Repeatability, Reproducibility and Inter-Device Agreement of Anterior and Posterior Corneal Parameters Measured with Visionix 130+ and Pentacam HR in Healthy Eyes

Assessing the validity and measurement performance of optical instrumentation is essential to ensure its clinical utility. Therefore, the study’s aims were to analyze the repeatability and reproducibility of two multi-diagnostic platforms (Visionix 130+ and Pentacam HR) in the measurement of corneal characteristics, and to assess inter-device agreement. Anterior and posterior corneal curvature, eccentricity and central corneal thickness were measured using both devices in 31 healthy subjects. Each participant was subjected to six measurements per device, divided equally between two observers (three per observer). Both devices showed moderate-to-high reliability for anterior parameters and central corneal thickness (ICC > 79.89). For posterior parameters, Pentacam HR maintained moderate-to-high reliability (ICC: 78.28–99.34), whereas Visionix 130+ revealed decreased reliability (ICC: 5.16–77.77). Regarding eccentricity, high reliability was observed for Pentacam HR (ICC: 96.55), while Visionix 130+ exhibited low reliability (ICC: 68.81). No agreement was found between the results provided by the two devices for most of the parameters assessed. It is concluded that both devices demonstrate good repeatability and reproducibility when assessing anterior corneal surface and central corneal thickness, with the exception of eccentricity measured by Visionix 130+. For posterior surface measurements, Pentacam HR demonstrates moderate-to-high reliability, while Visionix 130+ exhibits low reliability for most parameters. Finally, the results of both devices are not interchangeable.

Authors

Institutions

Publication Details

Journal
Photonics
Published
2026-11-26
DOI
https://doi.org/10.3390/photonics13121120
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Repeatability, Reproducibility and Inter-Device Agreement of Anterior and Posterior Corneal Parameters Measured with Visionix 130+ and Pentacam HR in Healthy Eyes

Maria Jesus Giraldez, Eva Yebra‐Pimentel, Verónica Noya-Padin, César Sánchez-Sellero et al.
Photonics
Corneal surgery and disorders
article

Repeatability, Reproducibility and Inter-Device Agreement of Anterior and Posterior Corneal Parameters Measured with Visionix 130+ and Pentacam HR in Healthy Eyes

Maria Jesus Giraldez, Eva Yebra‐Pimentel, Verónica Noya-Padin, César Sánchez-Sellero, Juan F. Otero-Mouteira, Lua Souto-Blanco
article en

Abstract

Assessing the validity and measurement performance of optical instrumentation is essential to ensure its clinical utility. Therefore, the study’s aims were to analyze the repeatability and reproducibility of two multi-diagnostic platforms (Visionix 130+ and Pentacam HR) in the measurement of corneal characteristics, and to assess inter-device agreement. Anterior and posterior corneal curvature, eccentricity and central corneal thickness were measured using both devices in 31 healthy subjects. Each participant was subjected to six measurements per device, divided equally between two observers (three per observer). Both devices showed moderate-to-high reliability for anterior parameters and central corneal thickness (ICC > 79.89). For posterior parameters, Pentacam HR maintained moderate-to-high reliability (ICC: 78.28–99.34), whereas Visionix 130+ revealed decreased reliability (ICC: 5.16–77.77). Regarding eccentricity, high reliability was observed for Pentacam HR (ICC: 96.55), while Visionix 130+ exhibited low reliability (ICC: 68.81). No agreement was found between the results provided by the two devices for most of the parameters assessed. It is concluded that both devices demonstrate good repeatability and reproducibility when assessing anterior corneal surface and central corneal thickness, with the exception of eccentricity measured by Visionix 130+. For posterior surface measurements, Pentacam HR demonstrates moderate-to-high reliability, while Visionix 130+ exhibits low reliability for most parameters. Finally, the results of both devices are not interchangeable.

PhotonicsVol. 13(12)
Universidade de Santiago de Compostela (ES), Instituto de Investigación Sanitaria de Santiago (ES)
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon