A patient-specific Winston-Lutz system for multi-target, single isocentre radiotherapy

Objective: To develop a quality assurance (QA) system for verifying patient-specific radiation targeting accuracy of radiotherapy linear accelerators (linacs) for multi-target, single isocentre stereotactic radiosurgery (SRS). Approach: The QA system consists of a phantom containing a 3D array of possible radiopaque target (BB) positions and associated software. Software determines the in-phantom BB positions most closely corresponding to the patient’s lesion locations. A dynamic conformal arc (DCA) plan was generated using the couch/collimator/gantry angles corresponding to the patient’s treatment plan. During plan delivery, images are acquired at 25 frames per second using the linac’s electronic portal imaging device (EPID). BB positions relative to the field centroid were detected using a Hough circle detection algorithm. The radiation field was detected using image thresholding. Main results: Validation with synthetic data showed BB position discrepancies of 0.23 mm (mean) and 0.53 mm (maximum), respectively, which do not exceed the synthetic phantom resolution, meaning that the test can be considered successful. Field area discrepancies were 2.60% (mean) and 3.33% (maximum) of the expected field size. Intentional translation and rotations were applied, demonstrating the sensitivity and accuracy of the system. A combined analysis of BB positions and field area discrepancies are necessary to fully assess linac targeting accuracy. As a further validation step, results for a single BB case were found to agree with the analysis results from independent software within 0.19 mm. For a 6 BB test case, maximum and mean 2D differences in expected and detected BB positions were 2.21 mm and 1.03 mm respectively. Field area discrepancies were 2.17 cm2 (mean) and 4.20 cm2 (maximum) or 3.75% and 7.98% of the expected field size. Significance: This QA system defines plan-specific machine isocentricity for single-isocentre multi-target SRS according to a chosen tolerance.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-11-23
DOI
https://doi.org/10.1088/2057-1976/aec569
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A patient-specific Winston-Lutz system for multi-target, single isocentre radiotherapy

R Lee MacDonald, Patricia A. K. Oliver, Kassia M. M. Schenkels
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

A patient-specific Winston-Lutz system for multi-target, single isocentre radiotherapy

R Lee MacDonald, Patricia A. K. Oliver, Kassia M. M. Schenkels
article en

Abstract

Objective: To develop a quality assurance (QA) system for verifying patient-specific radiation targeting accuracy of radiotherapy linear accelerators (linacs) for multi-target, single isocentre stereotactic radiosurgery (SRS). Approach: The QA system consists of a phantom containing a 3D array of possible radiopaque target (BB) positions and associated software. Software determines the in-phantom BB positions most closely corresponding to the patient’s lesion locations. A dynamic conformal arc (DCA) plan was generated using the couch/collimator/gantry angles corresponding to the patient’s treatment plan. During plan delivery, images are acquired at 25 frames per second using the linac’s electronic portal imaging device (EPID). BB positions relative to the field centroid were detected using a Hough circle detection algorithm. The radiation field was detected using image thresholding. Main results: Validation with synthetic data showed BB position discrepancies of 0.23 mm (mean) and 0.53 mm (maximum), respectively, which do not exceed the synthetic phantom resolution, meaning that the test can be considered successful. Field area discrepancies were 2.60% (mean) and 3.33% (maximum) of the expected field size. Intentional translation and rotations were applied, demonstrating the sensitivity and accuracy of the system. A combined analysis of BB positions and field area discrepancies are necessary to fully assess linac targeting accuracy. As a further validation step, results for a single BB case were found to agree with the analysis results from independent software within 0.19 mm. For a 6 BB test case, maximum and mean 2D differences in expected and detected BB positions were 2.21 mm and 1.03 mm respectively. Field area discrepancies were 2.17 cm2 (mean) and 4.20 cm2 (maximum) or 3.75% and 7.98% of the expected field size. Significance: This QA system defines plan-specific machine isocentricity for single-isocentre multi-target SRS according to a chosen tolerance.

Biomedical Physics & Engineering Express
Dalhousie University (CA), Nova Scotia Health Authority (CA)
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon