Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions

This exploratory bench study characterized pressure behavior in a roller pump-driven low-flow valve-type semi-closed extracorporeal circulation (VACC) mock circuit under experimentally reduced patient-side effective filling pressure conditions with the de-airing pump inactive. Pressures were measured at four sites: circuit-reservoir pressure (Pc), patient-side mock-reservoir outlet pressure (Pv), roller pump inlet pressure (Pi), and roller pump outlet pressure (Po). Pump flow-rate settings were 50–300 mL/min. Circuit-reservoir negative-pressure setpoints were 0, −10, −20, −30, and −40 mmHg, and patient-side mock-reservoir negative-pressure setpoints were 0, −5, and −10 mmHg. A 32 wt% aqueous glycerol solution was used. Pc generally reflected the circuit-reservoir negative-pressure setpoint, with a mean difference of −0.18 mmHg (95% CI, −0.37 to 0.00). The 95% CIs summarize variability among technical replicates within this bench system and should not be interpreted as population-inferential estimates. Pv decreased stepwise according to the patient-side setpoint. Within the common evaluation range, mean differences versus 0 mmHg were −5.7 mmHg (95% CI, −6.0 to −5.3) for the −5 mmHg condition and −9.2 mmHg (95% CI, −9.5 to −8.9) for the −10 mmHg condition. Pi shifted in the negative direction as the pump flow-rate setting increased, whereas Po increased monotonically under the fixed occluder condition. Delivered flow was independently verified volumetrically before the main experiments at 100, 200, and 300 mL/min but was not continuously measured during the pressure experiments. Therefore, Po was interpreted primarily in relation to the pump flow-rate setting rather than as a directly measured hydraulic pressure–flow relationship. These data provide baseline pressure-characterization information for future mock-circuit studies and do not establish clinical safety, efficacy, or a stable operating range.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2026-11-25
DOI
https://doi.org/10.1371/journal.pone.0363658
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions

Masahiro Miyata, Takahiro Okumura, Kenji Tani, Keisuke Matsuda et al.
PLoS ONE
Mechanical Circulatory Support Devices
article

Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions

Masahiro Miyata, Takahiro Okumura, Kenji Tani, Keisuke Matsuda, Keizo Wauchi, Hiroki Sato
article en

Abstract

This exploratory bench study characterized pressure behavior in a roller pump-driven low-flow valve-type semi-closed extracorporeal circulation (VACC) mock circuit under experimentally reduced patient-side effective filling pressure conditions with the de-airing pump inactive. Pressures were measured at four sites: circuit-reservoir pressure (Pc), patient-side mock-reservoir outlet pressure (Pv), roller pump inlet pressure (Pi), and roller pump outlet pressure (Po). Pump flow-rate settings were 50–300 mL/min. Circuit-reservoir negative-pressure setpoints were 0, −10, −20, −30, and −40 mmHg, and patient-side mock-reservoir negative-pressure setpoints were 0, −5, and −10 mmHg. A 32 wt% aqueous glycerol solution was used. Pc generally reflected the circuit-reservoir negative-pressure setpoint, with a mean difference of −0.18 mmHg (95% CI, −0.37 to 0.00). The 95% CIs summarize variability among technical replicates within this bench system and should not be interpreted as population-inferential estimates. Pv decreased stepwise according to the patient-side setpoint. Within the common evaluation range, mean differences versus 0 mmHg were −5.7 mmHg (95% CI, −6.0 to −5.3) for the −5 mmHg condition and −9.2 mmHg (95% CI, −9.5 to −8.9) for the −10 mmHg condition. Pi shifted in the negative direction as the pump flow-rate setting increased, whereas Po increased monotonically under the fixed occluder condition. Delivered flow was independently verified volumetrically before the main experiments at 100, 200, and 300 mL/min but was not continuously measured during the pressure experiments. Therefore, Po was interpreted primarily in relation to the pump flow-rate setting rather than as a directly measured hydraulic pressure–flow relationship. These data provide baseline pressure-characterization information for future mock-circuit studies and do not establish clinical safety, efficacy, or a stable operating range.

PLoS ONEVol. 21(11)
Yamaguchi University (JP), Saitama Women's Junior College (JP), Saitama Children's Medical Center (JP), Hidaka Hospital (JP), Sendai Medical Center (JP), National Hospital Organization (JP), Teikyo University (JP)
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon