Agreement and systematic bias between the InBody 770 and VisBody S30 for body composition assessment in competitive athletes: a method-comparison study
Body composition results from different multifrequency bioelectrical impedance analysis systems may not be interchangeable because device architecture and proprietary algorithms differ. This study quantified agreement and systematic bias between the InBody 770 and VisBody S30 in competitive athletes. In this cross-sectional method-comparison study, 105 athletes completed consecutive measurements with both analyzers. Six records failed internal plausibility checks, leaving 99 paired observations. Body weight, body mass index, percentage body fat, body fat mass, fat-free mass, skeletal muscle mass, and total body water were evaluated using paired differences, Bland-Altman limits of agreement, Lin’s concordance correlation coefficient, absolute-agreement intraclass correlation coefficients, and proportional-bias tests. Concordance coefficients ranged from 0.940 to 1.000. Body weight and body mass index differed minimally. VisBody reported higher percentage body fat by 2.42% points and higher body fat mass by 1.74 kg. InBody reported higher fat-free mass by 1.81 kg, higher skeletal muscle mass by 0.61 kg, and higher total body water by 1.45 L. Proportional bias was present for percentage body fat, fat-free mass, and total body water. Despite high concordance, the analyzers produced outcome-specific systematic offsets and should not be used interchangeably for individual longitudinal monitoring. Repeated assessments should use the same device under standardized conditions.
Authors
- Csaba Melczer (ORCID: https://orcid.org/0000-0002-8197-0572)
- Iván Petrov (ORCID: https://orcid.org/0009-0003-0773-1877)
Institutions
- University of Pecs (HU)
Publication Details
- Journal
- BMC Sports Science Medicine and Rehabilitation
- Published
- 2026-11-26
- DOI
- https://doi.org/10.1186/s13102-026-02286-4
- Primary Topic
- Body Composition Measurement Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00