Introduction The rapid expansion of radiopharmaceutical therapies (RPTs) has increased the need for multi-center clinical trials, yet achieving consistent dosimetry across institutions with different imaging equipment remains challenging. This study developed a practical and reliable calibration approach for a multi-center pan-Canadian 177 Lu-PSMA-617 trial. Specifically, a centralized core laboratory was established to cross-calibrate equipment and support dosimetry calculations across all sites. This approach reduced inter-site variability and minimized the need for radioactive source handling at remote centers. The method is applicable to future RPT clinical trials requiring harmonized dosimetry. Methods The core laboratory radionuclide calibrator was first calibrated against the manufacturer’s (Novartis AG, Basel, Switzerland) reference standard and validated with a high-purity germanium (HPGe) detector calibrated with a NIST-traceable source. Reference radioactive sources were then prepared, documented, and shipped to the participating sites. Sites performed planar and SPECT/CT acquisitions using these sources and adjusted their radionuclide calibrator dial settings to match the core laboratory activity reference. Sites submitted acquisition data to the core laboratory for verification and protocol compliance. The sensitivity and dead-time constant for each SPECT/CT system were calculated centrally. Calibration consistency between manufacturer and the core laboratory was evaluated by comparing reported patient vial activities with measurements from the core laboratory radionuclide calibrator. Results Eleven sites across Canada including a total of 22 radionuclide calibrators and 12 SPECT/CT systems were calibrated using this method. Although the core-laboratory calibrator was independently anchored to a NIST-traceable HPGe measurement (agreement within 0.5%), manufacturer-stated activities of subsequent vials were systematically lower, with discrepancies up to 13% (median 10%) over three years. The mean SPECT/CT sensitivity was 9.23 ± 0.15 cps/MBq for a Siemens camera with a 3/8″ crystal (n = 5) and 12.54 cps/MBq for a Siemens camera with a 5/8" crystal (n = 1). The mean sensitivity was 4.91 ± 0.53 cps/MBq for the GE camera with a 3/8" crystal (n = 6). The mean dead-time constant was 1.08 ± 0.16 µs (Siemens, n = 6) and 2.95 ± 0.75 µs (GE, n = 5). Harmonization reduced inter-site activity variability across all geometries (e.g., point source: SD 10.3% → 1.6%). Conclusion Centralized calibration harmonization using a core laboratory is feasible and effective for multi-center RPT trials. Preparing reference sources centrally and performing centralized dosimetry calculations ensures consistent and comparable measurements across participating sites. Critically, manufacturer-stated activities drifted by up to 13% during the trial; independent verification against a stable, traceable reference is therefore essential for multi-center dosimetry.

Authors

Institutions

Publication Details

Journal
EJNMMI Physics
Published
2026-11-25
DOI
https://doi.org/10.1186/s40658-026-00957-2
Primary Topic
Radiopharmaceutical Chemistry and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Harmonizing calibrations for multi-center 177Lu dosimetry: insights from the Pan-Canadian PR.21 trial

Pedro Luis Esquinas, Sara Kurkowska, François Bénard, Nadine Colpo et al.
4 links
EJNMMI Physics
Radiopharmaceutical Chemistry and Applications
article

Harmonizing calibrations for multi-center 177Lu dosimetry: insights from the Pan-Canadian PR.21 trial

Pedro Luis Esquinas, Sara Kurkowska, François Bénard, Nadine Colpo, Conor Dellar, Daniel Juneau, Carlos Uribe, Jean‐Mathieu Beauregard, Fred Saad, Wendy Parulekar, Kim N. Chi, Katherine Zukotynski
article en

Abstract

Introduction The rapid expansion of radiopharmaceutical therapies (RPTs) has increased the need for multi-center clinical trials, yet achieving consistent dosimetry across institutions with different imaging equipment remains challenging. This study developed a practical and reliable calibration approach for a multi-center pan-Canadian 177 Lu-PSMA-617 trial. Specifically, a centralized core laboratory was established to cross-calibrate equipment and support dosimetry calculations across all sites. This approach reduced inter-site variability and minimized the need for radioactive source handling at remote centers. The method is applicable to future RPT clinical trials requiring harmonized dosimetry. Methods The core laboratory radionuclide calibrator was first calibrated against the manufacturer’s (Novartis AG, Basel, Switzerland) reference standard and validated with a high-purity germanium (HPGe) detector calibrated with a NIST-traceable source. Reference radioactive sources were then prepared, documented, and shipped to the participating sites. Sites performed planar and SPECT/CT acquisitions using these sources and adjusted their radionuclide calibrator dial settings to match the core laboratory activity reference. Sites submitted acquisition data to the core laboratory for verification and protocol compliance. The sensitivity and dead-time constant for each SPECT/CT system were calculated centrally. Calibration consistency between manufacturer and the core laboratory was evaluated by comparing reported patient vial activities with measurements from the core laboratory radionuclide calibrator. Results Eleven sites across Canada including a total of 22 radionuclide calibrators and 12 SPECT/CT systems were calibrated using this method. Although the core-laboratory calibrator was independently anchored to a NIST-traceable HPGe measurement (agreement within 0.5%), manufacturer-stated activities of subsequent vials were systematically lower, with discrepancies up to 13% (median 10%) over three years. The mean SPECT/CT sensitivity was 9.23 ± 0.15 cps/MBq for a Siemens camera with a 3/8″ crystal (n = 5) and 12.54 cps/MBq for a Siemens camera with a 5/8" crystal (n = 1). The mean sensitivity was 4.91 ± 0.53 cps/MBq for the GE camera with a 3/8" crystal (n = 6). The mean dead-time constant was 1.08 ± 0.16 µs (Siemens, n = 6) and 2.95 ± 0.75 µs (GE, n = 5). Harmonization reduced inter-site activity variability across all geometries (e.g., point source: SD 10.3% → 1.6%). Conclusion Centralized calibration harmonization using a core laboratory is feasible and effective for multi-center RPT trials. Preparing reference sources centrally and performing centralized dosimetry calculations ensures consistent and comparable measurements across participating sites. Critically, manufacturer-stated activities drifted by up to 13% during the trial; independent verification against a stable, traceable reference is therefore essential for multi-center dosimetry.

EJNMMI Physics
BC Cancer Agency (CA), University of British Columbia (CA), Queen's University (CA), Centre hospitalier universitaire de Québec (CA), Centre Hospitalier de l’Université de Montréal (CA), Université Laval (CA), Canadian Cancer Trials Group (CA), BC Cancer Research Institute (CA), Pomeranian Medical University (PL), McMaster University (CA)
Canadian Cancer Society
Good health and well-being
Radiopharmaceutical Chemistry and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon