Markerless augmented reality guidance for mastoidectomy: a phantom feasibility study using a head-mounted display

Introduction Cochlear implantation requires precise mastoid drilling to avoid facial nerve injury. Markerless augmented reality (AR) provides non-disruptive guidance by overlaying 3D anatomy directly onto the surgical view. Methods We developed a markerless AR system using a Magic Leap 2 head-mounted display. Preoperative CT models are coloured for RGB matching. In the proposed pipeline, the Segment Anything Model 3 and FoundationPose automatically achieve six-degree-of-freedom (6-DOF) registration of the anatomy without fiducials or external trackers (instrument tracking remains marker-based). Registration was evaluated on the intact bone surface before drilling. Results Across nine sessions spanning three working distances and three viewing angles, the mean lateral target registration error was 1.56 ± 1.01 mm in magnitude (root-mean-square 1.82 mm), with a rotational error of 0.86° ± 0.28°. Depth deviation averaged 6.90 ± 3.41 mm and fell steadily as working distance increased, while no statistically significant effect of viewing angle was detected. Measured directly in image space, without HMD-based visual alignment, the in-plane registration error was about 2 mm and nearly constant across all conditions; this analysis does not estimate depth, so the source of the depth error remains unresolved. Mean measured software-pipeline latency was 153 ms, excluding sensor-to-application delay and display buffering. A single cadaveric specimen provided a preliminary demonstration that registration is maintained on natural bone under partial occlusion. Discussion This markerless system, using a single wearable sensing and display device with off-board processing, demonstrates low-millimetre lateral accuracy for preoperative registration. Despite the observed depth error, it offers a non-invasive, ergonomic alternative to traditional trackers for projecting vital otologic structures; clinical evaluation and continuous registration through bone removal are the next steps towards intraoperative navigation.

Authors

Institutions

Publication Details

Journal
Frontiers in Virtual Reality
Published
2026-11-24
DOI
https://doi.org/10.3389/frvir.2026.1887382
Primary Topic
Augmented Reality Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Markerless augmented reality guidance for mastoidectomy: a phantom feasibility study using a head-mounted display

Nicolas Verhaert, Tristan Putzeys, Carlos Rodriguez‐Guerrero, Yang Li
Frontiers in Virtual Reality
Augmented Reality Applications
article

Markerless augmented reality guidance for mastoidectomy: a phantom feasibility study using a head-mounted display

Nicolas Verhaert, Tristan Putzeys, Carlos Rodriguez‐Guerrero, Yang Li
article en

Abstract

Introduction Cochlear implantation requires precise mastoid drilling to avoid facial nerve injury. Markerless augmented reality (AR) provides non-disruptive guidance by overlaying 3D anatomy directly onto the surgical view. Methods We developed a markerless AR system using a Magic Leap 2 head-mounted display. Preoperative CT models are coloured for RGB matching. In the proposed pipeline, the Segment Anything Model 3 and FoundationPose automatically achieve six-degree-of-freedom (6-DOF) registration of the anatomy without fiducials or external trackers (instrument tracking remains marker-based). Registration was evaluated on the intact bone surface before drilling. Results Across nine sessions spanning three working distances and three viewing angles, the mean lateral target registration error was 1.56 ± 1.01 mm in magnitude (root-mean-square 1.82 mm), with a rotational error of 0.86° ± 0.28°. Depth deviation averaged 6.90 ± 3.41 mm and fell steadily as working distance increased, while no statistically significant effect of viewing angle was detected. Measured directly in image space, without HMD-based visual alignment, the in-plane registration error was about 2 mm and nearly constant across all conditions; this analysis does not estimate depth, so the source of the depth error remains unresolved. Mean measured software-pipeline latency was 153 ms, excluding sensor-to-application delay and display buffering. A single cadaveric specimen provided a preliminary demonstration that registration is maintained on natural bone under partial occlusion. Discussion This markerless system, using a single wearable sensing and display device with off-board processing, demonstrates low-millimetre lateral accuracy for preoperative registration. Despite the observed depth error, it offers a non-invasive, ergonomic alternative to traditional trackers for projecting vital otologic structures; clinical evaluation and continuous registration through bone removal are the next steps towards intraoperative navigation.

Frontiers in Virtual RealityVol. 7
Universitair Ziekenhuis Leuven (BE), KU Leuven (BE)
Fonds Wetenschappelijk Onderzoek, KU Leuven
Good health and well-being
Augmented Reality Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon