Impact Attenuation Performance of a Lightweight, Ventilated Two-Zone Natural Rubber Hip Protector: Experimental Testing and Finite Element Optimization

Hip fractures caused by sideways falls impose a major health and economic burden on older adults, yet hip protectors are often rejected because they are thick, heavy, and uncomfortable. This study used finite element (FE) modeling and drop-tower testing to develop a lightweight two-zone natural rubber hip protector (TZNRHP), comprising a stiff solid-rubber inner zone over the greater trochanter and a closed-cell rubber-foam outer zone. A hyperelastic–viscoelastic FE model was validated against drop-tower tests on a surrogate hip (4.75 kg impactor, 0.52 m drop), and protectors 9–15 mm thick with six ventilation-hole patterns were evaluated (n = 5 repeated impacts; one-way ANOVA with Tukey’s test). The FE model predicted peak forces within 5.4% for all validation conditions. Thickness significantly affected peak force (p < 0.001, η2 = 0.99), and the 11 mm protector was the thinnest design significantly below the 3100 N fracture criterion, reducing the peak force from 6232 ± 160 N (unprotected) to 2785 ± 79 N (55.3% attenuation) while being 28.3% lighter than the original 15 mm design. For a female-specific threshold of 2800 N, the 12 mm design was required. Thirteen 5 mm ventilation holes (8.4% open area) raised the peak force by only 2.6% (2856 ± 73 N; not significantly different from the least-perforated pattern), and the 11 mm protector reduced the maximum femoral von Mises stress from 9.2 to 5.83 MPa. The optimized TZNRHP is a promising lightweight, ventilated hip protector for older adults.

Authors

Institutions

Publication Details

Journal
Polymers
Published
2026-11-25
DOI
https://doi.org/10.3390/polym18232967
Primary Topic
Automotive and Human Injury Biomechanics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Impact Attenuation Performance of a Lightweight, Ventilated Two-Zone Natural Rubber Hip Protector: Experimental Testing and Finite Element Optimization

Theerawat Petdee, Boonsin Tangtrakulwanich, Wiriya Thongruang, Varah Yuenyongviwat et al.
Polymers
Automotive and Human Injury Biomechanics
article

Impact Attenuation Performance of a Lightweight, Ventilated Two-Zone Natural Rubber Hip Protector: Experimental Testing and Finite Element Optimization

Theerawat Petdee, Boonsin Tangtrakulwanich, Wiriya Thongruang, Varah Yuenyongviwat, Satta Srewaradachpisal, Charoenyutr Dechwayukul
article en

Abstract

Hip fractures caused by sideways falls impose a major health and economic burden on older adults, yet hip protectors are often rejected because they are thick, heavy, and uncomfortable. This study used finite element (FE) modeling and drop-tower testing to develop a lightweight two-zone natural rubber hip protector (TZNRHP), comprising a stiff solid-rubber inner zone over the greater trochanter and a closed-cell rubber-foam outer zone. A hyperelastic–viscoelastic FE model was validated against drop-tower tests on a surrogate hip (4.75 kg impactor, 0.52 m drop), and protectors 9–15 mm thick with six ventilation-hole patterns were evaluated (n = 5 repeated impacts; one-way ANOVA with Tukey’s test). The FE model predicted peak forces within 5.4% for all validation conditions. Thickness significantly affected peak force (p < 0.001, η2 = 0.99), and the 11 mm protector was the thinnest design significantly below the 3100 N fracture criterion, reducing the peak force from 6232 ± 160 N (unprotected) to 2785 ± 79 N (55.3% attenuation) while being 28.3% lighter than the original 15 mm design. For a female-specific threshold of 2800 N, the 12 mm design was required. Thirteen 5 mm ventilation holes (8.4% open area) raised the peak force by only 2.6% (2856 ± 73 N; not significantly different from the least-perforated pattern), and the 11 mm protector reduced the maximum femoral von Mises stress from 9.2 to 5.83 MPa. The optimized TZNRHP is a promising lightweight, ventilated hip protector for older adults.

PolymersVol. 18(23)
Prince of Songkla University (TH)
Automotive and Human Injury Biomechanics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon