Verification of Manufacturer Provided Biological Reference Intervals in a Healthy Adult Cohort: A Pragmatic Laboratory Study

Background Biological reference intervals (RIs) supplied by in-vitro diagnostic manufacturers are typically established in carefully selected populations and under tightly controlled pre-analytical conditions. Before their routine adoption, international guidelines mandate that each laboratory verifies these intervals in a local reference population to ensure clinical validity. Despite the importance of this process, real-world data describing verification outcomes across multiple routine chemistry panels remain scarce. Methods We conducted a single-centre, cross-sectional verification study in 63 apparently healthy adults (24 men, 39 women; mean ± SD age 40.4 ± 12.0 years) recruited according to CLSI EP28-A3c criteria. Fasting blood was analysed on a calibrated Beckman Coulter AU5800 chemistry analyser for lipid, kidney (KFT) and liver (LFT) panels. Manufacturer RIs were assessed by a binomial test (≤ 2 results outside limits = verification pass). Sex and age effects were evaluated with Student’s t-test or one-way ANOVA as appropriate. Two visual analytics—(i) creatinine distribution by sex and (ii) eGFR versus age—were generated to illustrate key findings. Results All 21 KFT and 22 LFT measurements (100 %) fell within the stated RIs. For the lipid panel, 92–100 % of results were within limits, except for the cholesterol/HDL ratio (36 %), prompting further review. Sex differences were significant for HDL-C (higher in women, p = 0.014) and serum creatinine (higher in men, p = 0.004). Age had no material impact on most analytes, although eGFR declined across age tertiles (p = 0.001). Overall, manufacturer RIs were verified for 26 of 27 routine chemistry tests (96 %). Conclusion Using a streamlined protocol that required fewer than 70 participants, we confirmed the applicability of the manufacturer’s chemistry RIs to our local adult population, with the exception of the cholesterol/HDL ratio. Laboratories serving demographically similar populations may leverage these data to prioritise verification resources.

Authors

Publication Details

Journal
International Journal of Drug Delivery Technology
Published
2026-11-24
DOI
https://doi.org/10.25258/ijddt.16.82s.38
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Verification of Manufacturer Provided Biological Reference Intervals in a Healthy Adult Cohort: A Pragmatic Laboratory Study

Shivani Sharma, Onkar Nath Tiwari, Neetesh Kumar Yadav
1 links
International Journal of Drug Delivery Technology
Clinical Laboratory Practices and Quality Control
article

Verification of Manufacturer Provided Biological Reference Intervals in a Healthy Adult Cohort: A Pragmatic Laboratory Study

Shivani Sharma, Onkar Nath Tiwari, Neetesh Kumar Yadav
article en

Abstract

Background Biological reference intervals (RIs) supplied by in-vitro diagnostic manufacturers are typically established in carefully selected populations and under tightly controlled pre-analytical conditions. Before their routine adoption, international guidelines mandate that each laboratory verifies these intervals in a local reference population to ensure clinical validity. Despite the importance of this process, real-world data describing verification outcomes across multiple routine chemistry panels remain scarce. Methods We conducted a single-centre, cross-sectional verification study in 63 apparently healthy adults (24 men, 39 women; mean ± SD age 40.4 ± 12.0 years) recruited according to CLSI EP28-A3c criteria. Fasting blood was analysed on a calibrated Beckman Coulter AU5800 chemistry analyser for lipid, kidney (KFT) and liver (LFT) panels. Manufacturer RIs were assessed by a binomial test (≤ 2 results outside limits = verification pass). Sex and age effects were evaluated with Student’s t-test or one-way ANOVA as appropriate. Two visual analytics—(i) creatinine distribution by sex and (ii) eGFR versus age—were generated to illustrate key findings. Results All 21 KFT and 22 LFT measurements (100 %) fell within the stated RIs. For the lipid panel, 92–100 % of results were within limits, except for the cholesterol/HDL ratio (36 %), prompting further review. Sex differences were significant for HDL-C (higher in women, p = 0.014) and serum creatinine (higher in men, p = 0.004). Age had no material impact on most analytes, although eGFR declined across age tertiles (p = 0.001). Overall, manufacturer RIs were verified for 26 of 27 routine chemistry tests (96 %). Conclusion Using a streamlined protocol that required fewer than 70 participants, we confirmed the applicability of the manufacturer’s chemistry RIs to our local adult population, with the exception of the cholesterol/HDL ratio. Laboratories serving demographically similar populations may leverage these data to prioritise verification resources.

International Journal of Drug Delivery TechnologyVol. 16(82s)
Good health and well-being
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon