Robotic Venipuncture with Sequential Needle-Tip Realignment and Probe-Induced Vessel Displacement Compensation

Robotic venipuncture systems can accurately localize vascular targets using near-infrared (NIR) and ultrasound (US) imaging. However, image-based localization does not guarantee accurate physical needle placement. Residual errors arising from sensor-to-robot registration, mechanical assembly, and needle mounting can cause misalignment of the physical needle tip, while contact pressure of the US probe on soft tissue can displace the vessel from its initially estimated position. This study presents a sequential error-compensation method that addresses residual positioning errors before needle insertion, followed by US-based vessel-arrival detection and bounded stopping during insertion. Monocular NIR imaging is used to determine the vessel centerline and planar target on the skin surface, while US imaging provides the corresponding vessel depth coordinate for three-dimensional target localization. Probe-induced vessel displacement is incorporated into the insertion geometry to determine the required needle angle and insertion length. After insertion-path planning, the physical needle tip is redetected in the NIR image and realigned with the planar vessel target. The final insertion angle is then established and needle advancement is performed using the independent linear insertion axis. By compensating these error sources prior to penetration, the subsequent insertion stage is simplified to linear needle advancement without continuous high-DOF steering. During insertion, US-based vessel-arrival detection is combined with an independent maximum insertion-length constraint to provide bounded stopping. In phantom experiments, NIR-based needle-tip alignment increased the pre-insertion alignment success rate from 14.0% to 94.0%, while probe-induced vessel displacement compensation increased insertion success from 58.0% to 98.0%. The complete autonomous procedure achieved a 96.7% success rate. These results demonstrate that sequential compensation of probe-induced vessel displacement and physical needle-tip misalignment enables accurate and repeatable robotic venipuncture using a comparatively simple insertion architecture without continuous multi-DOF needle steering.

Authors

Institutions

Publication Details

Journal
Actuators
Published
2026-11-23
DOI
https://doi.org/10.3390/act15120619
Primary Topic
Soft Robotics and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Robotic Venipuncture with Sequential Needle-Tip Realignment and Probe-Induced Vessel Displacement Compensation

Kiwon Yeom, Seungyeon Shin
Actuators
Soft Robotics and Applications
article

Robotic Venipuncture with Sequential Needle-Tip Realignment and Probe-Induced Vessel Displacement Compensation

Kiwon Yeom, Seungyeon Shin
article en

Abstract

Robotic venipuncture systems can accurately localize vascular targets using near-infrared (NIR) and ultrasound (US) imaging. However, image-based localization does not guarantee accurate physical needle placement. Residual errors arising from sensor-to-robot registration, mechanical assembly, and needle mounting can cause misalignment of the physical needle tip, while contact pressure of the US probe on soft tissue can displace the vessel from its initially estimated position. This study presents a sequential error-compensation method that addresses residual positioning errors before needle insertion, followed by US-based vessel-arrival detection and bounded stopping during insertion. Monocular NIR imaging is used to determine the vessel centerline and planar target on the skin surface, while US imaging provides the corresponding vessel depth coordinate for three-dimensional target localization. Probe-induced vessel displacement is incorporated into the insertion geometry to determine the required needle angle and insertion length. After insertion-path planning, the physical needle tip is redetected in the NIR image and realigned with the planar vessel target. The final insertion angle is then established and needle advancement is performed using the independent linear insertion axis. By compensating these error sources prior to penetration, the subsequent insertion stage is simplified to linear needle advancement without continuous high-DOF steering. During insertion, US-based vessel-arrival detection is combined with an independent maximum insertion-length constraint to provide bounded stopping. In phantom experiments, NIR-based needle-tip alignment increased the pre-insertion alignment success rate from 14.0% to 94.0%, while probe-induced vessel displacement compensation increased insertion success from 58.0% to 98.0%. The complete autonomous procedure achieved a 96.7% success rate. These results demonstrate that sequential compensation of probe-induced vessel displacement and physical needle-tip misalignment enables accurate and repeatable robotic venipuncture using a comparatively simple insertion architecture without continuous multi-DOF needle steering.

ActuatorsVol. 15(12)
Sangmyung University (KR)
Soft Robotics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon