Coupled Transport–Capture in Aligned ZIF-8/Polyacrylonitrile Electrospun Fibrous Membranes: In-Line Hemoglobin Removal for Enzymatic and pH-Responsive Colorimetric Test Strips

Hemoglobin (Hb) interferes with colorimetric assays because its strong visible absorption masks the chromogenic signal. We report a proof-of-concept test strip in which a separation pad capturing Hb in-line during capillary flow is coupled to downstream colorimetric detection pads. The separation pad is an aligned electrospun polyacrylonitrile membrane bearing in-situ-grown zeolitic imidazolate framework-8 (ZIF-8). Although the as-spun mats show apparent water contact angles of 110-128°, all wick Hb solution laterally; a two-barrier wetting analysis attributes the apparent nonwetting to the outermost fiber layer and indicates that fiber alignment accelerates transport mainly by lowering this entry barrier. Increasing the Zn2+ precursor concentration from 100 to 400 mM raises the 90 min Freundlich constant for Hb 2.6-fold, whereas bovine serum albumin is not measurably depleted, and the framework remains intact after Hb exposure, indicating Zn2+-mediated affinity at the external ZIF-8 surface. The same loading lowers permeability and narrows the operating window under flow from 80 to 20 mg mL-1 Hb; the 200 mM formulation combines 42% more capacity than the 100 mM formulation with a 17% lower intrinsic flow rate and generates filtrate up to 40 mg mL-1 Hb. Integrated with pyrrolidonyl arylamidase and oxidase pads for necrotizing fasciitis triage and with an m-cresol purple pad for meat-spoilage monitoring, the separation pad preserved 74-126% of the stand-alone colorimetric signal at matched bacterial load or pH in Hb-spiked simulated samples. Undiluted whole blood lies outside the present operating window, and validation with real samples remains to be performed.

Authors

Institutions

Publication Details

Journal
ACS Applied Bio Materials
Published
2026-11-25
DOI
https://doi.org/10.1021/acsabm.6c01242
Primary Topic
Biosensors and Analytical Detection
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Coupled Transport–Capture in Aligned ZIF-8/Polyacrylonitrile Electrospun Fibrous Membranes: In-Line Hemoglobin Removal for Enzymatic and pH-Responsive Colorimetric Test Strips

Dzul Fadli Alimin, Pitt Supaphol, Piyachat Chuysinuan, Pairayaphak Ngamplang et al.
2 links
ACS Applied Bio Materials
Biosensors and Analytical Detection
article

Coupled Transport–Capture in Aligned ZIF-8/Polyacrylonitrile Electrospun Fibrous Membranes: In-Line Hemoglobin Removal for Enzymatic and pH-Responsive Colorimetric Test Strips

Dzul Fadli Alimin, Pitt Supaphol, Piyachat Chuysinuan, Pairayaphak Ngamplang, Paranee Phuking
article en

Abstract

Hemoglobin (Hb) interferes with colorimetric assays because its strong visible absorption masks the chromogenic signal. We report a proof-of-concept test strip in which a separation pad capturing Hb in-line during capillary flow is coupled to downstream colorimetric detection pads. The separation pad is an aligned electrospun polyacrylonitrile membrane bearing in-situ-grown zeolitic imidazolate framework-8 (ZIF-8). Although the as-spun mats show apparent water contact angles of 110-128°, all wick Hb solution laterally; a two-barrier wetting analysis attributes the apparent nonwetting to the outermost fiber layer and indicates that fiber alignment accelerates transport mainly by lowering this entry barrier. Increasing the Zn2+ precursor concentration from 100 to 400 mM raises the 90 min Freundlich constant for Hb 2.6-fold, whereas bovine serum albumin is not measurably depleted, and the framework remains intact after Hb exposure, indicating Zn2+-mediated affinity at the external ZIF-8 surface. The same loading lowers permeability and narrows the operating window under flow from 80 to 20 mg mL-1 Hb; the 200 mM formulation combines 42% more capacity than the 100 mM formulation with a 17% lower intrinsic flow rate and generates filtrate up to 40 mg mL-1 Hb. Integrated with pyrrolidonyl arylamidase and oxidase pads for necrotizing fasciitis triage and with an m-cresol purple pad for meat-spoilage monitoring, the separation pad preserved 74-126% of the stand-alone colorimetric signal at matched bacterial load or pH in Hb-spiked simulated samples. Undiluted whole blood lies outside the present operating window, and validation with real samples remains to be performed.

ACS Applied Bio Materials
Chulalongkorn University (TH), Chulabhorn Research Institute (TH)
Chulalongkorn University, Chulabhorn Research Institute
Biosensors and Analytical Detection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon