Detector-side neural timing calibration for TOF-PET from coincidence supervision

Objective: Machine-learning methods can model non-linear event-dependent timing effects in time-of-flight positron emission tomography (TOF-PET), but calibration measurements naturally provide timing supervision only at the coincidence-level rather than for individual detector interactions. We investigate whether coincidence measurements can be used to learn reusable detector-side timing calibration functions without individually labelled interaction times. Approach: A shared detector-conditioned neural network independently processes each single-event readout and predicts a detector-side timing correction and variance contribution. During training, two detector-side timing estimates are differenced and compared with the geometrically expected time difference obtained from point-source coincidence measurements. We evaluate the method on three heterogeneous PET systems against conventional time-skew and time-walk calibration and matched direct coincidence-level neural models. Main results: Under a common energy-window setting, the proposed model reduced coincidence timing resolution from 321.7 to 263.0ps, 462.4 to 431.7ps, and 417.0 to 382.0ps relative to charge-based time-walk correction on the three systems. In detector-pair holdout experiments, detector-side models achieved comparable or slightly better timing performance than direct coincidence-level neural models while retaining independently applicable singles-level outputs. The learned calibration generalized to unseen source positions and detector pairs, enabled recombination of independently trained scanner components after scalar timing-reference alignment, provided event-wise coincidence-timing uncertainty stratification in both full-ring systems, and remained compatible with low-precision inference after quantization-aware training. Significance: Coincidence supervision enables reusable detector-side timing corrections that improve CTR without single-event timing labels. This preserves a singles-level calibration interface while enabling machine-learning-based nonlinear timing correction, modular scanner calibration, and event-specific coincidence-timing uncertainty estimates.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-11-26
DOI
https://doi.org/10.1088/1361-6560/aec75b
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Detector-side neural timing calibration for TOF-PET from coincidence supervision

Yannick Kuhl, Bjoern Weissler, S. Gundacker, Vanessa Nadig et al.
Physics in Medicine and Biology
Medical Imaging Techniques and Applications
article

Detector-side neural timing calibration for TOF-PET from coincidence supervision

Yannick Kuhl, Bjoern Weissler, S. Gundacker, Vanessa Nadig, Stephan Naunheim, Volkmar Schulz, David Schug, Julian Thull, Marc von Henning, Florian Mueller, L. Lopes de Paiva, Thore Meyer
article en

Abstract

Objective: Machine-learning methods can model non-linear event-dependent timing effects in time-of-flight positron emission tomography (TOF-PET), but calibration measurements naturally provide timing supervision only at the coincidence-level rather than for individual detector interactions. We investigate whether coincidence measurements can be used to learn reusable detector-side timing calibration functions without individually labelled interaction times. Approach: A shared detector-conditioned neural network independently processes each single-event readout and predicts a detector-side timing correction and variance contribution. During training, two detector-side timing estimates are differenced and compared with the geometrically expected time difference obtained from point-source coincidence measurements. We evaluate the method on three heterogeneous PET systems against conventional time-skew and time-walk calibration and matched direct coincidence-level neural models. Main results: Under a common energy-window setting, the proposed model reduced coincidence timing resolution from 321.7 to 263.0ps, 462.4 to 431.7ps, and 417.0 to 382.0ps relative to charge-based time-walk correction on the three systems. In detector-pair holdout experiments, detector-side models achieved comparable or slightly better timing performance than direct coincidence-level neural models while retaining independently applicable singles-level outputs. The learned calibration generalized to unseen source positions and detector pairs, enabled recombination of independently trained scanner components after scalar timing-reference alignment, provided event-wise coincidence-timing uncertainty stratification in both full-ring systems, and remained compatible with low-precision inference after quantization-aware training. Significance: Coincidence supervision enables reusable detector-side timing corrections that improve CTR without single-event timing labels. This preserves a singles-level calibration interface while enabling machine-learning-based nonlinear timing correction, modular scanner calibration, and event-specific coincidence-timing uncertainty estimates.

Physics in Medicine and Biology
RWTH Aachen University (DE)
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon