A laboratory case of delipidation of a blood sample with lipemia for hematological analysis

The pre-analytical stage in laboratory practice critically affects the reliability of research results. Blood samples with signs of chylemia, hemolysis, or ictericity arriving at the laboratory can cause significant distortion of laboratory test results. Lipemia, in particular, affects optical density when measuring hemoglobin concentration, which entails errors in obtaining the result of a complete blood count, and, consequently, in diagnosis and therapy monitoring. At the same time, existing guidelines [1] lack protocols for pre-treatment of lipemic samples for hematological testing, which underscores the relevance of developing and validating effective delipidation methods. The aim of the study was to validate the blood sample delipidation method by Zeng S.G. et al. [2] for the correction of hemoglobin measurement on an automated hematology analyzer. Materials and methods. Biochemical and hematological analyses were performed on a blood sample from patient A. (40 years old, diagnosis — diabetes mellitus), in whom chylous sample was detected during fasting blood collection. Lipid profile analysis was performed using a COBAS INTEGRA 400 plus analyzer; hematological testing was performed on a UniCel DxH 800 (Beckman Coulter) instrument. The analysis was performed twice: before and after sample preparation according to the Zeng S.G. method. Sample preparation was as follows: the sample with K3EDTA was centrifuged at 500 rpm for 15 minutes, then the lipemic plasma was removed and the volume was replenished with an equal amount of physiological saline. Results. In the chylous sample, the hemoglobin level reached 270 g/L, which is above the reference values. Due to lipemia, MCH and MCHC parameters were also overestimated. After delipidation, the hemoglobin concentration decreased to 119 g/L; a decrease in other hematological parameters was also noted: leukocytes from 9.5 to 4.8·109/L, platelets from 277 to 146·106/L, erythrocytes from 4.35 to 3.59 ·109/L. Hypertriglyceridemia was confirmed by the triglyceride level of 132.05 mmol/L (reference values up to 2.3 mmol/L). Conclusions. The blood delipidation method by Zeng S.G. [2] allows obtaining correct hemoglobin values in chylous samples, but causes a decrease in blood cell parameters, which limits its application for performing a complete blood count. Further studies are needed to adapt this method and correct the measurement of all hematological parameters.

Authors

Institutions

Publication Details

Journal
Laboratory Service
Published
2026-11-26
DOI
https://doi.org/10.17116/labs20261503127
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A laboratory case of delipidation of a blood sample with lipemia for hematological analysis

L.B. Gaikovaya, S.A. Karpich, N.G. Kobzeva, Y.S. Lantratova et al.
Laboratory Service
Clinical Laboratory Practices and Quality Control
article

A laboratory case of delipidation of a blood sample with lipemia for hematological analysis

L.B. Gaikovaya, S.A. Karpich, N.G. Kobzeva, Y.S. Lantratova, E.F. Gulyakina
article en

Abstract

The pre-analytical stage in laboratory practice critically affects the reliability of research results. Blood samples with signs of chylemia, hemolysis, or ictericity arriving at the laboratory can cause significant distortion of laboratory test results. Lipemia, in particular, affects optical density when measuring hemoglobin concentration, which entails errors in obtaining the result of a complete blood count, and, consequently, in diagnosis and therapy monitoring. At the same time, existing guidelines [1] lack protocols for pre-treatment of lipemic samples for hematological testing, which underscores the relevance of developing and validating effective delipidation methods. The aim of the study was to validate the blood sample delipidation method by Zeng S.G. et al. [2] for the correction of hemoglobin measurement on an automated hematology analyzer. Materials and methods. Biochemical and hematological analyses were performed on a blood sample from patient A. (40 years old, diagnosis — diabetes mellitus), in whom chylous sample was detected during fasting blood collection. Lipid profile analysis was performed using a COBAS INTEGRA 400 plus analyzer; hematological testing was performed on a UniCel DxH 800 (Beckman Coulter) instrument. The analysis was performed twice: before and after sample preparation according to the Zeng S.G. method. Sample preparation was as follows: the sample with K3EDTA was centrifuged at 500 rpm for 15 minutes, then the lipemic plasma was removed and the volume was replenished with an equal amount of physiological saline. Results. In the chylous sample, the hemoglobin level reached 270 g/L, which is above the reference values. Due to lipemia, MCH and MCHC parameters were also overestimated. After delipidation, the hemoglobin concentration decreased to 119 g/L; a decrease in other hematological parameters was also noted: leukocytes from 9.5 to 4.8·109/L, platelets from 277 to 146·106/L, erythrocytes from 4.35 to 3.59 ·109/L. Hypertriglyceridemia was confirmed by the triglyceride level of 132.05 mmol/L (reference values up to 2.3 mmol/L). Conclusions. The blood delipidation method by Zeng S.G. [2] allows obtaining correct hemoglobin values in chylous samples, but causes a decrease in blood cell parameters, which limits its application for performing a complete blood count. Further studies are needed to adapt this method and correct the measurement of all hematological parameters.

Laboratory ServiceVol. 15(3)
North-Western State Medical University named after I.I. Mechnikov (RU)
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon