Detector-side neural timing calibration for TOF-PET from coincidence supervision
Objective: Machine-learning methods can model non-linear event-dependent timing effects in time-of-flight positron emission tomography (TOF-PET), but calibration measurements naturally provide timing supervision only at the coincidence-level rather than for individual detector interactions. We investigate whether coincidence measurements can be used to learn reusable detector-side timing calibration functions without individually labelled interaction times. Approach: A shared detector-conditioned neural network independently processes each single-event readout and predicts a detector-side timing correction and variance contribution. During training, two detector-side timing estimates are differenced and compared with the geometrically expected time difference obtained from point-source coincidence measurements. We evaluate the method on three heterogeneous PET systems against conventional time-skew and time-walk calibration and matched direct coincidence-level neural models. Main results: Under a common energy-window setting, the proposed model reduced coincidence timing resolution from 321.7 to 263.0ps, 462.4 to 431.7ps, and 417.0 to 382.0ps relative to charge-based time-walk correction on the three systems. In detector-pair holdout experiments, detector-side models achieved comparable or slightly better timing performance than direct coincidence-level neural models while retaining independently applicable singles-level outputs. The learned calibration generalized to unseen source positions and detector pairs, enabled recombination of independently trained scanner components after scalar timing-reference alignment, provided event-wise coincidence-timing uncertainty stratification in both full-ring systems, and remained compatible with low-precision inference after quantization-aware training. Significance: Coincidence supervision enables reusable detector-side timing corrections that improve CTR without single-event timing labels. This preserves a singles-level calibration interface while enabling machine-learning-based nonlinear timing correction, modular scanner calibration, and event-specific coincidence-timing uncertainty estimates.
Authors
- Yannick Kuhl (ORCID: https://orcid.org/0000-0002-4548-0111)
- Bjoern Weissler (ORCID: https://orcid.org/0000-0003-1119-785X)
- S. Gundacker (ORCID: https://orcid.org/0000-0003-2087-3266)
- Vanessa Nadig (ORCID: https://orcid.org/0000-0002-1566-0568)
- Stephan Naunheim (ORCID: https://orcid.org/0000-0003-0306-7641)
- Volkmar Schulz (ORCID: https://orcid.org/0000-0003-1341-9356)
- David Schug (ORCID: https://orcid.org/0000-0002-5154-8303)
- Julian Thull
- Marc von Henning
- Florian Mueller (ORCID: https://orcid.org/0000-0002-9496-4710)
- L. Lopes de Paiva
- Thore Meyer (ORCID: https://orcid.org/0009-0004-8632-9647)
Institutions
- RWTH Aachen University (DE)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2026-11-26
- DOI
- https://doi.org/10.1088/1361-6560/aec75b
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00