Metric-Dependent Dosimetric Divergence Between 106Ru and 125I Eye Plaques: A TOPAS Monte Carlo Dose-Volume Histogram Analysis

Uveal melanoma is commonly treated with episcleral plaque brachytherapy, yet photon-emitting and beta-emitting plaques are rarely compared across the full clinically relevant range of tumor depth. This study compared dose-volume histogram (DVH) characteristics of an 125 I Collaborative Ocular Melanoma Study (COMS) plaque and an 106 Ru Concave Circular Applicator (CCA) plaque using TOPAS Monte Carlo simulations in an eye phantom. Seven tumor apical heights (4 to 12 mm) were simulated. Each plaque was independently normalized to 85 Gy at 4 mm depth on the central axis, a reference achievable by both sources. Tumor and organ-at-risk DVHs were extracted for the lens, cornea, and optic nerve. The 106 Ru plaque delivered a higher median tumor dose than 125 I through 7 mm (502.8 vs 202.7 Gy at 4 mm; 157.2 vs 97.2 Gy at 7 mm), but its apex dose was lower at every height beyond 4 mm (2.47 vs 35.84 Gy at 7 mm; 1.86 vs 29.25 Gy at 8 mm). The D₅₀/D₉₈ heterogeneity ratio peaked at 52.6 at 7 mm for ¹⁰⁶Ru vs 2.5 for 125 I. Mean organ-at-risk doses were approximately tenfold lower with 106 Ru at all heights (lens 1.63 vs 15.8 Gy). The two plaques are indistinguishable by prescription-dose coverage, and a comparison based on that metric alone would incorrectly conclude that they are equivalent. Photon and beta plaques differ in the shape rather than the depth of their intratumoral dose distributions, and comparative plaque dosimetry should report sub-prescription coverage and near-minimum dose rather than prescription-dose coverage alone.

Authors

Institutions

Publication Details

Journal
Research Square
Published
2026-11-23
DOI
https://doi.org/10.21203/rs.3.rs-11724309/v1
Primary Topic
Advanced Radiotherapy Techniques
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Metric-Dependent Dosimetric Divergence Between 106Ru and 125I Eye Plaques: A TOPAS Monte Carlo Dose-Volume Histogram Analysis

Parvin Ahmadi, Saleh Rouhi, Shahab Atefi
Research Square
Advanced Radiotherapy Techniques
preprint

Metric-Dependent Dosimetric Divergence Between 106Ru and 125I Eye Plaques: A TOPAS Monte Carlo Dose-Volume Histogram Analysis

Parvin Ahmadi, Saleh Rouhi, Shahab Atefi
preprint en

Abstract

Uveal melanoma is commonly treated with episcleral plaque brachytherapy, yet photon-emitting and beta-emitting plaques are rarely compared across the full clinically relevant range of tumor depth. This study compared dose-volume histogram (DVH) characteristics of an 125 I Collaborative Ocular Melanoma Study (COMS) plaque and an 106 Ru Concave Circular Applicator (CCA) plaque using TOPAS Monte Carlo simulations in an eye phantom. Seven tumor apical heights (4 to 12 mm) were simulated. Each plaque was independently normalized to 85 Gy at 4 mm depth on the central axis, a reference achievable by both sources. Tumor and organ-at-risk DVHs were extracted for the lens, cornea, and optic nerve. The 106 Ru plaque delivered a higher median tumor dose than 125 I through 7 mm (502.8 vs 202.7 Gy at 4 mm; 157.2 vs 97.2 Gy at 7 mm), but its apex dose was lower at every height beyond 4 mm (2.47 vs 35.84 Gy at 7 mm; 1.86 vs 29.25 Gy at 8 mm). The D₅₀/D₉₈ heterogeneity ratio peaked at 52.6 at 7 mm for ¹⁰⁶Ru vs 2.5 for 125 I. Mean organ-at-risk doses were approximately tenfold lower with 106 Ru at all heights (lens 1.63 vs 15.8 Gy). The two plaques are indistinguishable by prescription-dose coverage, and a comparison based on that metric alone would incorrectly conclude that they are equivalent. Photon and beta plaques differ in the shape rather than the depth of their intratumoral dose distributions, and comparative plaque dosimetry should report sub-prescription coverage and near-minimum dose rather than prescription-dose coverage alone.

Research Square
Shahed University (IR), University of Pavia (IT), Kermanshah University of Medical Sciences (IR)
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon