Femoral Surface Stress Redistribution After Cemented Stem Insertion With Distal Lateral Stress Concentration in an Osteoporotic Composite Femur Model

Introduction Previous thermoelastic studies have shown that cemented stem implantation shifts femoral stress distally, but regional averages provide limited information about the location of individual stress peaks. This study aimed to characterize femoral surface stress redistribution after implantation of the Trilliance polished taper-slip cemented stem in normal and osteoporotic composite femora. Materials and Methods Three normal and three osteoporotic composite femora were tested before and after stem implantation. Size 10 and size 16 high-offset stems were used in the normal and osteoporotic models, respectively. Cyclic compressive loads of 100–1900 N and 100–1000 N, respectively, were applied at 5 Hz. Thermoelastic stress analysis measured changes in the sum of the surface principal stresses. Longitudinal stress profiles along the medial and lateral surfaces were evaluated descriptively. For each implanted specimen, the location of peak lateral stress was recorded relative to the stem tip. Results In both models, proximal surface stress decreased after implantation, and the peaks of the mean medial and lateral stress profiles shifted distally. In the normal model, distal stress changed little, and individual lateral peaks were located 50–82 mm proximal to the stem tip. In the osteoporotic model, distal stress increased on both surfaces. All three lateral peaks were within 16 mm of the stem tip, including one 16 mm distal to the tip. Conclusions Under the conditions tested, the osteoporotic specimens showed increased distal surface stress after implantation, with the lateral stress peak consistently located near the stem tip. Differences in stem size and loading conditions precluded attribution of this pattern to bone properties alone.

Authors

Institutions

Publication Details

Journal
Research Square
Published
2026-11-23
DOI
https://doi.org/10.21203/rs.3.rs-11672958/v1
Primary Topic
Orthopaedic implants and arthroplasty
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Femoral Surface Stress Redistribution After Cemented Stem Insertion With Distal Lateral Stress Concentration in an Osteoporotic Composite Femur Model

Ryunosuke Watanabe, Yoshihisa Harada, Yasuhiro Homma, Tomohiro Yoshizawa et al.
Research Square
Orthopaedic implants and arthroplasty
preprint

Femoral Surface Stress Redistribution After Cemented Stem Insertion With Distal Lateral Stress Concentration in an Osteoporotic Composite Femur Model

Ryunosuke Watanabe, Yoshihisa Harada, Yasuhiro Homma, Tomohiro Yoshizawa, Hajime Mishima, Shota Yasunaga
preprint en

Abstract

Introduction Previous thermoelastic studies have shown that cemented stem implantation shifts femoral stress distally, but regional averages provide limited information about the location of individual stress peaks. This study aimed to characterize femoral surface stress redistribution after implantation of the Trilliance polished taper-slip cemented stem in normal and osteoporotic composite femora. Materials and Methods Three normal and three osteoporotic composite femora were tested before and after stem implantation. Size 10 and size 16 high-offset stems were used in the normal and osteoporotic models, respectively. Cyclic compressive loads of 100–1900 N and 100–1000 N, respectively, were applied at 5 Hz. Thermoelastic stress analysis measured changes in the sum of the surface principal stresses. Longitudinal stress profiles along the medial and lateral surfaces were evaluated descriptively. For each implanted specimen, the location of peak lateral stress was recorded relative to the stem tip. Results In both models, proximal surface stress decreased after implantation, and the peaks of the mean medial and lateral stress profiles shifted distally. In the normal model, distal stress changed little, and individual lateral peaks were located 50–82 mm proximal to the stem tip. In the osteoporotic model, distal stress increased on both surfaces. All three lateral peaks were within 16 mm of the stem tip, including one 16 mm distal to the tip. Conclusions Under the conditions tested, the osteoporotic specimens showed increased distal surface stress after implantation, with the lateral stress peak consistently located near the stem tip. Differences in stem size and loading conditions precluded attribution of this pattern to bone properties alone.

Research Square
University of Tsukuba (JP), Tsukuba Central Hospital (JP), National Institute of Advanced Industrial Science and Technology (JP)
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon