Optimization of a Locally Shielded 16-Channel Transmit Array for Human Brain MRI at 10.5 T via Electromagnetic/Circuit Co-Simulation and Parallel Transmission Strategies

Modern neuroimaging and neuroscience increasingly demand high spatial and temporal resolution, driving the use of ultrahigh-field (7 T and above) MRI for improved signal-to-noise ratio. Realizing this advantage depends critically on advanced radiofrequency (RF) and gradient technologies, particularly multi-channel transmit arrays and high-performance head-gradient systems. In integrated systems, a local RF shield must provide sufficient field confinement while preserving transmit performance and minimizing sensitivity to nearby conductive structures. In this work, we sought to optimize a locally shielded 16-channel transmit RF array for human neuroimaging at 10.5 T using electromagnetic/circuit co-simulation combined with parallel-transmission strategies. Candidate transmit-array and RF-shield configurations were evaluated before and after gradient-shield integration to identify a configuration with suitable RF transmit and shielding performance for the integrated system. The results show that no single standalone metric, including static RF shimming performance, dynamic parallel transmission (pTx) performance, or external-field leakage, reliably predicts robustness after gradient-shield integration. Progressive screening identified the aligned loop-folded dipole (LFD) array combined with a compact cylindrical shield featuring a conical top and three windows (160-mm radius and 70-mm window length) as the final configuration. Without RF-circuit retuning, the selected design maintained external-field leakage below 1% and after gradient shield integration achieved whole-brain excitation coefficient of variation (CoV) of approximately 27% with static RF shimming and 9% with dynamic pTx. Experimental validation of the unshielded LFD array showed good agreement between simulated and measured B1+ efficiency distributions, with a normalized root-mean-square error (NRMSE) of approximately 8%.

Authors

Publication Details

Journal
arXiv
Published
2026-11-25
DOI
https://doi.org/10.48550/arxiv.2611.34395
Primary Topic
Advanced MRI Techniques and Applications
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Optimization of a Locally Shielded 16-Channel Transmit Array for Human Brain MRI at 10.5 T via Electromagnetic/Circuit Co-Simulation and Parallel Transmission Strategies

Alireza Sadeghi‐Tarakameh, Mengxuan Zheng, Mazin M. Mustafa, Xiaoping Wu et al.
arXiv
Advanced MRI Techniques and Applications
preprint

Optimization of a Locally Shielded 16-Channel Transmit Array for Human Brain MRI at 10.5 T via Electromagnetic/Circuit Co-Simulation and Parallel Transmission Strategies

Alireza Sadeghi‐Tarakameh, Mengxuan Zheng, Mazin M. Mustafa, Xiaoping Wu, Kamil Ugurbil, Xin Shao, Xiaotong Zhang, Matt Waks
preprint en

Abstract

Modern neuroimaging and neuroscience increasingly demand high spatial and temporal resolution, driving the use of ultrahigh-field (7 T and above) MRI for improved signal-to-noise ratio. Realizing this advantage depends critically on advanced radiofrequency (RF) and gradient technologies, particularly multi-channel transmit arrays and high-performance head-gradient systems. In integrated systems, a local RF shield must provide sufficient field confinement while preserving transmit performance and minimizing sensitivity to nearby conductive structures. In this work, we sought to optimize a locally shielded 16-channel transmit RF array for human neuroimaging at 10.5 T using electromagnetic/circuit co-simulation combined with parallel-transmission strategies. Candidate transmit-array and RF-shield configurations were evaluated before and after gradient-shield integration to identify a configuration with suitable RF transmit and shielding performance for the integrated system. The results show that no single standalone metric, including static RF shimming performance, dynamic parallel transmission (pTx) performance, or external-field leakage, reliably predicts robustness after gradient-shield integration. Progressive screening identified the aligned loop-folded dipole (LFD) array combined with a compact cylindrical shield featuring a conical top and three windows (160-mm radius and 70-mm window length) as the final configuration. Without RF-circuit retuning, the selected design maintained external-field leakage below 1% and after gradient shield integration achieved whole-brain excitation coefficient of variation (CoV) of approximately 27% with static RF shimming and 9% with dynamic pTx. Experimental validation of the unshielded LFD array showed good agreement between simulated and measured B1+ efficiency distributions, with a normalized root-mean-square error (NRMSE) of approximately 8%.

arXiv
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon