Case report: When quality control goes wrong — an incidental finding of Hb E compound heterozygosity with Filipino β0-thalassemia deletion in a commercial EQA material mistaken as Hb E homozygote

Objective Hb E homozygosity and compound heterozygosity for Hb E and β 0 -thalassemia show highly similar profiles on hemoglobin analysis, creating a diagnostic pitfall. We aimed to resolve an incidental genotype discrepancy in an external quality assessment (EQA) sample originally labeled as Hb E homozygote. Methods Routine thalassemia genetic testing covered 26 mutations commonly found in the Chinese population. Multiplex ligation-dependent probe amplification (MLPA) was used to screen for potential large deletions or duplications. Third-generation sequencing (TGS) was performed to define the exact location of the large deletion, and the breakpoints it identified were subsequently validated by Gap-PCR and Sanger sequencing. Results The sample showed Hb A 0%, Hb A 2 0%, Hb F 43.6%, Hb E 56.4%, Hb 8.3 g/dL, and MCV 64.5 fL. PCR-reverse dot blot hybridization indicated Hb E homozygosity. MLPA revealed heterozygous deletions across multiple probes, consistent with the Filipino deletion. TGS identified a 118,475 bp deletion at chr11:5112883-5231358del. Gap-PCR and Sanger sequencing confirmed a 118,475 bp deletion. The supplier initially disputed these findings and reported Hb E homozygosity without deletion by NGS. After Nanopore-based TGS rechecking, the Filipino deletion was detected, and cell line trace-back confirmed compound heterozygosity for Hb E and the Filipino β 0 -thalassemia deletion. Conclusion EQA samples with complex or overlapping hemoglobin phenotypes require comprehensive molecular validation before distribution. Reliance on a single assay or manufacturer-provided assignment can lead to erroneous EQA evaluation.

Authors

Institutions

Publication Details

Journal
Frontiers in Genetics
Published
2026-11-24
DOI
https://doi.org/10.3389/fgene.2026.2013668
Primary Topic
Hemoglobinopathies and Related Disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Case report: When quality control goes wrong — an incidental finding of Hb E compound heterozygosity with Filipino β0-thalassemia deletion in a commercial EQA material mistaken as Hb E homozygote

李友琼, Lihong Zheng, Juan Tang, Liang Liang et al.
3 links
Frontiers in Genetics
Hemoglobinopathies and Related Disorders
article

Case report: When quality control goes wrong — an incidental finding of Hb E compound heterozygosity with Filipino β0-thalassemia deletion in a commercial EQA material mistaken as Hb E homozygote

李友琼, Lihong Zheng, Juan Tang, Liang Liang, Yun Yu, Xiong Huang
article en

Abstract

Objective Hb E homozygosity and compound heterozygosity for Hb E and β 0 -thalassemia show highly similar profiles on hemoglobin analysis, creating a diagnostic pitfall. We aimed to resolve an incidental genotype discrepancy in an external quality assessment (EQA) sample originally labeled as Hb E homozygote. Methods Routine thalassemia genetic testing covered 26 mutations commonly found in the Chinese population. Multiplex ligation-dependent probe amplification (MLPA) was used to screen for potential large deletions or duplications. Third-generation sequencing (TGS) was performed to define the exact location of the large deletion, and the breakpoints it identified were subsequently validated by Gap-PCR and Sanger sequencing. Results The sample showed Hb A 0%, Hb A 2 0%, Hb F 43.6%, Hb E 56.4%, Hb 8.3 g/dL, and MCV 64.5 fL. PCR-reverse dot blot hybridization indicated Hb E homozygosity. MLPA revealed heterozygous deletions across multiple probes, consistent with the Filipino deletion. TGS identified a 118,475 bp deletion at chr11:5112883-5231358del. Gap-PCR and Sanger sequencing confirmed a 118,475 bp deletion. The supplier initially disputed these findings and reported Hb E homozygosity without deletion by NGS. After Nanopore-based TGS rechecking, the Filipino deletion was detected, and cell line trace-back confirmed compound heterozygosity for Hb E and the Filipino β 0 -thalassemia deletion. Conclusion EQA samples with complex or overlapping hemoglobin phenotypes require comprehensive molecular validation before distribution. Reliance on a single assay or manufacturer-provided assignment can lead to erroneous EQA evaluation.

Frontiers in GeneticsVol. 17
Jinan University (CN), The People's Hospital of Guangxi Zhuang Autonomous Region (CN)
Guangxi Key Research and Development Program
Good health and well-being
Hemoglobinopathies and Related Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon