Are Distance Stereoacuity Tests Interchangeable? Effects of Measurement Range and Disparity Resolution on Agreement

Purpose. To assess the repeatability and interchangeability of four distance stereoacuity methods and examine how differences in measurement range and disparity resolution affect between-method comparisons. Methods. Distance stereoacuity was assessed at 3 m in 65 young adults using COI, Eyenext, Distance Randot, and frontal Senaptec, in randomised order, with retesting after 10 minutes. Valid measurable thresholds were analysed on a log₁₀(seconds of arc) scale. Repeatability was evaluated using paired Wilcoxon tests, intraclass correlation coefficients (ICC(A,1)), concordance correlation coefficients (CCC), and Bland–Altman limits of agreement. Between-method differences and agreement were evaluated in 61 participants with complete initial measurements, first on the native scales, then after capping thresholds to 60–200 arcsec and finally after harmonising to three ordinal levels (60, 100, and 200 arcsec). Results. Median initial thresholds were 60, 80, 100, and 41 arcsec for COI, Eyenext, Distance Randot, and Senaptec, respectively. The methods differed on their native scales (Friedman χ²(3) = 28.007; p < 0.001; Kendall’s W = 0.153), with significant Holm-adjusted differences for Distance Randot versus COI and Senaptec. The global difference was smaller after capping to 60–200 arcsec (χ²(3) = 9.348; p = 0.025; W = 0.051), with no significant adjusted pairwise differences, and was not statistically significant after three-level harmonisation (χ²(3) = 6.371; p = 0.095; W = 0.035). Nevertheless, individual agreement remained limited. COI–Eyenext had the highest native-scale agreement (ICC = 0.665; CCC = 0.661) and three-level linear weighted kappa (κ = 0.477), whereas the remaining pairwise ICCs were 0.114–0.254 and linear weighted kappas 0.103–0.196. No method showed a statistically significant short-term test–retest shift after Holm correction; repeatability ICCs ranged from 0.527 to 0.636, with wide limits of agreement. Conclusions. Harmonising measurement scales attenuated group-level differences but did not establish equivalence or adequate agreement between individuals. These four methods should not be used interchangeably; the same instrument and protocol should be retained for longitudinal assessment.

Authors

Institutions

Publication Details

Journal
Research Square
Published
2026-11-24
DOI
https://doi.org/10.21203/rs.3.rs-11707158/v1
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
preprint

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Are Distance Stereoacuity Tests Interchangeable? Effects of Measurement Range and Disparity Resolution on Agreement

Miguel Coelho, João Pedro Jorge, Jorge Jorge¹
1 links
Research Square
Ophthalmology and Visual Impairment Studies
preprint

Are Distance Stereoacuity Tests Interchangeable? Effects of Measurement Range and Disparity Resolution on Agreement

Miguel Coelho, João Pedro Jorge, Jorge Jorge¹
preprint en

Abstract

Purpose. To assess the repeatability and interchangeability of four distance stereoacuity methods and examine how differences in measurement range and disparity resolution affect between-method comparisons. Methods. Distance stereoacuity was assessed at 3 m in 65 young adults using COI, Eyenext, Distance Randot, and frontal Senaptec, in randomised order, with retesting after 10 minutes. Valid measurable thresholds were analysed on a log₁₀(seconds of arc) scale. Repeatability was evaluated using paired Wilcoxon tests, intraclass correlation coefficients (ICC(A,1)), concordance correlation coefficients (CCC), and Bland–Altman limits of agreement. Between-method differences and agreement were evaluated in 61 participants with complete initial measurements, first on the native scales, then after capping thresholds to 60–200 arcsec and finally after harmonising to three ordinal levels (60, 100, and 200 arcsec). Results. Median initial thresholds were 60, 80, 100, and 41 arcsec for COI, Eyenext, Distance Randot, and Senaptec, respectively. The methods differed on their native scales (Friedman χ²(3) = 28.007; p < 0.001; Kendall’s W = 0.153), with significant Holm-adjusted differences for Distance Randot versus COI and Senaptec. The global difference was smaller after capping to 60–200 arcsec (χ²(3) = 9.348; p = 0.025; W = 0.051), with no significant adjusted pairwise differences, and was not statistically significant after three-level harmonisation (χ²(3) = 6.371; p = 0.095; W = 0.035). Nevertheless, individual agreement remained limited. COI–Eyenext had the highest native-scale agreement (ICC = 0.665; CCC = 0.661) and three-level linear weighted kappa (κ = 0.477), whereas the remaining pairwise ICCs were 0.114–0.254 and linear weighted kappas 0.103–0.196. No method showed a statistically significant short-term test–retest shift after Holm correction; repeatability ICCs ranged from 0.527 to 0.636, with wide limits of agreement. Conclusions. Harmonising measurement scales attenuated group-level differences but did not establish equivalence or adequate agreement between individuals. These four methods should not be used interchangeably; the same instrument and protocol should be retained for longitudinal assessment.

Research Square
Universidade Nova de Lisboa (PT), University of Minho (PT)
Fundação para a Ciência e a Tecnologia
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon