Capillary-Driven Surface-Enhanced Raman Scattering Hollow Microneedle Patch for Glucose and pH Sensing in Interstitial Fluids

Minimally invasive monitoring of interstitial fluid (ISF) biomarkers requires reliable sample collection and molecular analysis without external pumping. Here, we present an integrated, capillary-driven hollow microneedle (HMN) patch that combines passive fluid collection, nitrocellulose (NC) filtration, and surface-enhanced Raman scattering (SERS) sensing of glucose and pH. A three-dimensionally printed hydrophilic HMN structure enables skin access and capillary transport, while a flexible and optically transparent substrate permits Raman interrogation from the backside. The SERS substrate was fabricated by Langmuir-Blodgett assembly of four layers of silver-coated gold nanorods (GNR@Ag) on polydimethylsiloxane to create dense plasmonic hotspots. The substrate exhibited good spatial uniformity (RSD = 8.4%) and interbatch reproducibility (RSD = 4.8%). Functionalization with 4-mercaptophenylboronic acid and 4-mercaptobenzoic acid generated glucose- and pH-responsive signals, respectively. Glucose-dependent SERS responses were calibrated up to 30 mM (R2 = 0.97), and pH sensing was linear across pH 4-8 (R2 = 0.98). The NC filter reduced protein-related interference, and low-power excitation (0.25 mW) maintained spectral stability for 200 s. Crosstalk experiments revealed a pronounced pH dependence in glucose sensing but comparatively limited glucose interference with the pH readout. In mice, the patch exhibited fluid uptake after a 20 min collection period and glucose-dose-dependent SERS responses following intravenous administration. Ex vivo tests using excised mouse skin conditioned on agarose media at different pH values further demonstrated pH-responsive sensing under tissue-contact conditions. These results support the potential of a capillary-driven SERS HMN platform for minimally invasive glucose and pH sensing in ISF.

Authors

Institutions

Publication Details

Journal
ACS Sensors
Published
2026-11-24
DOI
https://doi.org/10.1021/acssensors.6c02862
Primary Topic
Gold and Silver Nanoparticles Synthesis and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Capillary-Driven Surface-Enhanced Raman Scattering Hollow Microneedle Patch for Glucose and pH Sensing in Interstitial Fluids

Young‐Seong Kim, Kyuil Chung, Joong Yeon Lim, Jeon Woong Kang et al.
6 links
ACS Sensors
Gold and Silver Nanoparticles Synthesis and Applications
article

Capillary-Driven Surface-Enhanced Raman Scattering Hollow Microneedle Patch for Glucose and pH Sensing in Interstitial Fluids

Young‐Seong Kim, Kyuil Chung, Joong Yeon Lim, Jeon Woong Kang, Minwook Park
article en

Abstract

Minimally invasive monitoring of interstitial fluid (ISF) biomarkers requires reliable sample collection and molecular analysis without external pumping. Here, we present an integrated, capillary-driven hollow microneedle (HMN) patch that combines passive fluid collection, nitrocellulose (NC) filtration, and surface-enhanced Raman scattering (SERS) sensing of glucose and pH. A three-dimensionally printed hydrophilic HMN structure enables skin access and capillary transport, while a flexible and optically transparent substrate permits Raman interrogation from the backside. The SERS substrate was fabricated by Langmuir-Blodgett assembly of four layers of silver-coated gold nanorods (GNR@Ag) on polydimethylsiloxane to create dense plasmonic hotspots. The substrate exhibited good spatial uniformity (RSD = 8.4%) and interbatch reproducibility (RSD = 4.8%). Functionalization with 4-mercaptophenylboronic acid and 4-mercaptobenzoic acid generated glucose- and pH-responsive signals, respectively. Glucose-dependent SERS responses were calibrated up to 30 mM (R2 = 0.97), and pH sensing was linear across pH 4-8 (R2 = 0.98). The NC filter reduced protein-related interference, and low-power excitation (0.25 mW) maintained spectral stability for 200 s. Crosstalk experiments revealed a pronounced pH dependence in glucose sensing but comparatively limited glucose interference with the pH readout. In mice, the patch exhibited fluid uptake after a 20 min collection period and glucose-dose-dependent SERS responses following intravenous administration. Ex vivo tests using excised mouse skin conditioned on agarose media at different pH values further demonstrated pH-responsive sensing under tissue-contact conditions. These results support the potential of a capillary-driven SERS HMN platform for minimally invasive glucose and pH sensing in ISF.

ACS Sensors
Dongguk University (KR), Massachusetts Institute of Technology (US)
Dongguk University, Ministry of Trade, Industry and Energy, Korea Institute for Advancement of Technology, Ministry of Science and ICT, South Korea, Commercializations Promotion Agency for R and D Outcomes
Good health and well-being
Gold and Silver Nanoparticles Synthesis and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon