Microbial barrier integrity of 3-way-stopcocks following disinfection with and without mechanical challenge
Background The study assessed the microbial barrier integrity of 3-way stopcocks exposed to various disinfecting agents in terms of stress cracks with potential relevance for infection prevention and control. Methods Four different 3-way stopcocks were filled with medium containing each more than 100,000 colony-forming units of Staphylococcus epidermidis, Pseudomonas aeruginosa, Escherichia coli and Candida albicans, with subsequent exposure to four different disinfecting agents for 6 and 24 h as well as for 28 days, exceeding realistic real-word exposure times. Mechanical challenge was simulated by repeated docking of syringes, combined with 24 h disinfection turns. After the disinfection terms, lacking bacterial growth from closed 3-way stopcocks as well as positive growth from the bacterial suspension inside was confirmed in brain heart infusion broth. Results Absence of bacterial growth from closed 3-way stopcocks and remaining vitality of the bacteria in the suspension within the stopcocks could be confirmed in all instances. Conclusions The 3-way stopcocks showed a robust barrier microbial integrity against long-term disinfection without and with combined mechanical challenging. The results suggest that infection prevention and control risks due to stress cracks caused by disinfection of 3-way stopcocks seems to be negligible and that their disinfection is safe to be performed.
Authors
- Philipp Warnke
- Hagen Frickmann (ORCID: https://orcid.org/0000-0002-8967-9528)
- Andreas Podbielski (ORCID: https://orcid.org/0000-0003-1017-439X)
- Tobias Hans Conrad
Institutions
- Bundeswehrkrankenhaus (DE)
- Universitätsmedizin Rostock (DE)
Publication Details
- Journal
- European Journal of Microbiology and Immunology
- Published
- 2026-11-26
- DOI
- https://doi.org/10.1556/1886.2026.00056
- Primary Topic
- Infection Control in Healthcare
- Type
- article
- Field-Weighted Citation Impact
- 0.00