Within-Session Repeatability of Spectral-Domain OCT Angiography and Inter-Device Agreement with Swept-Source OCT Angiography in Diabetic Eyes

Purpose: The aim of this study was to assess the repeatability of a spectral-domain (SD) OCTA device, the reproducibility of its measurements between raters, and its agreement with a swept-source (SS) OCTA device in diabetic eyes. Methods: Each right eye underwent three consecutive 12 × 12 mm macular OCTA scans on the SD-OCTA device and one on the SS-OCTA device, acquired in the same session by the same examiner. Retinal thickness and vascular density in the fovea, parafovea and perifovea, and the areas of the foveal avascular zone (FAZ) and capillary non-perfusion (CNP), were exported for analysis. Repeatability and inter-rater reproducibility were assessed by ICCs, and inter-device agreement by Bland–Altman analysis. Results: A total of 282 right eyes of 282 participants with type 2 diabetes were imaged. Inter-rater reproducibility was good for retinal thickness in all ETDRS areas (ICCs 0.82–0.85) but poor for all vascular parameters (all ICCs < 0.3). Repeatability of the SD-OCTA device was excellent for retinal thickness, FAZ area and CNP area (ICCs > 0.95) and moderate for foveal vascular density (ICC = 0.77). CNP detection was concordant between the two devices (κ = 0.854, 95% CI 0.77–0.94), but the SS-OCTA device read 32.91 ± 58.35 μm (fovea) to 62.01 ± 28.50 μm (perifovea) thicker than the SD-OCTA device did (95% limits of agreement: −81.5 to 147.3 μm in the fovea and 6.2 to 117.9 μm in the perifovea). Conclusions: The SD-OCTA device gave repeatable measurements of macular thickness, FAZ area and CNP area, similar to those reported for SS-OCTA systems. The two devices also agreed on whether CNP was present. However, their quantitative values were not interchangeable, and the vascular parameters were poorly reproducible between raters. The device may be a more accessible alternative, but its value for screening has not been tested, and device-specific reference values should be used.

Authors

Institutions

Publication Details

Journal
Vision
Published
2026-11-24
DOI
https://doi.org/10.3390/vision10050100
Primary Topic
Retinal Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Within-Session Repeatability of Spectral-Domain OCT Angiography and Inter-Device Agreement with Swept-Source OCT Angiography in Diabetic Eyes

Shanhu Qiu, 张忠红, Jie Luan, Zilin Sun et al.
Vision
Retinal Diseases and Treatments
article

Within-Session Repeatability of Spectral-Domain OCT Angiography and Inter-Device Agreement with Swept-Source OCT Angiography in Diabetic Eyes

Shanhu Qiu, 张忠红, Jie Luan, Zilin Sun, Jian Xu, Yuzhi Ding, Ruibo Chen
article en

Abstract

Purpose: The aim of this study was to assess the repeatability of a spectral-domain (SD) OCTA device, the reproducibility of its measurements between raters, and its agreement with a swept-source (SS) OCTA device in diabetic eyes. Methods: Each right eye underwent three consecutive 12 × 12 mm macular OCTA scans on the SD-OCTA device and one on the SS-OCTA device, acquired in the same session by the same examiner. Retinal thickness and vascular density in the fovea, parafovea and perifovea, and the areas of the foveal avascular zone (FAZ) and capillary non-perfusion (CNP), were exported for analysis. Repeatability and inter-rater reproducibility were assessed by ICCs, and inter-device agreement by Bland–Altman analysis. Results: A total of 282 right eyes of 282 participants with type 2 diabetes were imaged. Inter-rater reproducibility was good for retinal thickness in all ETDRS areas (ICCs 0.82–0.85) but poor for all vascular parameters (all ICCs < 0.3). Repeatability of the SD-OCTA device was excellent for retinal thickness, FAZ area and CNP area (ICCs > 0.95) and moderate for foveal vascular density (ICC = 0.77). CNP detection was concordant between the two devices (κ = 0.854, 95% CI 0.77–0.94), but the SS-OCTA device read 32.91 ± 58.35 μm (fovea) to 62.01 ± 28.50 μm (perifovea) thicker than the SD-OCTA device did (95% limits of agreement: −81.5 to 147.3 μm in the fovea and 6.2 to 117.9 μm in the perifovea). Conclusions: The SD-OCTA device gave repeatable measurements of macular thickness, FAZ area and CNP area, similar to those reported for SS-OCTA systems. The two devices also agreed on whether CNP was present. However, their quantitative values were not interchangeable, and the vascular parameters were poorly reproducible between raters. The device may be a more accessible alternative, but its value for screening has not been tested, and device-specific reference values should be used.

VisionVol. 10(5)
Zhongda Hospital Southeast University (CN), Nanjing Drum Tower Hospital (CN), Southeast University (CN), Nanjing University (CN)
Retinal Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon