Factorial Sensitivity of Prosthetic Reaction Forces to Mass, Inertia, and Alignment

Background: Prosthesis mass, inertia, and alignment can alter mechanical loading during gait, but their relative influence may depend on how loading is quantified. This study examined their effects on absolute and relative joint reaction force outcomes in a transfemoral prosthesis model. Methods: A deterministic 3 × 3 × 3 full-factorial design generated 27 configurations from three mass, inertia, and mediolateral alignment levels, driven by identical prescribed kinematics and external loads in OpenSim. Resultant reactions were evaluated at the socket–prosthesis junction and prosthetic ankle using peak resultant reaction force, force–time integral (FTI), socket-to-ankle FTI ratio, and a secondary Peak Reaction Ratio. Factorial decomposition and range-normalized sensitivity analysis quantified parameter effects. Results: Within this deterministic computational framework, prosthesis mass was the dominant contributor to variation in the absolute loading outcomes, representing 90.28% and 94.06% of socket and ankle peak-force variation and 94.26% and 95.63% of their FTI variation, respectively. The FTI ratio showed a more distributed sensitivity to mass (53.55%), inertia (28.90%), and alignment (13.41%), whereas the Peak Reaction Ratio was primarily influenced by inertia (51.99%) and mass (45.85%). Peak timing separation ranged from 2.7% to 48.0% of the normalized gait cycle. Conclusions: Within the tested parameter ranges and fixed reference gait, prosthesis mass was the dominant contributor to absolute reaction-force variation in the rigid-body model, whereas relative socket-to-ankle outcomes showed greater contributions from inertia and alignment. These findings represent model-level sensitivities and should not be interpreted as direct clinical fitting recommendations.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-11-24
DOI
https://doi.org/10.3390/bioengineering13121426
Primary Topic
Prosthetics and Rehabilitation Robotics
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Factorial Sensitivity of Prosthetic Reaction Forces to Mass, Inertia, and Alignment

Linas Puodžiukynas, Aurelijus Domeika, Arūnas Žvironas, Donatas Daublys et al.
1 links
Bioengineering
Prosthetics and Rehabilitation Robotics
article

Factorial Sensitivity of Prosthetic Reaction Forces to Mass, Inertia, and Alignment

Linas Puodžiukynas, Aurelijus Domeika, Arūnas Žvironas, Donatas Daublys, Danas Jackunas
article en

Abstract

Background: Prosthesis mass, inertia, and alignment can alter mechanical loading during gait, but their relative influence may depend on how loading is quantified. This study examined their effects on absolute and relative joint reaction force outcomes in a transfemoral prosthesis model. Methods: A deterministic 3 × 3 × 3 full-factorial design generated 27 configurations from three mass, inertia, and mediolateral alignment levels, driven by identical prescribed kinematics and external loads in OpenSim. Resultant reactions were evaluated at the socket–prosthesis junction and prosthetic ankle using peak resultant reaction force, force–time integral (FTI), socket-to-ankle FTI ratio, and a secondary Peak Reaction Ratio. Factorial decomposition and range-normalized sensitivity analysis quantified parameter effects. Results: Within this deterministic computational framework, prosthesis mass was the dominant contributor to variation in the absolute loading outcomes, representing 90.28% and 94.06% of socket and ankle peak-force variation and 94.26% and 95.63% of their FTI variation, respectively. The FTI ratio showed a more distributed sensitivity to mass (53.55%), inertia (28.90%), and alignment (13.41%), whereas the Peak Reaction Ratio was primarily influenced by inertia (51.99%) and mass (45.85%). Peak timing separation ranged from 2.7% to 48.0% of the normalized gait cycle. Conclusions: Within the tested parameter ranges and fixed reference gait, prosthesis mass was the dominant contributor to absolute reaction-force variation in the rigid-body model, whereas relative socket-to-ankle outcomes showed greater contributions from inertia and alignment. These findings represent model-level sensitivities and should not be interpreted as direct clinical fitting recommendations.

BioengineeringVol. 13(12)
Kaunas University of Technology (LT)
Lietuvos Mokslo Taryba
Prosthetics and Rehabilitation Robotics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Factorial Sensitivity of Prosthetic Reaction Forces to Mass, Inertia, and Alignment — Linas Puodžiukynas, Aurelijus Domeika, et al. · Bioengineering (2026) | TGRS Research Map | TGRS