Atrial Pacing Capability of a Single‐Lead Implantable Cardioverter‐Defibrillator Incorporating an Atrial Floating Dipole

INTRODUCTION: The single-lead DX implantable cardioverter-defibrillator (ICD) combines the advantages of a conventional single-lead ICD with atrial sensing capability provided by an integrated atrial floating dipole. While atrial sensing performance has been extensively documented, the ability of the floating atrial dipole to achieve atrial capture has not previously been investigated. METHODS AND RESULTS: Consecutive patients undergoing de novo implantation or generator replacement of a single-lead DX ICD system at two Italian centers were prospectively enrolled. Acute atrial pacing was assessed using three configurations: bipolar pacing and two unipolar configurations employing either the proximal or distal atrial electrode. Atrial capture, pacing threshold (PT), impedance, and phrenic nerve stimulation were evaluated at pulse widths of 0.4 and 1.0 ms. Twenty-five patients (92% male; mean age 66.2 ± 10.1 years) were included. Acute atrial capture was achieved in 22 patients (88%) with bipolar pacing, compared with 18 (72%) and 17 (68%) using proximal and distal unipolar pacing, respectively (p = 0.122). Bipolar pacing yielded significantly lower PTs than unipolar configurations at both pulse widths (p < 0.020). Median atrial sensing amplitude measured after connection to the ICD generator was 3.9 mV (2.8-5.1). CONCLUSION: Atrial capture through the floating atrial dipole of a single-lead DX ICD system is feasible in the majority of patients, with bipolar pacing providing the highest capture rate and the lowest PTs. These findings provide proof-of-concept evidence supporting further development of single-lead dual-chamber sensing and pacing strategies using the DX platform.

Authors

Institutions

Publication Details

Journal
Journal of Cardiovascular Electrophysiology
Published
2026-11-25
DOI
https://doi.org/10.1111/jce.70571
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Atrial Pacing Capability of a Single‐Lead Implantable Cardioverter‐Defibrillator Incorporating an Atrial Floating Dipole

P Fornaro, Gianmarco Arabia, Manuel Cerini, Emiliano Calvi et al.
Journal of Cardiovascular Electrophysiology
Cardiac pacing and defibrillation studies
article

Atrial Pacing Capability of a Single‐Lead Implantable Cardioverter‐Defibrillator Incorporating an Atrial Floating Dipole

P Fornaro, Gianmarco Arabia, Manuel Cerini, Emiliano Calvi, Alessio Nicolai, Gianfranco Mitacchione, Gioele Papi, Antonio Curnis, Aurora Giavarini
article en

Abstract

INTRODUCTION: The single-lead DX implantable cardioverter-defibrillator (ICD) combines the advantages of a conventional single-lead ICD with atrial sensing capability provided by an integrated atrial floating dipole. While atrial sensing performance has been extensively documented, the ability of the floating atrial dipole to achieve atrial capture has not previously been investigated. METHODS AND RESULTS: Consecutive patients undergoing de novo implantation or generator replacement of a single-lead DX ICD system at two Italian centers were prospectively enrolled. Acute atrial pacing was assessed using three configurations: bipolar pacing and two unipolar configurations employing either the proximal or distal atrial electrode. Atrial capture, pacing threshold (PT), impedance, and phrenic nerve stimulation were evaluated at pulse widths of 0.4 and 1.0 ms. Twenty-five patients (92% male; mean age 66.2 ± 10.1 years) were included. Acute atrial capture was achieved in 22 patients (88%) with bipolar pacing, compared with 18 (72%) and 17 (68%) using proximal and distal unipolar pacing, respectively (p = 0.122). Bipolar pacing yielded significantly lower PTs than unipolar configurations at both pulse widths (p < 0.020). Median atrial sensing amplitude measured after connection to the ICD generator was 3.9 mV (2.8-5.1). CONCLUSION: Atrial capture through the floating atrial dipole of a single-lead DX ICD system is feasible in the majority of patients, with bipolar pacing providing the highest capture rate and the lowest PTs. These findings provide proof-of-concept evidence supporting further development of single-lead dual-chamber sensing and pacing strategies using the DX platform.

Journal of Cardiovascular Electrophysiology
Azienda Socio Sanitaria Territoriale degli Spedali Civili di Brescia (IT), Istituti Ospitalieri di Cremona (IT), University of Brescia (IT), Biotronik (Switzerland) (CH)
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon