Detection of Manifest Strabismus Using the Spot Vision Screener at a University Hospital in Japan: A Retrospective Study

Background This study aimed to evaluate the performance of the Spot Vision Screener (SVS) for the detection of manifest strabismus in children using same-day clinical ocular alignment examination as the reference standard. Methodology This retrospective study included 364 pediatric examinations wherein SVS screening and clinical ocular alignment examination were performed on the same day between December 2021 and November 2025. SVS screening was routinely performed by trained orthoptists before the clinical examination. Clinical ocular alignment examinations were performed by ophthalmologists or orthoptists. SVS screening results were compared with the presence or absence of manifest strabismus determined by the clinical ocular alignment examination. Sensitivity, specificity, positive predictive value, and negative predictive value were calculated with 95% confidence intervals. Results The median patient age was 37 months (interquartile range = 22-50 months). Manifest strabismus was identified in 117 (32.1%) examinations. The SVS demonstrated high specificity (98.0%) but moderate sensitivity (65.0%) for the detection of manifest strabismus. Most false-negative results occurred in children with intermittent exotropia. In an analysis restricted to constant esotropia and constant exotropia, sensitivity increased to 94.3%. Conclusions In this pediatric ophthalmology population, the SVS demonstrated high specificity but limited sensitivity for the detection of manifest strabismus, primarily because intermittent exotropia was difficult to detect. Although the SVS is a useful instrument-based screening tool, a negative screening result should not preclude referral for clinical ocular alignment examination when a parent, caregiver, or clinician has concerns regarding possible strabismus. These findings support the intended role of the SVS as a screening tool rather than as a diagnostic test.

Authors

Publication Details

Journal
Cureus
Published
2026-11-25
DOI
https://doi.org/10.7759/cureus.120236
Primary Topic
Ophthalmology and Eye Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Detection of Manifest Strabismus Using the Spot Vision Screener at a University Hospital in Japan: A Retrospective Study

Nobuyuki Ebihara, Noriko Tsutsumi, Asuka Nakamura
Cureus
Ophthalmology and Eye Disorders
article

Detection of Manifest Strabismus Using the Spot Vision Screener at a University Hospital in Japan: A Retrospective Study

Nobuyuki Ebihara, Noriko Tsutsumi, Asuka Nakamura
article en

Abstract

Background This study aimed to evaluate the performance of the Spot Vision Screener (SVS) for the detection of manifest strabismus in children using same-day clinical ocular alignment examination as the reference standard. Methodology This retrospective study included 364 pediatric examinations wherein SVS screening and clinical ocular alignment examination were performed on the same day between December 2021 and November 2025. SVS screening was routinely performed by trained orthoptists before the clinical examination. Clinical ocular alignment examinations were performed by ophthalmologists or orthoptists. SVS screening results were compared with the presence or absence of manifest strabismus determined by the clinical ocular alignment examination. Sensitivity, specificity, positive predictive value, and negative predictive value were calculated with 95% confidence intervals. Results The median patient age was 37 months (interquartile range = 22-50 months). Manifest strabismus was identified in 117 (32.1%) examinations. The SVS demonstrated high specificity (98.0%) but moderate sensitivity (65.0%) for the detection of manifest strabismus. Most false-negative results occurred in children with intermittent exotropia. In an analysis restricted to constant esotropia and constant exotropia, sensitivity increased to 94.3%. Conclusions In this pediatric ophthalmology population, the SVS demonstrated high specificity but limited sensitivity for the detection of manifest strabismus, primarily because intermittent exotropia was difficult to detect. Although the SVS is a useful instrument-based screening tool, a negative screening result should not preclude referral for clinical ocular alignment examination when a parent, caregiver, or clinician has concerns regarding possible strabismus. These findings support the intended role of the SVS as a screening tool rather than as a diagnostic test.

Cureus
Ophthalmology and Eye Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon