Two Is Not too Small; a Case of a Ventricular Aveir AR2 for Ventricular Pacing in a 12.2-kg Two-Year-Old
Pacemaker implants in pediatric patients can be filled with potential complications. Epicardial pacemaker implants are the usual mainstay for those under 15 kg and transvenous devices can have infectious and pocket erosion risk. Retrievable leadless pacemakers seem to be a potential option. Retrospective review of a 2-year-old patient with high grade AV block and overall lowering average junctional rate was performed. Internal review board approval was obtained. A 2-year-old, 12.2-kg female underwent successful right internal jugular vein implant of an Aveir AR2 into the right ventricle for backup ventricular pacing need. No complications occurred. Follow-up at 12 months demonstrated a threshold of 0.5 Volts @0.2 ms, impedance at 420 ohms and sensing at 10milliVolts with 30% pacing noted with programming at VVI 50 bpm, with predicted longevity at 20.2 years. Feasibility of internal jugular vein implant of an Aveir AR2 is possible without complication in a 2-year-old, 12.2 kg child. Further prospective studies are needed to assess true feasibility in this population.
Authors
- Daniel Cortez (ORCID: https://orcid.org/0000-0001-9874-0733)
- Nandita Sharma (ORCID: https://orcid.org/0000-0003-2191-8462)
Institutions
- University of California Davis Medical Center (US)
- University of California, Davis (US)
Publication Details
- Journal
- Pediatric Cardiology
- Published
- 2026-11-24
- DOI
- https://doi.org/10.1007/s00246-026-04536-1
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- University of California, Davis