Feedback-controlled ultrafiltration as a useful tool for the diagnosis of volume status assessment in dialysis patients: a pilot study

BACKGROUND: Volume overload is one of the most critical challenges in dialysis patients and is monitored using bioelectrical impedance analysis (BIA). The feedback-controlled ultrafiltration (fUF) is used to reduce hypovolemic complications, but its diagnostic power has not been tested.Our analysis aims to evaluate the ability of the fUF to detect volume status alterations. METHODS: This study enrolled 29 dialysis patients. Volume status was defined by BIA as hypo/normo/hyper-volemia. The volume status assessment using the fUF was defined through the difference between the prescribed ultrafiltration and the effective ultrafiltration. Sensitivity, Specificity, Positive Predictive Value (PPV), Negative Predictive Value (NPV), and Accuracy were evaluated. The analysis was performed in subsamples based on the extracellular/intracellular (E/I) ratio of fluid distribution. RESULTS: For hypervolemia, the model showed a sensitivity of 83.3%, a specificity of 91.3%, a PPV of 71.4%, and an NPV of 95.5%. For hypovolemia the analysis showed a sensitivity of 70.0%, a specificity of 73.7%, a PPV of 58.3%, and an NPV of 82.4%. The correlation coefficient between BIA and fUF was 0.502 (p=0.006). Stratification by the E/I ratio revealed improved diagnostic performance in the subgroup with an E/I ratio > 0.8 (n=17). Global accuracy was 70.6%. The metrics for hypervolemia reached a sensitivity of 100% and an NPV of 100%, a specificity of 92.9%, a PPV of 75%, and a balanced Accuracy of 96.4%. CONCLUSIONS: Although further larger studies are needed, our pilot findings indicate potential exploratory utility for feedback-controlled ultrafiltration in assessing volume status among dialysis patients.

Authors

Institutions

Publication Details

Journal
Journal of Nephrology
Published
2026-11-25
DOI
https://doi.org/10.1093/joneph/aajag287
Primary Topic
Dialysis and Renal Disease Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Feedback-controlled ultrafiltration as a useful tool for the diagnosis of volume status assessment in dialysis patients: a pilot study

Maria Eva Sberna, Maria Rita Stancanelli, Vincenzo Calabrese, Giuseppe Restivo et al.
2 links
Journal of Nephrology
Dialysis and Renal Disease Management
article

Feedback-controlled ultrafiltration as a useful tool for the diagnosis of volume status assessment in dialysis patients: a pilot study

Maria Eva Sberna, Maria Rita Stancanelli, Vincenzo Calabrese, Giuseppe Restivo, Veronica Cucci, Veronica Fortunato, Salvuccio Siciliano, Pietro Suriano, Gabriele Carmelo Morreale
article en

Abstract

BACKGROUND: Volume overload is one of the most critical challenges in dialysis patients and is monitored using bioelectrical impedance analysis (BIA). The feedback-controlled ultrafiltration (fUF) is used to reduce hypovolemic complications, but its diagnostic power has not been tested.Our analysis aims to evaluate the ability of the fUF to detect volume status alterations. METHODS: This study enrolled 29 dialysis patients. Volume status was defined by BIA as hypo/normo/hyper-volemia. The volume status assessment using the fUF was defined through the difference between the prescribed ultrafiltration and the effective ultrafiltration. Sensitivity, Specificity, Positive Predictive Value (PPV), Negative Predictive Value (NPV), and Accuracy were evaluated. The analysis was performed in subsamples based on the extracellular/intracellular (E/I) ratio of fluid distribution. RESULTS: For hypervolemia, the model showed a sensitivity of 83.3%, a specificity of 91.3%, a PPV of 71.4%, and an NPV of 95.5%. For hypovolemia the analysis showed a sensitivity of 70.0%, a specificity of 73.7%, a PPV of 58.3%, and an NPV of 82.4%. The correlation coefficient between BIA and fUF was 0.502 (p=0.006). Stratification by the E/I ratio revealed improved diagnostic performance in the subgroup with an E/I ratio > 0.8 (n=17). Global accuracy was 70.6%. The metrics for hypervolemia reached a sensitivity of 100% and an NPV of 100%, a specificity of 92.9%, a PPV of 75%, and a balanced Accuracy of 96.4%. CONCLUSIONS: Although further larger studies are needed, our pilot findings indicate potential exploratory utility for feedback-controlled ultrafiltration in assessing volume status among dialysis patients.

Journal of Nephrology
Università degli Studi di Enna Kore (IT)
Good health and well-being
Dialysis and Renal Disease Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon