Performance Evaluation of HBV RNA Real-Time PCR Kits for Quantitative Detection and Clinical Comparison Across Different Detection Systems

Serum hepatitis B virus (HBV) RNA has emerged as a potential biomarker of residual viral transcriptional activity in patients with chronic hepatitis B (CHB). However, insufficient assay standardization and variability across testing systems limit the reliability and comparability of HBV RNA measurements. This study aimed to verify the analytical performance of a commercial HBV RNA detection assay (Kit A) and assess its methodological comparability with another commercial assay (Kit B) using clinical serum samples under different testing conditions. This study comprised analytical performance verification of Kit A, clinical sample comparison between Kit A and Kit B using contemporaneous fresh samples, long-term frozen samples, and non-contemporaneous refrigerated samples, and an exploratory comparison of amplification platforms. Both kits were tested in parallel on the same amplification instrument using different nucleic acid extraction platforms, with operators blinded to the other reagent’s results, and analyses included qualitative agreement, quantitative bias, and low-concentration detection concordance. Kit A met all predefined analytical performance criteria, including trueness (mean biases of −0.06 and −0.03 log10 copies/mL), linearity (R2 = 0.9997; slope 1.008), the 50 copies/mL limit of detection (20/20 detected), precision (CVs of 1.22% and 2.29%), interference resistance (|B| < 7.5%), and analytical specificity (no HBV RNA target detected in 30 tests). In clinical comparisons, qualitative agreement was 88.7% for contemporaneous fresh samples, 92.5% for long-term frozen samples, and 72.9% for non-contemporaneous refrigerated samples, with discordant results concentrated in low-concentration samples; among samples with Kit B results ≥ 300 copies/mL, Kit A showed strong quantitative correlation but positive systematic bias (mean biases of 0.428 and 0.668 log10 copies/mL, respectively, with proportional bias on Passing–Bablok regression). Exploratory platform comparison suggested a higher detection rate on the Applied Biosystems 7500 Fast Real-Time PCR System and the Applied Biosystems™ QuantStudio™ 5 Real-Time PCR System, but the limited data were insufficient to determine overall platform differences. Kit A met the predefined analytical performance requirements under this laboratory’s conditions, but quantitative bias and low-concentration detection differences between the two HBV RNA testing systems underscore the need to standardize calibrators, assay target regions, sample pretreatment, and amplification platforms to improve result comparability.

Authors

Institutions

Publication Details

Journal
Microorganisms
Published
2026-11-26
DOI
https://doi.org/10.3390/microorganisms14122760
Primary Topic
Hepatitis B Virus Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Performance Evaluation of HBV RNA Real-Time PCR Kits for Quantitative Detection and Clinical Comparison Across Different Detection Systems

Yuyuan Jia, 张海莹, Huiying Rao, Runling Zhang et al.
Microorganisms
Hepatitis B Virus Studies
article

Performance Evaluation of HBV RNA Real-Time PCR Kits for Quantitative Detection and Clinical Comparison Across Different Detection Systems

Yuyuan Jia, 张海莹, Huiying Rao, Runling Zhang, Feng Liu, Hongsong Chen, Xin Wang, Xin Ai
article en

Abstract

Serum hepatitis B virus (HBV) RNA has emerged as a potential biomarker of residual viral transcriptional activity in patients with chronic hepatitis B (CHB). However, insufficient assay standardization and variability across testing systems limit the reliability and comparability of HBV RNA measurements. This study aimed to verify the analytical performance of a commercial HBV RNA detection assay (Kit A) and assess its methodological comparability with another commercial assay (Kit B) using clinical serum samples under different testing conditions. This study comprised analytical performance verification of Kit A, clinical sample comparison between Kit A and Kit B using contemporaneous fresh samples, long-term frozen samples, and non-contemporaneous refrigerated samples, and an exploratory comparison of amplification platforms. Both kits were tested in parallel on the same amplification instrument using different nucleic acid extraction platforms, with operators blinded to the other reagent’s results, and analyses included qualitative agreement, quantitative bias, and low-concentration detection concordance. Kit A met all predefined analytical performance criteria, including trueness (mean biases of −0.06 and −0.03 log10 copies/mL), linearity (R2 = 0.9997; slope 1.008), the 50 copies/mL limit of detection (20/20 detected), precision (CVs of 1.22% and 2.29%), interference resistance (|B| < 7.5%), and analytical specificity (no HBV RNA target detected in 30 tests). In clinical comparisons, qualitative agreement was 88.7% for contemporaneous fresh samples, 92.5% for long-term frozen samples, and 72.9% for non-contemporaneous refrigerated samples, with discordant results concentrated in low-concentration samples; among samples with Kit B results ≥ 300 copies/mL, Kit A showed strong quantitative correlation but positive systematic bias (mean biases of 0.428 and 0.668 log10 copies/mL, respectively, with proportional bias on Passing–Bablok regression). Exploratory platform comparison suggested a higher detection rate on the Applied Biosystems 7500 Fast Real-Time PCR System and the Applied Biosystems™ QuantStudio™ 5 Real-Time PCR System, but the limited data were insufficient to determine overall platform differences. Kit A met the predefined analytical performance requirements under this laboratory’s conditions, but quantitative bias and low-concentration detection differences between the two HBV RNA testing systems underscore the need to standardize calibrators, assay target regions, sample pretreatment, and amplification platforms to improve result comparability.

MicroorganismsVol. 14(12)
Peking University People's Hospital (CN)
Hepatitis B Virus Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon