Association between FMEA occurrence rankings and linear accelerator maintenance episode counts

Objective. Failure mode and effects analysis (FMEA) guides quality control (QC) priorities, but occurrence rankings based on tolerance exceedances may not represent corrective maintenance. We assessed whether higher historical QC occurrence ranks corresponded to more corrective maintenance episodes in the same QC domain. Methods. We analyzed 643 maintenance records and 30 523 numerical QC observations from three accelerators. For each accelerator, consecutive corrective maintenance entries with the same documented mechanism and gaps of no more than three calendar days were grouped into an episode under the 72 h rule. A fixed crosswalk assigned eligible episodes to 12 historical FMEA items through direct or indirect numerical mappings. We compared occurrence ranks from 2017–2021 with episode counts from 2018–2021 across 36 accelerator and QC item units using Spearman correlation, a QC item cluster bootstrap, block permutation, and Kendall tau- b. Sensitivity analyses examined the unbinned occurrence probabilities, episode definitions, and exclusion of the multileaf collimator (MLC) domain. Results. The primary maintenance period contained 393 eligible episodes. Of 224 episodes with numerical representation, 213 (95.1%) had only indirect or surrogate mappings; 209 entered the 12 item crosswalk. The QC worksheets contained 41 confirmed tolerance exceedances in 10 of 36 units. Laser localization accounted for 18 exceedances but no mapped maintenance episode. MLC leaf position accounted for two exceedances and 137 mapped episodes, including 74 leaf drive motor failures. Spearman correlation was 0.19 (95% QC item cluster resampling interval, −0.36 to 0.65); Kendall tau- b was 0.17 (95% interval, −0.33 to 0.57). Numerical coverage was absent for 169 episodes (43.0%). Conclusion. Historical occurrence ranks showed a weak, imprecisely estimated association with maintenance episode counts. Differences between QC domains indicate a need to check how QC measures relate to documented maintenance mechanisms before using these scores to represent maintenance activity.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-11-26
DOI
https://doi.org/10.1088/2057-1976/aec0a7
Primary Topic
Quality and Safety in Healthcare
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Association between FMEA occurrence rankings and linear accelerator maintenance episode counts

Jianrong Dai, Min Ma, Hongkai Wang, Ke Zhang
Biomedical Physics & Engineering Express
Quality and Safety in Healthcare
article

Association between FMEA occurrence rankings and linear accelerator maintenance episode counts

Jianrong Dai, Min Ma, Hongkai Wang, Ke Zhang
article en

Abstract

Objective. Failure mode and effects analysis (FMEA) guides quality control (QC) priorities, but occurrence rankings based on tolerance exceedances may not represent corrective maintenance. We assessed whether higher historical QC occurrence ranks corresponded to more corrective maintenance episodes in the same QC domain. Methods. We analyzed 643 maintenance records and 30 523 numerical QC observations from three accelerators. For each accelerator, consecutive corrective maintenance entries with the same documented mechanism and gaps of no more than three calendar days were grouped into an episode under the 72 h rule. A fixed crosswalk assigned eligible episodes to 12 historical FMEA items through direct or indirect numerical mappings. We compared occurrence ranks from 2017–2021 with episode counts from 2018–2021 across 36 accelerator and QC item units using Spearman correlation, a QC item cluster bootstrap, block permutation, and Kendall tau- b. Sensitivity analyses examined the unbinned occurrence probabilities, episode definitions, and exclusion of the multileaf collimator (MLC) domain. Results. The primary maintenance period contained 393 eligible episodes. Of 224 episodes with numerical representation, 213 (95.1%) had only indirect or surrogate mappings; 209 entered the 12 item crosswalk. The QC worksheets contained 41 confirmed tolerance exceedances in 10 of 36 units. Laser localization accounted for 18 exceedances but no mapped maintenance episode. MLC leaf position accounted for two exceedances and 137 mapped episodes, including 74 leaf drive motor failures. Spearman correlation was 0.19 (95% QC item cluster resampling interval, −0.36 to 0.65); Kendall tau- b was 0.17 (95% interval, −0.33 to 0.57). Numerical coverage was absent for 169 episodes (43.0%). Conclusion. Historical occurrence ranks showed a weak, imprecisely estimated association with maintenance episode counts. Differences between QC domains indicate a need to check how QC measures relate to documented maintenance mechanisms before using these scores to represent maintenance activity.

Biomedical Physics & Engineering ExpressVol. 12(6)
Chinese Academy of Medical Sciences & Peking Union Medical College (CN)
Quality and Safety in Healthcare
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon