A Wired Pressure-Sensing Smart Biliary Stent: Bench Characterization and Feasibility in a Porcine Model

Biliary stents restore drainage in malignant biliary obstruction, but recurrent dysfunction creates a need to monitor the patency of the stent. We developed a pressure-sensing smart biliary stent by integrating a flexible contact-resistive sensor, comprising opposing Au mesh electrodes on liquid crystal polymer films, into a commercial self-expandable metal stent connected to an external monitoring system. Under loading through a compliant pad matched to the 1 mm × 2 mm sensing window (n = 4 sensors), the normalized resistance decreased approximately linearly up to 2.5 kPa and approached a plateau at approximately 3.0 kPa, with a sensitivity of −37.0 ± 0.4% kPa−1 in the linear region, a pressure hysteresis of 4.1 ± 1.5% of full scale, and, in a representative sensor, a variation within ±2.1% of the mean over five loading–unloading cycles. Between 30 and 45 °C, measured in the solution at the stent, the step means of ΔR/R0 remained within 0.10% of zero (n = 4 sensors). In recirculating model bile inoculated with Escherichia coli, the sensor output in the run with a complete recording began to change after approximately 400 min and reached the plateau of the response after approximately 840 min, whereas in sterile model bile the mean output of four sensors remained within 0.15% of baseline over 14 h. In a porcine biliary stricture model, the device was deployed endoscopically in one animal, its wires were externalized through a percutaneous endoscopic gastrostomy tube, and the monitor, zeroed in vivo after connection, displayed a change in readout after start-up, which was not verified against a reference pressure. The connection wire broke inside the percutaneous endoscopic gastrostomy tube within 1 week and had also broken at several steps of the procedure when it was pulled; in bench tests, it fractured at a monotonic tensile load of 1.878 N. As an initial feasibility study toward clinical application, these results indicate that sensor integration, endoscopic delivery, wire routing, and external readout can be achieved, identify the procedural steps at which the sensor and its wires failed, and define the extension of the measurement range and the protection of the connection that are required before occlusion monitoring can be evaluated in vivo.

Authors

Institutions

Publication Details

Journal
Biosensors
Published
2026-11-24
DOI
https://doi.org/10.3390/bios16120683
Primary Topic
Gallbladder and Bile Duct Disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A Wired Pressure-Sensing Smart Biliary Stent: Bench Characterization and Feasibility in a Porcine Model

Dong Chan Joo, Seung Min Hong, Kiyoun Yi, Sung Yong Han et al.
1 links
Biosensors
Gallbladder and Bile Duct Disorders
article

A Wired Pressure-Sensing Smart Biliary Stent: Bench Characterization and Feasibility in a Porcine Model

Dong Chan Joo, Seung Min Hong, Kiyoun Yi, Sung Yong Han, Young Woo Kwon, Jonghyun Lee, Jae Won Lee
article en

Abstract

Biliary stents restore drainage in malignant biliary obstruction, but recurrent dysfunction creates a need to monitor the patency of the stent. We developed a pressure-sensing smart biliary stent by integrating a flexible contact-resistive sensor, comprising opposing Au mesh electrodes on liquid crystal polymer films, into a commercial self-expandable metal stent connected to an external monitoring system. Under loading through a compliant pad matched to the 1 mm × 2 mm sensing window (n = 4 sensors), the normalized resistance decreased approximately linearly up to 2.5 kPa and approached a plateau at approximately 3.0 kPa, with a sensitivity of −37.0 ± 0.4% kPa−1 in the linear region, a pressure hysteresis of 4.1 ± 1.5% of full scale, and, in a representative sensor, a variation within ±2.1% of the mean over five loading–unloading cycles. Between 30 and 45 °C, measured in the solution at the stent, the step means of ΔR/R0 remained within 0.10% of zero (n = 4 sensors). In recirculating model bile inoculated with Escherichia coli, the sensor output in the run with a complete recording began to change after approximately 400 min and reached the plateau of the response after approximately 840 min, whereas in sterile model bile the mean output of four sensors remained within 0.15% of baseline over 14 h. In a porcine biliary stricture model, the device was deployed endoscopically in one animal, its wires were externalized through a percutaneous endoscopic gastrostomy tube, and the monitor, zeroed in vivo after connection, displayed a change in readout after start-up, which was not verified against a reference pressure. The connection wire broke inside the percutaneous endoscopic gastrostomy tube within 1 week and had also broken at several steps of the procedure when it was pulled; in bench tests, it fractured at a monotonic tensile load of 1.878 N. As an initial feasibility study toward clinical application, these results indicate that sensor integration, endoscopic delivery, wire routing, and external readout can be achieved, identify the procedural steps at which the sensor and its wires failed, and define the extension of the measurement range and the protection of the connection that are required before occlusion monitoring can be evaluated in vivo.

BiosensorsVol. 16(12)
Pusan National University Hospital (KR), Ulsan National Institute of Science and Technology (KR), Pusan National University (KR)
National Research Foundation, Ministry of Science and ICT, South Korea
Good health and well-being
Gallbladder and Bile Duct Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon