Concurrent validity and agreement between VO 2 master and COSMED Quark during maximal treadmill testing

This study evaluated the concurrent validity and agreement between the VO 2 Master (VM) portable metabolic analyzer and the COSMED Quark (CQ) cardiopulmonary exercise testing (CPET) system during maximal incremental treadmill exercise and examined whether sex moderated the agreement between the devices. Twenty-six recreationally active adults (16 men and 10 women), aged 22.1 ± 3.8 years, completed two randomized, counterbalanced cardiopulmonary exercise tests on separate days, one with each device, under standardized laboratory conditions. Oxygen uptake (VO 2) and minute ventilation (V · E) were recorded across 10 incremental stages. Agreement was assessed using intraclass correlation coefficients, Bland–Altman analysis with 95% limits of agreement, stage-wise post hoc comparisons, Cohen’s d, and standard error of measurement. A mixed-design analysis of variance examined sex × equipment interactions. Agreement for VO 2 ranged from poor at Stage 2 to moderate at Stage 8, with moderate intraclass correlation coefficients at Stages 5, 6, 8, and 9 (0.56–0.63); no stage reached the 0.75 threshold for acceptable agreement. Agreement for V · E was stronger, reaching good agreement at Stages 5 and 10. The VM under-reported VO 2 relative to the CQ reference at Stages 1–7 and over-reported it at Stages 8–10; the mean bias across stages was −0.6 mL/kg/min (range of stage-wise biases: −3.4 mL/kg/min at Stage 2 to +5.8 mL/kg/min at Stage 10), with wide stage-wise 95% limits of agreement (approximately ±9–±17 mL/kg/min). Standard error of measurement was generally lower at moderate-to-high intensities. Men showed higher VO 2 peak and peak V · E than women, but no significant sex × equipment interaction was observed. The VM showed moderate agreement with the CQ at moderate-to-maximal intensities, particularly for V · E, but poor agreement at low intensities; therefore, the devices should not be considered interchangeable. Measurement bias did not differ significantly between sexes at the stages analyzed.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part P Journal of Sports Engineering and Technology
Published
2026-11-25
DOI
https://doi.org/10.1177/17543371261509779
Primary Topic
Cardiovascular and exercise physiology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Concurrent validity and agreement between VO 2 master and COSMED Quark during maximal treadmill testing

Roberto Espinoza-Gutiérrez, José Andrés Trejos-Montoya, Jorge Alberto Aburto Corona, Juan J. Calleja-Núñez et al.
3 links
Proceedings of the Institution of Mechanical Engineers Part P Journal of Sports Engineering and Technology
Cardiovascular and exercise physiology
article

Concurrent validity and agreement between VO 2 master and COSMED Quark during maximal treadmill testing

Roberto Espinoza-Gutiérrez, José Andrés Trejos-Montoya, Jorge Alberto Aburto Corona, Juan J. Calleja-Núñez, Daniel Rojas‐Valverde, Cindy Castro-Ramírez
article en

Abstract

This study evaluated the concurrent validity and agreement between the VO 2 Master (VM) portable metabolic analyzer and the COSMED Quark (CQ) cardiopulmonary exercise testing (CPET) system during maximal incremental treadmill exercise and examined whether sex moderated the agreement between the devices. Twenty-six recreationally active adults (16 men and 10 women), aged 22.1 ± 3.8 years, completed two randomized, counterbalanced cardiopulmonary exercise tests on separate days, one with each device, under standardized laboratory conditions. Oxygen uptake (VO 2) and minute ventilation (V · E) were recorded across 10 incremental stages. Agreement was assessed using intraclass correlation coefficients, Bland–Altman analysis with 95% limits of agreement, stage-wise post hoc comparisons, Cohen’s d, and standard error of measurement. A mixed-design analysis of variance examined sex × equipment interactions. Agreement for VO 2 ranged from poor at Stage 2 to moderate at Stage 8, with moderate intraclass correlation coefficients at Stages 5, 6, 8, and 9 (0.56–0.63); no stage reached the 0.75 threshold for acceptable agreement. Agreement for V · E was stronger, reaching good agreement at Stages 5 and 10. The VM under-reported VO 2 relative to the CQ reference at Stages 1–7 and over-reported it at Stages 8–10; the mean bias across stages was −0.6 mL/kg/min (range of stage-wise biases: −3.4 mL/kg/min at Stage 2 to +5.8 mL/kg/min at Stage 10), with wide stage-wise 95% limits of agreement (approximately ±9–±17 mL/kg/min). Standard error of measurement was generally lower at moderate-to-high intensities. Men showed higher VO 2 peak and peak V · E than women, but no significant sex × equipment interaction was observed. The VM showed moderate agreement with the CQ at moderate-to-maximal intensities, particularly for V · E, but poor agreement at low intensities; therefore, the devices should not be considered interchangeable. Measurement bias did not differ significantly between sexes at the stages analyzed.

Proceedings of the Institution of Mechanical Engineers Part P Journal of Sports Engineering and Technology
Universidad Autónoma de Baja California (MX), Universidad Nacional (CR), Universidad Santo Tomás (CO)
Good health and well-being
Cardiovascular and exercise physiology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon