Off-axis Winston-Lutz quality assurance for static and rotational delivery on a gimbal-mounted O-ring linear accelerator

Objective: OXRAY is a gimbal-mounted O-ring linear accelerator that can shift the radiation isocenter away from the machine isocenter and deliver biaxial rotational dynamic radiation therapy (BROAD-RT). Published commissioning studies have characterized dynamic beam positioning with in-house analysis, but static gimbal-shifted positions have not been evaluated with a commercially available quality assurance system. We assessed a commercial Winston-Lutz (WL) and off-axis Winston-Lutz (OAWL) procedure for field-to-target coincidence at the machine isocenter and at ±30 mm along the superior-inferior axis. Approach: A MultiMet-WL Cube phantom and DoseLab Pro were used to analyze cine electronic portal imaging device images acquired during static four-port irradiation, 360° gantry rotations, and a BROAD-RT trajectory. Main results: Static 2D displacement was 0.38 ± 0.20 mm (maximum 0.73 mm) at the machine isocenter, 0.45 ± 0.15 mm (0.70 mm) at +30 mm, and 0.45 ± 0.20 mm (0.79 mm) at -30 mm. During gantry rotation, mean displacements were 0.25-0.34 mm and the maximum was 0.68 mm. BROAD-RT yielded 0.32 ± 0.16 mm with a maximum of 0.65 mm. Clockwise and counterclockwise rotations showed similar angle-dependent profiles within the measurement session. The maximum recorded-minus-commanded gimbal tilt residual was 0.004°, showing high agreement within the control system. All measured values were below 1 mm. Significance: This commercial procedure directly tests static gimbal-shifted positions that are not covered by center-only WL testing and provides the same angle-resolved endpoint for rotational and BROAD-RT delivery without custom phantom hardware or analysis software.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-11-23
DOI
https://doi.org/10.1088/2057-1976/aec566
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Off-axis Winston-Lutz quality assurance for static and rotational delivery on a gimbal-mounted O-ring linear accelerator

Tomoki Kitagawa, Yasunori Ishiguro, Takahiro Aoyama, Takeshi Kodaira et al.
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

Off-axis Winston-Lutz quality assurance for static and rotational delivery on a gimbal-mounted O-ring linear accelerator

Tomoki Kitagawa, Yasunori Ishiguro, Takahiro Aoyama, Takeshi Kodaira, Yutaro Koide, Ryoma Tomoda, Maho Morishita
article en

Abstract

Objective: OXRAY is a gimbal-mounted O-ring linear accelerator that can shift the radiation isocenter away from the machine isocenter and deliver biaxial rotational dynamic radiation therapy (BROAD-RT). Published commissioning studies have characterized dynamic beam positioning with in-house analysis, but static gimbal-shifted positions have not been evaluated with a commercially available quality assurance system. We assessed a commercial Winston-Lutz (WL) and off-axis Winston-Lutz (OAWL) procedure for field-to-target coincidence at the machine isocenter and at ±30 mm along the superior-inferior axis. Approach: A MultiMet-WL Cube phantom and DoseLab Pro were used to analyze cine electronic portal imaging device images acquired during static four-port irradiation, 360° gantry rotations, and a BROAD-RT trajectory. Main results: Static 2D displacement was 0.38 ± 0.20 mm (maximum 0.73 mm) at the machine isocenter, 0.45 ± 0.15 mm (0.70 mm) at +30 mm, and 0.45 ± 0.20 mm (0.79 mm) at -30 mm. During gantry rotation, mean displacements were 0.25-0.34 mm and the maximum was 0.68 mm. BROAD-RT yielded 0.32 ± 0.16 mm with a maximum of 0.65 mm. Clockwise and counterclockwise rotations showed similar angle-dependent profiles within the measurement session. The maximum recorded-minus-commanded gimbal tilt residual was 0.004°, showing high agreement within the control system. All measured values were below 1 mm. Significance: This commercial procedure directly tests static gimbal-shifted positions that are not covered by center-only WL testing and provides the same angle-resolved endpoint for rotational and BROAD-RT delivery without custom phantom hardware or analysis software.

Biomedical Physics & Engineering Express
Aichi Cancer Center (JP)
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon