Hour-long live-cell TIRF imaging of GFP-talin at ensemble and single-molecule level with low-duty-cycle picosecond supercontinuum excitation
Photobleaching and phototoxicity limit how long living cells can be followed by total internal reflection fluorescence (TIRF) microscopy. We describe an objective-type TIRF microscope illuminated by a picosecond supercontinuum laser spectrally windowed to 20-nm bands, pulse-picked to 1 MHz and delivered through two TIRF illuminators, one per polarisation direction. Single GFP-talin molecules in focal adhesions of living cells were detected throughout uninterrupted recordings of 10-31 min, and drift-corrected, time-stamped localisations resolved adhesion motion of tens of nanometres per minute. In diffraction-limited time-lapse at one frame per second, adhesion brightness, adhesion area, cell footprint and adhesion motility were unchanged after an hour. Under continuous-wave excitation on a comparable commercial TIRF microscope, the same schedule preserved cell behaviour at a low laser setting but adhesions dimmed, and higher settings bleached and retracted cells within a minute of exposure. We provide the datasets and open ImageJ tools for drift correction and localisation.
Authors
- Oddmund Bakke (ORCID: https://orcid.org/0000-0003-4843-7626)
- Michael P. Sheetz (ORCID: https://orcid.org/0000-0001-8747-3430)
- Felix Margadant (ORCID: https://orcid.org/0000-0002-6729-6155)
- Xian Hu
Institutions
- National University of Singapore (SG)
- University of Oslo (NO)
- Mechanobiology Institute (SG)
Publication Details
- Journal
- bioRxiv (Cold Spring Harbor Laboratory)
- Published
- 2026-11-24
- DOI
- https://doi.org/10.64898/2026.11.23.768601
- Primary Topic
- Advanced Fluorescence Microscopy Techniques
- Type
- preprint