Performance and comparative analysis of the touchless home sleep monitor Sleepiz One+ with polysomnography in women with sleep disturbance associated with menopause (Preprint)

BACKGROUND Sleep disturbances can affect up to 60% of postmenopausal women and can have a profound impact on quality of life, mental health, and physical health. Polysomnography (PSG) conducted within a sleep laboratory is the reference standard for assessing sleep disturbance, however, it has several limitations. Home sleep monitors, such as the contactless Sleepiz One+ device, counter some limitations of in-lab PSG and provide valuable insights into sleep health, as well as the variability of sleep between nights, due to longitudinal measures in the user’s natural sleep environment at home. It is important to gain insight into the level of agreement between PSG and Sleepiz One+ to identify its suitability for use in future clinical trials assessing sleep disturbances in a clinically relevant population. OBJECTIVE To assess the performance of Sleepiz One+ compared with reference standard in-lab PSG and further evaluate Sleepiz One+ as a home-based sleep monitoring tool for measuring sleep disturbance associated with menopause. METHODS Agreement between modalities was exploratorily assessed using epoch-level (sleep/wake classification in 30-second intervals) and sleep parameter-level analyses using data from postmenopausal women with sleep disturbances who participated in the randomized NIRVANA Phase II trial (NCT06112756). Synchronized PSG and Sleepiz One+ assessments were conducted over 3 visits (6 nights total) within a laboratory setting, and Sleepiz One+ data were also collected nightly in the participant’s home over a period of 12-weeks. RESULTS Epoch-level classification of sleep/wake status by Sleepiz One+ was mostly consistent across nights. In 237,258 epochs from 66 subjects, sensitivity (to detect sleep), specificity (to detect wake), accuracy, and inter-device reliability (prevalence-adjusted bias-adjusted kappa [PABAK]) in comparison to PSG were estimated to be 91%, 64%, 86%, and 0.72, respectively. Mean differences between Sleepiz One+ and PSG at the parameter-level were: wakefulness after sleep onset (WASO) −4.8 minutes; sleep efficiency (SE) +2.1%; number of awakenings (NAW) −0.4. Negative values indicated underestimation by Sleepiz One+ and positive indicated overestimation. Test-retest reliability was high (intraclass correlation coefficients: WASO 0.91–0.96; SE 0.87–0.98; NAW 0.94–0.97), with ≥4 nights per week recommended to yield a weekly average. CONCLUSIONS Sleepiz One+ is a reliable tool for sleep assessment in clinical trials with comparable performance to PSG, addressing the unmet need for longitudinal monitoring of natural sleep at-home, and should be considered for future studies assessing menopause-associated sleep disturbance. CLINICALTRIAL A Study to Learn About How Elinzanetant Works and How Safe it is in Women Having Sleep Disturbances Associated With Menopause (NIRVANA), Phase II, randomized clinical trial, NCT06112756, study registration date 2023-10-27, https://clinicaltrials.gov/study/NCT06112756

Authors

Publication Details

Published
2026-11-25
DOI
https://doi.org/10.2196/preprints.117364
Primary Topic
Sleep and related disorders
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Performance and comparative analysis of the touchless home sleep monitor Sleepiz One+ with polysomnography in women with sleep disturbance associated with menopause (Preprint)

Christian Seitz, Marc Rullan, Hoi-Shen Radcliffe, Gary K Zammit et al.
6 links
Sleep and related disorders
preprint

Performance and comparative analysis of the touchless home sleep monitor Sleepiz One+ with polysomnography in women with sleep disturbance associated with menopause (Preprint)

Christian Seitz, Marc Rullan, Hoi-Shen Radcliffe, Gary K Zammit, Ulrike Krahn, Andrew Trigg, Frank Kramer, Jonas Krauss, Harsha Umesh Babu
preprint en

Abstract

BACKGROUND Sleep disturbances can affect up to 60% of postmenopausal women and can have a profound impact on quality of life, mental health, and physical health. Polysomnography (PSG) conducted within a sleep laboratory is the reference standard for assessing sleep disturbance, however, it has several limitations. Home sleep monitors, such as the contactless Sleepiz One+ device, counter some limitations of in-lab PSG and provide valuable insights into sleep health, as well as the variability of sleep between nights, due to longitudinal measures in the user’s natural sleep environment at home. It is important to gain insight into the level of agreement between PSG and Sleepiz One+ to identify its suitability for use in future clinical trials assessing sleep disturbances in a clinically relevant population. OBJECTIVE To assess the performance of Sleepiz One+ compared with reference standard in-lab PSG and further evaluate Sleepiz One+ as a home-based sleep monitoring tool for measuring sleep disturbance associated with menopause. METHODS Agreement between modalities was exploratorily assessed using epoch-level (sleep/wake classification in 30-second intervals) and sleep parameter-level analyses using data from postmenopausal women with sleep disturbances who participated in the randomized NIRVANA Phase II trial (NCT06112756). Synchronized PSG and Sleepiz One+ assessments were conducted over 3 visits (6 nights total) within a laboratory setting, and Sleepiz One+ data were also collected nightly in the participant’s home over a period of 12-weeks. RESULTS Epoch-level classification of sleep/wake status by Sleepiz One+ was mostly consistent across nights. In 237,258 epochs from 66 subjects, sensitivity (to detect sleep), specificity (to detect wake), accuracy, and inter-device reliability (prevalence-adjusted bias-adjusted kappa [PABAK]) in comparison to PSG were estimated to be 91%, 64%, 86%, and 0.72, respectively. Mean differences between Sleepiz One+ and PSG at the parameter-level were: wakefulness after sleep onset (WASO) −4.8 minutes; sleep efficiency (SE) +2.1%; number of awakenings (NAW) −0.4. Negative values indicated underestimation by Sleepiz One+ and positive indicated overestimation. Test-retest reliability was high (intraclass correlation coefficients: WASO 0.91–0.96; SE 0.87–0.98; NAW 0.94–0.97), with ≥4 nights per week recommended to yield a weekly average. CONCLUSIONS Sleepiz One+ is a reliable tool for sleep assessment in clinical trials with comparable performance to PSG, addressing the unmet need for longitudinal monitoring of natural sleep at-home, and should be considered for future studies assessing menopause-associated sleep disturbance. CLINICALTRIAL A Study to Learn About How Elinzanetant Works and How Safe it is in Women Having Sleep Disturbances Associated With Menopause (NIRVANA), Phase II, randomized clinical trial, NCT06112756, study registration date 2023-10-27, https://clinicaltrials.gov/study/NCT06112756

Good health and well-being
Sleep and related disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon