac Imaging and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon