Implantation site influences tissue response to vaginal meshes in a material dependent manner
Polypropylene (PP) meshes are associated with significant complications when implanted transvaginally for the treatment of stress urinary incontinence (SUI) and pelvic organ prolapse (POP). Efforts to circumvent this include developing more biocompatible synthetic materials and improving the pre-clinical development pathways. In this study, we compared the tissue response to novel vaginal meshes in a more traditional abdominal wall implantation and a more representative vaginal implantation model. Two novel electrospun poly-urethane (PU) and poly-lactic acid (PLA) materials were implanted in the vesico-vaginal space (VVs) and abdominal wall of female rabbits. Estradiol (E 2 )-loaded PU (PU-E 2 ) was also implanted to compare it against unloaded PU mesh. A polypropylene (PP) mesh was used as a control. After three months, assessment of the implantation site revealed distinct reactions to the implants between the abdominal wall and vesicovaginal implantation sites. Extrusion and contraction of the meshes were only seen for PP and PLA in the vaginal implantation sites. Histologic examination revealed a mild inflammatory response to both PU and PP meshes in both implantation sites; while a strong inflammatory response was observed for PLA meshes. The capsule formation was similar between all groups and implantation sites, except for a thicker capsule for PP meshes implanted vaginally. Additionally, estradiol loaded PU meshes increased the cell infiltration and blood vessel formation compared to unloaded PU, particularly after vaginal implantation. The tissue response to mesh materials differed between abdominal and vaginal implantation in a material-dependent manner (significant for PLA and PU-E 2 but not PP or PU). Implant-related complications were observed only following vaginal implantation. These findings suggest that the implantation site may influence the preclinical evaluation of biomaterials intended for vaginal application. Further studies are warranted to determine whether anatomically relevant implantation models provide additional predictive value for the preclinical assessment of vaginal biomaterials.
Authors
- Özlem Bozdemir (ORCID: https://orcid.org/0000-0003-0592-1530)
- Vanessa L. Hearnden (ORCID: https://orcid.org/0000-0003-0838-7783)
- Kezban Ulubayram (ORCID: https://orcid.org/0000-0003-0365-298X)
- Şükrü Öztürk (ORCID: https://orcid.org/0000-0001-9684-634X)
- Sheila MacNeil (ORCID: https://orcid.org/0000-0002-9188-5769)
- Christopher Reginald Chapple (ORCID: https://orcid.org/0000-0002-2960-9931)
- Naşide Mangır (ORCID: https://orcid.org/0000-0002-3062-6480)
- Naciye Dilara Zeybek (ORCID: https://orcid.org/0000-0002-6161-5661)
- Sabiniano Roman Regueros
Institutions
- Sheffield Teaching Hospitals NHS Foundation Trust (GB)
- Hacettepe University (TR)
- University of Sheffield (GB)
Publication Details
- Journal
- Scientific Reports
- Published
- 2026-11-25
- DOI
- https://doi.org/10.1038/s41598-026-79231-5
- Primary Topic
- Pelvic floor disorders treatments
- Type
- article
- Field-Weighted Citation Impact
- 0.00