Rapid Volumetric Shimming of the Lung in Free‐Breathing
PURPOSE: To develop a rapid, free-breathing, automated framework for volumetric lung shimming. THEORY AND METHODS: Seven volunteers were scanned at 0.55, 1.5, and 3 T. First, shimming was performed offline from volumetric B0 field information using a comprehensive set of axial, coronal, and sagittal magnitude and phase images acquired with a 2D echo-shifted gradient echo sequence. A combinatorial search for a sparse selection of slice positions was then performed to yield highly similar shim coefficients. For this sparse set, image post-processing and shimming were performed inline and evaluated on a complete rescan of all volunteers at all fields. Finally, the prospects of dedicated lung-shimming were evaluated at 1.5 T using extended TR regimes for balanced steady-state free precession (bSSFP). RESULTS: From the volumetric scans, a fixed sparse selection, consisting of three coronal and two sagittal slice positions, was identified. The sparse set of slice selections resulted in no apparent differences in the shim field as compared to the full set, but data acquisition took only about 15 s (full set: 2-3 min) and shimming was completed in less than 1 min (full set: about 10 min). For all field strengths, shimming offered an increase of at least 30% in the upper limit of TR for artifact-free bSSFP MRI. CONCLUSION: Dedicated volumetric lung shimming is feasible and relaxes TR constraints for artifact-free bSSFP MRI. This offers new prospects for high-resolution imaging, lowering SAR, and reducing acoustic noise from low to high field strengths.
Authors
- Grzegorz Bauman (ORCID: https://orcid.org/0000-0001-6972-0776)
- Oliver Bieri (ORCID: https://orcid.org/0000-0002-6755-5495)
- Pavlos Panos (ORCID: https://orcid.org/0009-0009-4112-5292)
Institutions
- University of Basel (CH)
- University Hospital of Basel (CH)
Publication Details
- Journal
- Magnetic Resonance in Medicine
- Published
- 2026-11-25
- DOI
- https://doi.org/10.1002/mrm.70673
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00