Development of BaO–Fe2O3–SiO2 glass electrodes with enhanced pH response in low-conductivity water relative to commercial glass electrodes

Measuring the pH of low-conductivity water (LCW), including ultrapure water, remains difficult because conventional ion-conducting Li 2 O–SiO 2 glass electrodes respond slowly and lose accuracy at very low ionic strength. Here we introduce BaO–Fe 2 O 3 –SiO 2 glasses as a new class of electron-conducting pH-electrode materials that exploit small-polaron hopping between Fe 2+ and Fe 3+ sites. Unlike commercial lithium-silicate glass, these glasses contain neither lithium nor expensive GeO 2 nor Bi 2 O 3 , and they combine near-Nernstian pH sensitivity with mechanical strength comparable to that of the commercial glass. The glasses exhibited high pH sensitivity (96.7%–102.1%) over the pH 4–9 range, with electrical resistivities of 2.02 × 10 8 –2.21 × 10 9 Ω cm — about an order of magnitude lower than that of the commercial glass (1.06× 10 10 Ω cm) — and lower activation energies (0.31–0.47 eV vs. 0.66 eV). In a pH 7 buffer solution, the glasses reached 90% of their steady-state potential ( t 90 ≤ 1.5 s), versus t 90 = 14.0 s for the commercial glass. Lower activation energies corresponded consistently to shorter t 90 values, indicating that electronic conduction is the key factor accelerating the pH response. In LCW (<0.12 μS cm −1 ), all BaO–Fe 2 O 3 –SiO 2 electrodes read a near-neutral pH of approximately 7.0, whereas the commercial HORIBA glass read 6.96. These results establish BaO–Fe 2 O 3 –SiO 2 glasses as a promising, Li-free route to rapid and reliable pH sensing in ultrapure and other low-conductivity waters; notably, their short t 90 values allow measurements to be completed before significant KCl contamination from the reference electrode occurs, even during batch operation.

Authors

Institutions

Publication Details

Journal
Sensors International
Published
2026-11-24
DOI
https://doi.org/10.1016/j.sintl.2026.100413
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Development of BaO–Fe2O3–SiO2 glass electrodes with enhanced pH response in low-conductivity water relative to commercial glass electrodes

Atsushi Ishihara, Tadanori Hashimoto, Yuji Nishio, Kenta Suzuki et al.
Sensors International
Analytical Chemistry and Sensors
article

Development of BaO–Fe2O3–SiO2 glass electrodes with enhanced pH response in low-conductivity water relative to commercial glass electrodes

Atsushi Ishihara, Tadanori Hashimoto, Yuji Nishio, Kenta Suzuki, Kyota Fujioka
article en

Abstract

Measuring the pH of low-conductivity water (LCW), including ultrapure water, remains difficult because conventional ion-conducting Li 2 O–SiO 2 glass electrodes respond slowly and lose accuracy at very low ionic strength. Here we introduce BaO–Fe 2 O 3 –SiO 2 glasses as a new class of electron-conducting pH-electrode materials that exploit small-polaron hopping between Fe 2+ and Fe 3+ sites. Unlike commercial lithium-silicate glass, these glasses contain neither lithium nor expensive GeO 2 nor Bi 2 O 3 , and they combine near-Nernstian pH sensitivity with mechanical strength comparable to that of the commercial glass. The glasses exhibited high pH sensitivity (96.7%–102.1%) over the pH 4–9 range, with electrical resistivities of 2.02 × 10 8 –2.21 × 10 9 Ω cm — about an order of magnitude lower than that of the commercial glass (1.06× 10 10 Ω cm) — and lower activation energies (0.31–0.47 eV vs. 0.66 eV). In a pH 7 buffer solution, the glasses reached 90% of their steady-state potential ( t 90 ≤ 1.5 s), versus t 90 = 14.0 s for the commercial glass. Lower activation energies corresponded consistently to shorter t 90 values, indicating that electronic conduction is the key factor accelerating the pH response. In LCW (<0.12 μS cm −1 ), all BaO–Fe 2 O 3 –SiO 2 electrodes read a near-neutral pH of approximately 7.0, whereas the commercial HORIBA glass read 6.96. These results establish BaO–Fe 2 O 3 –SiO 2 glasses as a promising, Li-free route to rapid and reliable pH sensing in ultrapure and other low-conductivity waters; notably, their short t 90 values allow measurements to be completed before significant KCl contamination from the reference electrode occurs, even during batch operation.

Sensors InternationalVol. 8
Horiba (Japan) (JP), Mie University (JP)
Japan Society for the Promotion of Science London, Japan Society for the Promotion of Science
Analytical Chemistry and Sensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon