Effect of Nasal Continuous Positive Airway Pressure Delivered by RAM® Cannula Versus Standard Binasal Prongs on Diaphragm Electrical Activity in Very Low Birth Weight Preterm Infants: A Randomized Crossover Trial

Background/Objectives: The RAM cannula is widely used to deliver nasal continuous positive airway pressure (nCPAP) to preterm infants, but its non-occlusive fit permits leak and may reduce the distending pressure transmitted to the airway. Prescribing approximately 2 cmH2O more pressure than with conventional prongs is common but has not been validated physiologically. We compared diaphragm electrical activity (Edi) during nCPAP delivered through a RAM cannula and through Fisher & Paykel (F&P) short binasal prongs. Methods: In a randomized, open-label crossover trial (ClinicalTrials.gov NCT03121781), 20 clinically stable very low birth weight preterm infants on nCPAP received 2 h of nCPAP through each interface in randomized order. Prescribed positive end-expiratory pressure was 2 cmH2O higher during RAM cannula support. Edi was analyzed over the final 15 min of each period. The primary outcome was minimum (end-expiratory) Edi. Within-infant differences (RAM minus F&P) were analyzed with paired t tests. Results: All 20 infants completed both periods, with no missing outcome data. Minimum Edi was higher during RAM cannula support (mean (SD) 3.06 (1.85) versus 2.00 (1.12) microvolts; mean difference 1.06, 95% confidence interval (CI) 0.36 to 1.76; p = 0.005), as were peak Edi (2.02 microvolts, 0.80 to 3.24; p = 0.003) and neural respiratory rate (11.44 breaths/min, 8.07 to 14.82; p < 0.001); each was higher in all 20 infants. Mean fraction of inspired oxygen did not differ significantly (1.56 percentage points, −0.07 to 3.19; p = 0.06). No infant met a prespecified stopping criterion. Conclusions: In clinically stable very low birth weight preterm infants, nCPAP through a RAM cannula was associated with greater diaphragmatic electrical activity and a higher neural respiratory rate than binasal prongs, despite a 2 cmH2O higher prescribed pressure. A fixed 2 cmH2O increment cannot be assumed to restore equivalent support.

Authors

Institutions

Publication Details

Journal
Children
Published
2026-11-23
DOI
https://doi.org/10.3390/children13121633
Primary Topic
Neonatal Respiratory Health Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Effect of Nasal Continuous Positive Airway Pressure Delivered by RAM® Cannula Versus Standard Binasal Prongs on Diaphragm Electrical Activity in Very Low Birth Weight Preterm Infants: A Randomized Crossover Trial

Michael Dunn, Elizabeth V Asztalos, Patti Schurr, Maher Shahroor et al.
Children
Neonatal Respiratory Health Research
article

Effect of Nasal Continuous Positive Airway Pressure Delivered by RAM® Cannula Versus Standard Binasal Prongs on Diaphragm Electrical Activity in Very Low Birth Weight Preterm Infants: A Randomized Crossover Trial

Michael Dunn, Elizabeth V Asztalos, Patti Schurr, Maher Shahroor, Christer A. Sinderby, Nalini Singhal, Carolina López Sanguos, Norm Comtois, Jennifer Beck, Giselle Lai
article en

Abstract

Background/Objectives: The RAM cannula is widely used to deliver nasal continuous positive airway pressure (nCPAP) to preterm infants, but its non-occlusive fit permits leak and may reduce the distending pressure transmitted to the airway. Prescribing approximately 2 cmH2O more pressure than with conventional prongs is common but has not been validated physiologically. We compared diaphragm electrical activity (Edi) during nCPAP delivered through a RAM cannula and through Fisher & Paykel (F&P) short binasal prongs. Methods: In a randomized, open-label crossover trial (ClinicalTrials.gov NCT03121781), 20 clinically stable very low birth weight preterm infants on nCPAP received 2 h of nCPAP through each interface in randomized order. Prescribed positive end-expiratory pressure was 2 cmH2O higher during RAM cannula support. Edi was analyzed over the final 15 min of each period. The primary outcome was minimum (end-expiratory) Edi. Within-infant differences (RAM minus F&P) were analyzed with paired t tests. Results: All 20 infants completed both periods, with no missing outcome data. Minimum Edi was higher during RAM cannula support (mean (SD) 3.06 (1.85) versus 2.00 (1.12) microvolts; mean difference 1.06, 95% confidence interval (CI) 0.36 to 1.76; p = 0.005), as were peak Edi (2.02 microvolts, 0.80 to 3.24; p = 0.003) and neural respiratory rate (11.44 breaths/min, 8.07 to 14.82; p < 0.001); each was higher in all 20 infants. Mean fraction of inspired oxygen did not differ significantly (1.56 percentage points, −0.07 to 3.19; p = 0.06). No infant met a prespecified stopping criterion. Conclusions: In clinically stable very low birth weight preterm infants, nCPAP through a RAM cannula was associated with greater diaphragmatic electrical activity and a higher neural respiratory rate than binasal prongs, despite a 2 cmH2O higher prescribed pressure. A fixed 2 cmH2O increment cannot be assumed to restore equivalent support.

ChildrenVol. 13(12)
St. Michael's Hospital (CA), Sunnybrook Health Science Centre (CA), University of Toronto (CA), Complejo Hospitalario Universitario de Santiago (ES)
Neonatal Respiratory Health Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon