Lesion Formation Using a Dual-Energy Lattice-Tip Ablation Catheter: Effects of Repeated Application, Timing, and Spatial Arrangement

Purpose The dual-energy lattice-tip catheter enables point-by-point ablation using pulsed field (PF) or radiofrequency (RF) energy. In this study, we characterized PF- and RF-induced lesion formation and assessed the effects of repeated applications, inter-application timing, and inter-lesion distance in an ex vivo model. Methods Lesion dimensions were compared after single and repeated PF and RF applications. Inter-application intervals of 0, 30, 60, and 300 s were evaluated. Lesion continuity at depth was assessed using gap depth between adjacent lesions at different inter-lesion distances. RF lesions at 73 and 68°C were also compared. Results Single PF applications produced greater surface lesion length than RF. Repeated applications increased lesion depth with both PF and RF, whereas their effects on surface lesion length differed between the two energy modalities. Compared with immediate repetition, lesion depth did not differ significantly at 30 s but was significantly shallower at 60 and 300 s with both energies, whereas surface dimensions showed no clear time dependence. Increasing inter-lesion distance reduced continuity at depth; gap depth decreased significantly at ≥ 6 mm for PF compared with 4 mm and at 10 mm for RF compared with 7 mm. RF at 68°C produced shorter and shallower lesions than at 73°C. Conclusions Lattice-tip lesion formation is influenced by energy modality and the temporal and spatial pattern of delivery. Repetition increases lesion depth, while delayed additional applications and greater inter-lesion distance reduce depth augmentation and lesion continuity, respectively.

Authors

Institutions

Publication Details

Journal
Research Square
Published
2026-11-25
DOI
https://doi.org/10.21203/rs.3.rs-11699216/v1
Primary Topic
Cardiac Arrhythmias and Treatments
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Lesion Formation Using a Dual-Energy Lattice-Tip Ablation Catheter: Effects of Repeated Application, Timing, and Spatial Arrangement

Yusuke Sakamoto, Hiroyuki Osanai, Eiji Yoshida, Yuichiro Sakai et al.
3 links
Research Square
Cardiac Arrhythmias and Treatments
preprint

Lesion Formation Using a Dual-Energy Lattice-Tip Ablation Catheter: Effects of Repeated Application, Timing, and Spatial Arrangement

Yusuke Sakamoto, Hiroyuki Osanai, Eiji Yoshida, Yuichiro Sakai, Tomotsugu Izuchi
preprint en

Abstract

Purpose The dual-energy lattice-tip catheter enables point-by-point ablation using pulsed field (PF) or radiofrequency (RF) energy. In this study, we characterized PF- and RF-induced lesion formation and assessed the effects of repeated applications, inter-application timing, and inter-lesion distance in an ex vivo model. Methods Lesion dimensions were compared after single and repeated PF and RF applications. Inter-application intervals of 0, 30, 60, and 300 s were evaluated. Lesion continuity at depth was assessed using gap depth between adjacent lesions at different inter-lesion distances. RF lesions at 73 and 68°C were also compared. Results Single PF applications produced greater surface lesion length than RF. Repeated applications increased lesion depth with both PF and RF, whereas their effects on surface lesion length differed between the two energy modalities. Compared with immediate repetition, lesion depth did not differ significantly at 30 s but was significantly shallower at 60 and 300 s with both energies, whereas surface dimensions showed no clear time dependence. Increasing inter-lesion distance reduced continuity at depth; gap depth decreased significantly at ≥ 6 mm for PF compared with 4 mm and at 10 mm for RF compared with 7 mm. RF at 68°C produced shorter and shallower lesions than at 73°C. Conclusions Lattice-tip lesion formation is influenced by energy modality and the temporal and spatial pattern of delivery. Repetition increases lesion depth, while delayed additional applications and greater inter-lesion distance reduce depth augmentation and lesion continuity, respectively.

Research Square
Tosei General Hospital (JP)
Good health and well-being
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Lesion Formation Using a Dual-Energy Lattice-Tip Ablation Catheter: Effects of Repeated Application, Timing, and Spatial Arrangement — Yusuke Sakamoto, Hiroyuki Osanai, et al. · Research Square (2026) | TGRS Research Map | TGRS