IonGate: Circuit Memory and Waveform-Protected Learning for Functional Inference in Cardiac Electrophysiology

Background Functional assessment of SCN5A and hERG channels supports arrhythmia research, but voltage-clamp feedback can distort both recorded waveforms and derived functional quantities. We developed IonGate to coordinate voltage reconstruction and state-aware functional inference in one model. Results We retrospectively evaluated a frozen circuit-informed observer, branch-memory readout and waveform-protected learning architecture across 16 compensation settings from one electronic-model-cell session. An independent second-amplifier voltage channel provided the reconstruction reference. Affine and finite-impulse-response controls received matched incremental enhancement budgets. Fixed-waveform ablation isolated the readout contribution; a retained biological application comprised 450 sweeps from four cells. Median/P95 voltage root-mean-square errors were 0.252515/2.565794 mV; nominal resistance–capacitance functional errors were 0.105459%/0.443297%. Median functional error was 49.47% and 52.18% lower than affine and finite-impulse-response controls, respectively; voltage P95 was 28.99% and 28.00% lower. Removing branch memory from identical predicted waveforms increased median functional error to 0.349058%; the full readout improved 14/16 settings. Residual calibration contained 16/16 nominal references, with wider intervals and worse proper scores than matched controls. The unchanged biological adapter retained 20/24 identifiable activation fits, matching enhanced controls. Conclusions IonGate improved waveform and nominal circuit-functional point estimates together and separated the contributions of memory, bounded learning and calibration. References used measured voltage and nominal component values, not independent ionic-charge metrology. Clinical utility and cross-session generalization require further study.

Authors

Institutions

Publication Details

Journal
Research Square
Published
2026-11-25
DOI
https://doi.org/10.21203/rs.3.rs-11665941/v1
Primary Topic
Cardiac electrophysiology and arrhythmias
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

IonGate: Circuit Memory and Waveform-Protected Learning for Functional Inference in Cardiac Electrophysiology

Jicheng Li, Peng Xue, yu an xu, tianyi ma
Research Square
Cardiac electrophysiology and arrhythmias
preprint

IonGate: Circuit Memory and Waveform-Protected Learning for Functional Inference in Cardiac Electrophysiology

Jicheng Li, Peng Xue, yu an xu, tianyi ma
preprint en

Abstract

Background Functional assessment of SCN5A and hERG channels supports arrhythmia research, but voltage-clamp feedback can distort both recorded waveforms and derived functional quantities. We developed IonGate to coordinate voltage reconstruction and state-aware functional inference in one model. Results We retrospectively evaluated a frozen circuit-informed observer, branch-memory readout and waveform-protected learning architecture across 16 compensation settings from one electronic-model-cell session. An independent second-amplifier voltage channel provided the reconstruction reference. Affine and finite-impulse-response controls received matched incremental enhancement budgets. Fixed-waveform ablation isolated the readout contribution; a retained biological application comprised 450 sweeps from four cells. Median/P95 voltage root-mean-square errors were 0.252515/2.565794 mV; nominal resistance–capacitance functional errors were 0.105459%/0.443297%. Median functional error was 49.47% and 52.18% lower than affine and finite-impulse-response controls, respectively; voltage P95 was 28.99% and 28.00% lower. Removing branch memory from identical predicted waveforms increased median functional error to 0.349058%; the full readout improved 14/16 settings. Residual calibration contained 16/16 nominal references, with wider intervals and worse proper scores than matched controls. The unchanged biological adapter retained 20/24 identifiable activation fits, matching enhanced controls. Conclusions IonGate improved waveform and nominal circuit-functional point estimates together and separated the contributions of memory, bounded learning and calibration. References used measured voltage and nominal component values, not independent ionic-charge metrology. Clinical utility and cross-session generalization require further study.

Research Square
Second Affiliated Hospital of Xuzhou Medical College (CN), Monash University (AU)
Cardiac electrophysiology and arrhythmias
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.