Method-dependent drug interference in measuring serum creatinine in a deeply premature newborn: a clinical case and literature review

Background. Drug interference is an underappreciated cause of spurious laboratory test results, documented in 43% of hospitalised patients. In very preterm neonates receiving multidrug infusion therapy, the risk of diagnostic errors is particularly high due to low absolute analyte concentrations, age-dependent reference intervals, and a 30—50% incidence of acute kidney injury. Objective. To present a clinical case of method-dependent drug interference in serum creatinine measurement in such a patient and to propose a laboratory algorithm for artefact identification based on a literature review. Material and methods. Serum samples from a paediatric patient were studied. Two closed analytical systems (fully traceable to IDMS) were compared: the enzymatic method and the colorimetric Jaffe method. A preterm neonate (gestational age 28 weeks, birth weight <1000 g) received continuous infusions of ethamsylate and dobutamine. Results. Parallel creatinine measurements showed striking discrepancies: the colorimetric Jaffe method gave values of 206—236 µmol/L, whereas the enzymatic method (Trinder reaction) yielded only 76—120 µmol/L, which was inconsistent with the clinical picture. Reaction curve analysis and a serum dilution test (1:4) confirmed suppression of the enzymatic reaction caused by the reducing properties of ethamsylate and dobutamine, which compete with peroxidase for hydrogen peroxide. Conclusion. In very preterm neonates receiving ethamsylate and dobutamine infusions, enzymatic creatinine determination yields persistently falsely low values, whereas the Jaffe method remains diagnostically reliable. A serum dilution test can identify the artefact, and preanalytical prevention minimises the risk of errors.

Authors

Institutions

Publication Details

Journal
Laboratory Service
Published
2026-11-26
DOI
https://doi.org/10.17116/labs20261503139
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Method-dependent drug interference in measuring serum creatinine in a deeply premature newborn: a clinical case and literature review

Vladimir Leonidovich Emanuel, L. A. Kornoukhova, Ю. А. Чурюмова, Andrey M. Ivanov et al.
Laboratory Service
Clinical Laboratory Practices and Quality Control
article

Method-dependent drug interference in measuring serum creatinine in a deeply premature newborn: a clinical case and literature review

Vladimir Leonidovich Emanuel, L. A. Kornoukhova, Ю. А. Чурюмова, Andrey M. Ivanov, R. R. Ismagilov, R.F. Salikhova
article en

Abstract

Background. Drug interference is an underappreciated cause of spurious laboratory test results, documented in 43% of hospitalised patients. In very preterm neonates receiving multidrug infusion therapy, the risk of diagnostic errors is particularly high due to low absolute analyte concentrations, age-dependent reference intervals, and a 30—50% incidence of acute kidney injury. Objective. To present a clinical case of method-dependent drug interference in serum creatinine measurement in such a patient and to propose a laboratory algorithm for artefact identification based on a literature review. Material and methods. Serum samples from a paediatric patient were studied. Two closed analytical systems (fully traceable to IDMS) were compared: the enzymatic method and the colorimetric Jaffe method. A preterm neonate (gestational age 28 weeks, birth weight <1000 g) received continuous infusions of ethamsylate and dobutamine. Results. Parallel creatinine measurements showed striking discrepancies: the colorimetric Jaffe method gave values of 206—236 µmol/L, whereas the enzymatic method (Trinder reaction) yielded only 76—120 µmol/L, which was inconsistent with the clinical picture. Reaction curve analysis and a serum dilution test (1:4) confirmed suppression of the enzymatic reaction caused by the reducing properties of ethamsylate and dobutamine, which compete with peroxidase for hydrogen peroxide. Conclusion. In very preterm neonates receiving ethamsylate and dobutamine infusions, enzymatic creatinine determination yields persistently falsely low values, whereas the Jaffe method remains diagnostically reliable. A serum dilution test can identify the artefact, and preanalytical prevention minimises the risk of errors.

Laboratory ServiceVol. 15(3)
First Pavlov State Medical University of St. Petersburg (RU), S. M. Kirov Military Medical Academy (RU), Leningrad Regional Cancer Center (RU), Republican Clinical Hospital named after G.G. Kuvatova (RU)
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon