Estimating balloon-centre insertion depth for fluoroscopy-free REBOA with minimal body-surface measurements: development and external evaluation of a prediction model using CT angiography
Background In battlefield and austere emergency settings, resuscitative endovascular balloon occlusion of the aorta (REBOA) can temporarily control non-compressible haemorrhage below the diaphragm. Without fluoroscopy, the depth to the centre of the target aortic zone must be estimated. The gain in accuracy from additional body-surface measurements is uncertain. To evaluate midpoint error and simulated positioning with the smallest of the fixed input sets compared for REBOA. Methods In this retrospective study, of 10,453 patients screened across centres after CT angiography (CTA), 547 of 8,722 screened for development and 94 of 1,731 screened for external evaluation were included. Forward selection among 14 measurements and sex identified body thickness, suprasternal notch-to-pubic symphysis distance and body width. Ridge regression estimated four total distances to zone midpoints and six component distances. We compared continuous errors across models and also evaluated calibration and reference-window hits for the four total distances. Comparators included constants, larger input sets and other models. Intervals used patient-level bootstrapping for continuous measures and exact binomial limits for hit rates. Paired t-tests and exact McNemar tests were used in the relevant comparisons, with Holm adjustment. Results With three body-surface inputs, mean absolute errors in the external cohort were 15.8 and 16.3 mm on the right and left in zone I and 11.6 and 12.2 mm in zone Ⅲ. The three-input model and constants hit all zone I windows on both sides (94/94); the three-input model hit 88/94 right and 90/94 left zone Ⅲ windows. Conclusions Among the fixed input sets compared, the three body-surface measurements balanced measurement burden and midpoint error. They estimated the zone midpoint more accurately than constants even in zone I, where reference-window hit rates were identical. These CT findings may inform depth estimation for fluoroscopy-free REBOA in battlefield and austere emergency settings.
Authors
- SHAOFAN WANG (ORCID: https://orcid.org/0000-0002-3045-624X)
- Yangyang Ge (ORCID: https://orcid.org/0000-0002-6878-4943)
- Feiyang Zhong (ORCID: https://orcid.org/0000-0003-3833-6573)
- Jianhan Yin
- Guo Wei (ORCID: https://orcid.org/0000-0001-9988-0498)
- Dan Rong (ORCID: https://orcid.org/0000-0002-0680-8735)
- Tianfeng Ma (ORCID: https://orcid.org/0000-0001-7571-0996)
- Haibo Wang
- Long Cao
Institutions
- Nankai University (CN)
- Chinese PLA General Hospital (CN)
- Eastern Command Hospital (IN)
Publication Details
- Journal
- Research Square
- Published
- 2026-11-27
- DOI
- https://doi.org/10.21203/rs.3.rs-11812826/v1
- Primary Topic
- Trauma and Emergency Care Studies
- Type
- preprint